EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Biomechanical testing of virtual meniscus implants made from a bi-phasic silk fibroin-based hydrogel and polyurethane via finite element analysis

A C Moser, J Fritz, A Kesselring, F Schüssler, A Otahal, S Nehrer

Journal of the Mechanical Behavior of Biomedical Materials · 2025

DOI: 10.1016/j.jmbbm.2024.106830 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of the Mechanical Behavior of Biomedical Materials, 2025. Verfasst von A C Moser und 5 weiteren Autor:innen. Beteiligte Einrichtung: Centre for Regenerative Medicine, Department for Health Science, University for Continuing Education Krems, Krems an der Donau, Austria; Austrian Clu…. Hauptthemen laut MEDLINE: Finite Element Analysis, Fibroins, Polyurethanes, Meniscus. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 55 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Society for Research Promotion Lower Austria.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE: To investigate the suitability of different material compositions and structural designs for 3D-printed meniscus implants using finite element analysis (FEA) to improve joint function after meniscal injury and guide future implant development. DESIGN: This experimental study involved in-silico testing of a meniscus model developed from two materials: a specially formulated hydrogel composed of silk fibroin (SF), gelatine, and decellularized meniscus-derived extracellular matrix (MD-dECM), and polyurethane (PU) with stiffness levels of 54 and 205 MPa. Both single-material implants and a two-volumetric meniscus model with an SF/gelatine/MD-dECM core and a PU shell were analysed using FEA to simulate the biomechanical performance under physiological conditions. RESULTS: The hydrogel alone was found to be unsuitable for long-term use due to instability in material properties beyond two weeks. PU 54 closely replicated the biomechanical properties of an intact meniscus, particularly in terms of contact pressure and stress distribution. However, hybrid implants combining PU 54 with hydrogel showed potential but required further optimization to reduce stress peaks. In contrast, implants with a PU 205 shell generated higher induced stresses, increasing the risk of material failure. CONCLUSIONS: FEA proves to be a valuable tool in the design and optimization of meniscal implants. The findings suggest that softer PU 54 is a promising material for mimicking natural meniscus properties, while stiffer materials may require design modifications to mitigate stress concentrations. These insights are crucial for refining implant designs and selecting appropriate material combinations before physical prototype production, potentially reducing costs, time, and the risk of implant failure.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Finite Element AnalysisFibroinsPolyurethanesMeniscusMaterials TestingHydrogelsMechanical TestsBiomechanical PhenomenaProstheses and ImplantsStress, MechanicalMechanical Phenomena

Stichwörter der Autor:innen

3D printingFinite element analysisMeniscal implantsPolyurethaneSilk fibroin hydrogel

Substanzen

FibroinsPolyurethanesHydrogels

Autor:innen und Institutionen

  • Centre for Regenerative Medicine, Department for Health Science, University for Continuing Education Krems, Krems an der Donau, Austria; Austrian Cluster for Tissue Regeneration, Austria. Electronic address: anna-christ…
  • Centre for Regenerative Medicine, Department for Health Science, University for Continuing Education Krems, Krems an der Donau, Austria; Austrian Cluster for Tissue Regeneration, Austria.
  • Department of Industrial Engineering, Fachhochschule Technikum Vienna, Vienna, Austria.

Förderung

  • Society for Research Promotion Lower Austria

Zitieren

APA

Moser, A. C., Fritz, J., Kesselring, A., Schüssler, F., Otahal, A., & Nehrer, S. (2025). Biomechanical testing of virtual meniscus implants made from a bi-phasic silk fibroin-based hydrogel and polyurethane via finite element analysis. Journal of the Mechanical Behavior of Biomedical Materials. https://doi.org/10.1016/j.jmbbm.2024.106830

Vancouver

Moser AC, Fritz J, Kesselring A, Schüssler F, Otahal A, Nehrer S. Biomechanical testing of virtual meniscus implants made from a bi-phasic silk fibroin-based hydrogel and polyurethane via finite element analysis. Journal of the Mechanical Behavior of Biomedical Materials. 2025. doi:10.1016/j.jmbbm.2024.106830

BibTeX
@article{Moser2025,
  author = {Moser, A C and Fritz, J and Kesselring, A and Schüssler, F and Otahal, A and Nehrer, S},
  title = {Biomechanical testing of virtual meniscus implants made from a bi-phasic silk fibroin-based hydrogel and polyurethane via finite element analysis},
  journal = {Journal of the Mechanical Behavior of Biomedical Materials},
  year = {2025},
  doi = {10.1016/j.jmbbm.2024.106830}
}
Kostenlos starten →