Impact of different pressures on the drying time of flexible endoscopes

W Q Hu, J Y Li, C Y Zhou, Y X Ge, Q Gu

Journal of Hospital Infection · 2025

DOI: 10.1016/j.jhin.2025.04.007 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Hospital Infection, 2025. Verfasst von W Q Hu und 4 weiteren Autor:innen. Beteiligte Einrichtung: Nursing Department, First Affiliated Hospital of Zhejiang University School of Medicine, Hangzhou, Zhejiang, China. Hauptthemen laut MEDLINE: Endoscopes, Pressure, Desiccation. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: The complex structure of endoscopes causes non-uniform drying conditions. However, current clinical drying practices lack precise guidance, leading to the insufficient drying of some endoscopes, which increases the risk of infection. The pressure of compressed air has been shown to influence the effectiveness of endoscope drying; however, specific requirements are absent. AIM: To determine the time required for the sufficient drying of different endoscopes under various pressure conditions. METHODS: Six different types of commonly used Olympus flexible endoscopes were selected, and for each type, multiple drying time groups were set under pressures of 0.102 and 0.204 MPa, with 10 endoscopes observed in each group. FINDINGS: Drying times varied according to the type and pressure. At 0.102 MPa, the instrument channel drying ranged from 80 s to 160 s (average: 123 s), while suction channel ranged from 260 s to 540 s (average: 428 s). At 0.204 MPa, the instrument channel dried in 50 s-90 s (average: 72 s), and suction channel in 130 s-230 s (average: 186 s). CONCLUSION: Optimal drying times for endoscopes differ according to the type and pressure. A higher pressure (0.204 MPa) efficiently removes moisture and speeds up drying. Longer endoscopes with the same biopsy channel diameter require longer drying time. In addition, the instrument channel dries faster than the suction channel.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

EndoscopesPressureDesiccationTime FactorsHumans

Stichwörter der Autor:innen

Air injection timeDryingFlexible endoscopeInfection controlInstrument channelSuction channel

Autor:innen und Institutionen

  • Nursing Department, First Affiliated Hospital of Zhejiang University School of Medicine, Hangzhou, Zhejiang, China.
  • Nursing Department, First Affiliated Hospital of Zhejiang University School of Medicine, Hangzhou, Zhejiang, China. Electronic address: 1191016@zju.edu.cn.

Zitieren

APA

Hu, W. Q., Li, J. Y., Zhou, C. Y., Ge, Y. X., & Gu, Q. (2025). Impact of different pressures on the drying time of flexible endoscopes. Journal of Hospital Infection. https://doi.org/10.1016/j.jhin.2025.04.007

Vancouver

Hu WQ, Li JY, Zhou CY, Ge YX, Gu Q. Impact of different pressures on the drying time of flexible endoscopes. Journal of Hospital Infection. 2025. doi:10.1016/j.jhin.2025.04.007

BibTeX
@article{Hu2025,
  author = {Hu, W Q and Li, J Y and Zhou, C Y and Ge, Y X and Gu, Q},
  title = {Impact of different pressures on the drying time of flexible endoscopes},
  journal = {Journal of Hospital Infection},
  year = {2025},
  doi = {10.1016/j.jhin.2025.04.007}
}
Kostenlos starten →