Application of portable liquid scintillation counter for on-field measurement of tritium in aqueous samples during radiation emergency

Priyanka J Reddy, Rajesh Sankhla, Probal Chaudhury

Journal of Environmental Radioactivity · 2024

DOI: 10.1016/j.jenvrad.2023.107330 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Environmental Radioactivity, 2024. Verfasst von Priyanka J Reddy und 2 weiteren Autor:innen. Beteiligte Einrichtung: Radiation Safety Systems Division, Bhabha Atomic Research Centre, Mumbai, 400085, India. Electronic address: pjreddy@barc.gov.in. Hauptthemen laut MEDLINE: Radiation Monitoring, Drinking Water. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 27 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Screening of aqueous samples for 3H contamination is required to decide suitability of water intended for human consumption during radiation emergency. BARC, Trombay has recently procured Portable Triathler liquid scintillation counter as a screening tool for on-site response to radiation emergency. For this purpose, parameters like, different available scintillators, scintillator to sample ratio and influence of different capacity scintillation vials were optimized for 3H concentration measurement. The study indicated that for 7 mL vials, the optimized scintillator to sample ratio was 4:1 for Optiphase Hisafe II, Ultima Gold Ultra Low Level Tritium, Aqualight AB and Dioxane based scintillators whereas for Ultima Gold AB and Optiphase Hisafe III scintillators it was 3:1. In case of 20 mL vials, the ratio was optimized to 8:1 for all the above mentioned six scintillators. Additionally, the effect of applying counting efficiency using calibration curves generated using distilled water and that obtained using optimized scintillator to sample ratio in various spiked water samples was also studied. Results in 3H concentration (2000-5700 Bq/mL) indicate about ±10% deviation using both the methods for commercially available scintillators. However, in case of Dioxane based scintillator, the deviation was higher (20%). For analysis of 3H concentration ∼200 Bq/mL using commercially available scintillators results in higher deviation of about 21% due to unreliable quench indicating parameter for count rate less than 4000. The study indicated that calibration curves constructed using optimized parameters are universally applicable for determination of 3H concentration in wide variety of water samples.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiation MonitoringDrinking WaterHumansTritiumScintillation CountingDioxanes

Substanzen

TritiumDrinking WaterDioxanes

Autor:innen und Institutionen

  • Radiation Safety Systems Division, Bhabha Atomic Research Centre, Mumbai, 400085, India. Electronic address: pjreddy@barc.gov.in.
  • Radiation Safety Systems Division, Bhabha Atomic Research Centre, Mumbai, 400085, India; Homi Bhabha National Institute, Trombay, Mumbai, 400085, India.
  • Radiation Safety Systems Division, Bhabha Atomic Research Centre, Mumbai, 400085, India.

Zitieren

APA

Reddy, P. J., Sankhla, R., & Chaudhury, P. (2024). Application of portable liquid scintillation counter for on-field measurement of tritium in aqueous samples during radiation emergency. Journal of Environmental Radioactivity. https://doi.org/10.1016/j.jenvrad.2023.107330

Vancouver

Reddy PJ, Sankhla R, Chaudhury P. Application of portable liquid scintillation counter for on-field measurement of tritium in aqueous samples during radiation emergency. Journal of Environmental Radioactivity. 2024. doi:10.1016/j.jenvrad.2023.107330

BibTeX
@article{Reddy2024,
  author = {Reddy, Priyanka J and Sankhla, Rajesh and Chaudhury, Probal},
  title = {Application of portable liquid scintillation counter for on-field measurement of tritium in aqueous samples during radiation emergency},
  journal = {Journal of Environmental Radioactivity},
  year = {2024},
  doi = {10.1016/j.jenvrad.2023.107330}
}
Kostenlos starten →