Stability analysis of stock alloplastic temporomandibular joint prostheses based on clinical cases of screw loosening
Journal of Cranio-Maxillofacial Surgery · 2026
DOI: 10.1016/j.jcms.2026.109911 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Cranio-Maxillofacial Surgery, 2026. Verfasst von Ruoyi Ding und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Oral Surgery, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine, College of Stomatology, Shanghai Jiao…. Hauptthemen laut MEDLINE: Bone Screws, Joint Prosthesis, Temporomandibular Joint, Prosthesis Failure. Die Arbeit nennt 28 Literaturangaben. Gefördert von: School of Medicine, Shanghai Jiao Tong University, National Natural Science Foundation of China.
Abstract (Originalsprache)
Screw loosening is one of the common complications that affect stability after stock alloplastic temporomandibular joint (TMJ) replacement. Based on clinically observed cases of screw loosening, this study aimed to evaluate how prosthesis installation patterns influence screw loading and to provide recommendations for installation safety. A three-dimensional finite element model of unilateral total TMJ replacement was constructed to analyze Biomet standard TMJ prosthesis stability under the following situations: 1) fossa component installed at different angles (0°-20°) relative to the bone surface; 2) mandibular component installed at different angles (0°-15°) from the posterior border of the ramus. Under maximum occlusal loading and mouth-opening muscle loading, the characteristics of stress and strain, force on the fixation screws and displacement were analysed. The ratio of the maximum screw pull-out strength from the bone tested in vitro to the force on the screws was calculated to predict the safe installation angle. As the angle between the fossa component and the bone surface increased, the stress, strain and force on the prosthesis and fixation screws gradually enlarged especially the fossa screws increased by 3.05 to 9.36 times. As the angle between the mandibular prosthesis and the posterior border of the ramus increased, the stress, strain and force on the prosthesis and screws enlarged especially the fossa screws increased by 2.47 times to 4.54 times. A reduction in fixation safety was already observed at an installation angle of 5°, with individual fossa screws showing FoS values below 4. As the installation angle further increased, screw loading increased and the number of screws with reduced safety margins increased. These findings suggest that increasing component malposition progressively compromises fixation safety. The fossa component should therefore be adapted as closely as possible to the bone surface, and the mandibular component should be positioned as parallel as possible to the posterior border of the ramus.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Oral Surgery, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine, College of Stomatology, Shanghai Jiao Tong University, National Center for Stomatology, National Clinical R…
- Shanghai Key Laboratory of Orthopaedic Implants, Department of Orthopaedic Surgery, Shanghai Ninth People's Hospital, Shanghai Jiao Tong University School of Medicine, Shanghai, 200011, China.
- Key Laboratory of Pressure System and Safety, MOE, School of Mechanical and Power Engineering, East China University of Science and Technology, Shanghai, 200237, China. Electronic address: nwang@ecust.edu.cn.
Förderung
- School of Medicine, Shanghai Jiao Tong University
- National Natural Science Foundation of China
Zitieren
APA
Ding, R., Hong, J., Xue, W., Lu, C., Zhao, J., Wang, N., & He, D. (2026). Stability analysis of stock alloplastic temporomandibular joint prostheses based on clinical cases of screw loosening. Journal of Cranio-Maxillofacial Surgery. https://doi.org/10.1016/j.jcms.2026.109911
Vancouver
Ding R, Hong J, Xue W, Lu C, Zhao J, Wang N, et al. Stability analysis of stock alloplastic temporomandibular joint prostheses based on clinical cases of screw loosening. Journal of Cranio-Maxillofacial Surgery. 2026. doi:10.1016/j.jcms.2026.109911
BibTeX
@article{Ding2026,
author = {Ding, Ruoyi and Hong, Jiale and Xue, Wendong and Lu, Chuan and Zhao, Jieyun and Wang, Ning and He, Dongmei},
title = {Stability analysis of stock alloplastic temporomandibular joint prostheses based on clinical cases of screw loosening},
journal = {Journal of Cranio-Maxillofacial Surgery},
year = {2026},
doi = {10.1016/j.jcms.2026.109911}
}
EUVIMED