Multisize Electrode Field-of-View
JACC: Clinical Electrophysiology · 2024
DOI: 10.1016/j.jacep.2023.12.003 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift JACC: Clinical Electrophysiology, 2024. Verfasst von Sharif Omara und 14 weiteren Autor:innen. Beteiligte Einrichtung: Willem Einthoven Center for Cardiac Arrhythmia Research and Management, Leiden, the Netherlands, and Aarhus, Denmark; Department of Cardiology, Leide…. Hauptthemen laut MEDLINE: Myocardial Infarction. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Biosense Webster Inc.
Abstract (Originalsprache)
BACKGROUND: Voltage mapping to detect ventricular scar is important for guiding catheter ablation, but the field-of-view of unipolar, bipolar, conventional, and microelectrodes as it relates to the extent of viable myocardium (VM) is not well defined. OBJECTIVES: The purpose of this study was to evaluate electroanatomic voltage-mapping (EAVM) with different-size electrodes for identifying VM, validated against high-resolution ex-vivo cardiac magnetic resonance (HR-LGE-CMR). METHODS: A total of 9 swine with early-reperfusion myocardial infarction were mapped with the QDOT microcatheter. HR-LGE-CMR (0.3-mm slices) were merged with EAVM. At each EAVM point, the underlying VM in multisize transmural cylinders and spheres was quantified from ex vivo CMR and related to unipolar and bipolar voltages recorded from conventional and microelectrodes. RESULTS: In each swine, 220 mapping points (Q1, Q3: 216, 260 mapping points) were collected. Infarcts were heterogeneous and nontransmural. Unipolar and bipolar voltage increased with VM volumes from >175 mm3 up to >525 mm3 (equivalent to a 5-mm radius cylinder with height >6.69 mm). VM volumes in subendocardial cylinders with 1- or 3-mm depth correlated poorly with all voltages. Unipolar voltages recorded with conventional and microelectrodes were similar (difference 0.17 ± 2.66 mV) and correlated best to VM within a sphere of radius 10 and 8 mm, respectively. Distance-weighting did not improve the correlation. CONCLUSIONS: Voltage increases with transmural volume of VM but correlates poorly with small amounts of VM, which limits EAVM in defining heterogeneous scar. Microelectrodes cannot distinguish thin from thick areas of subendocardial VM. The field-of-view for unipolar recordings for microelectrodes and conventional electrodes appears to be 8 to 10 mm, respectively, and unexpectedly similar.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Willem Einthoven Center for Cardiac Arrhythmia Research and Management, Leiden, the Netherlands, and Aarhus, Denmark; Department of Cardiology, Leiden University Medical Center, Leiden, the Netherlands.
- Willem Einthoven Center for Cardiac Arrhythmia Research and Management, Leiden, the Netherlands, and Aarhus, Denmark; Department of Cardiology, Aarhus University Hospital, Aarhus, Denmark.
- Department of Mathematics, KU Leuven campus Kortrijk, Kortrijk, Belgium.
- Department of Physics and Astronomy, Ghent University, Ghent, Belgium.
- Department of Cardiology, Leiden University Medical Center, Leiden, the Netherlands.
- Department of Imaging Physics, Delft University of Technology, Delft, the Netherlands.
Förderung
- Biosense Webster Inc
Zitieren
APA
Omara, S., Glashan, C. A., Tofig, B. J., Leenknegt, L., Dierckx, H., Panfilov, A. V., Beukers, H. K. C., van Waasbergen, M. H., Tao, Q., Stevenson, W. G., Nielsen, J. C., Lukac, P., Kristiansen, S. B., van der Geest, R. J., & Zeppenfeld, K. (2024). Multisize Electrode Field-of-View. JACC: Clinical Electrophysiology. https://doi.org/10.1016/j.jacep.2023.12.003
Vancouver
Omara S, Glashan CA, Tofig BJ, Leenknegt L, Dierckx H, Panfilov AV, et al. Multisize Electrode Field-of-View. JACC: Clinical Electrophysiology. 2024. doi:10.1016/j.jacep.2023.12.003
BibTeX
@article{Omara2024,
author = {Omara, Sharif and Glashan, Claire A and Tofig, Bawer J and Leenknegt, Lore and Dierckx, Hans and Panfilov, Alexander V and Beukers, Hans K C and van Waasbergen, Michiel H and Tao, Qian and Stevenson, William G and Nielsen, Jens C and Lukac, Peter and Kristiansen, Steen B and van der Geest, Rob J and Zeppenfeld, Katja},