A simple electrochemical immunosensor based on MWCNTs-COOH/Fc-COOH@CoAl-LDH, nanocomposite for sensitive detection of the tumor marker CA724

Meng Miao, Ruyi Zhang, Hongmin Gao, Hehua Zhang, Longbin Que, Xin Gu, Dong Chang, Hongzhi Pan

International Immunopharmacology · 2024

DOI: 10.1016/j.intimp.2024.113406 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Immunopharmacology, 2024. Verfasst von Meng Miao und 7 weiteren Autor:innen. Beteiligte Einrichtung: Shanghai university of medicine & health Sciences Affiliated Zhoupu hospital, Shanghai 201318 China; School of Medical Technology, Shanghai Universit…. Hauptthemen laut MEDLINE: Nanotubes, Carbon, Electrochemical Techniques, Biosensing Techniques, Nanocomposites. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 48 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

A novel label-free electrochemical immunosensor for the ultrasensitive determination of cancer antigen (CA724) was developed using a novel composite of CoAl layered double hydroxides (CoAl-LDH) and carboxyl-functionalized multiwall carbon nanotubes (MWCNTs-COOH) as platform and ferrocenecarboxylic acid (Fc-COOH) as signal label. The MWCNTs-COOH/Fc-COOH@CoAl-LDH composite was prepared by a convenient and simple one-step ultrasonic method, and various characterization techniques consisting of scanning electron microscopy (SEM), transimission electronic microscopy (TEM), TEM-Mapping, fourier transform infrared (FT-IR), X-ray diffraction (XRD) and X-ray photoelectronic energy spectrum (XPS) were applied to study the size and morphological features. Due to its large specific surface area and multilayer structure, the CoAl-LDH can be post-doped to embed a large amount of signal probe to realize the amplification of the internal reference signal Fc. In addition, the higher conductivity of MWCNTs-COOH compensates for the deficiency of CoAl-LDH, which effectively strengthened the electron transfer efficiency of electrochemical signaling substances. The optimal experimental conditions were detected to be 2.5 mmol of Fc-COOH, 4.0 mg/mL of concentration, pH 6.0, incubation time of 40 min, and incubation temperature of 37 ℃. Under optimal conditions, the fabricated sensor exhibits linearity in a wide dynamic range covering the physiological concentration, from 0.001 to 100 U/mL and the limit of detection (LOD) was 0.03962 mU/mL, the calibration equation is stated as △I = 7.76363 log10CCA724 + 40.50351 (R2 = 0.99674). The sensor is successfully detects trace levels of CA724 in human serum with excellent recovery rates ranging from 100.52 %-102.30 %, proving the synergy of MWCNTs-COOH/Fc-COOH@CoAl-LDH as a promising platform for electrochemical sensing for clinical detection of other disease markers.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Nanotubes, CarbonElectrochemical TechniquesBiosensing TechniquesNanocompositesFerrous CompoundsLimit of DetectionMetallocenesHumansImmunoassayBiomarkers, TumorHydroxidesAntigens, Tumor-Associated, CarbohydrateAntigens, NeoplasmAntibodies, Immobilized

Stichwörter der Autor:innen

Cancer antigen 724Carboxyl-functionalized multiwall carbon nanotubeElectrochemical immunosensorFerrocenecarboxylic acidLabel-freeLayered double hydroxides

Substanzen

Nanotubes, Carbonferrocenecarboxylic acidFerrous CompoundsMetallocenesBiomarkers, TumorCA-72-4 antigenHydroxidesAntigens, Tumor-Associated, CarbohydrateAntigens, NeoplasmAntibodies, Immobilized

Autor:innen und Institutionen

  • Shanghai university of medicine & health Sciences Affiliated Zhoupu hospital, Shanghai 201318 China; School of Medical Technology, Shanghai University of Medicine & Health Sciences, Shanghai 201318 China.
  • School of Medical Technology, Shanghai University of Medicine & Health Sciences, Shanghai 201318 China.
  • School of Medical Instrument and Food Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China.
  • Collaborative Research Center, Shanghai University of Medicine and Health Sciences, Shanghai 201318, China.
  • Department of Laboratory Medicine, Shanghai Pudong Hospital, Shanghai 201399, China. Electronic address: dongchang1969@163.com.
  • Shanghai university of medicine & health Sciences Affiliated Zhoupu hospital, Shanghai 201318 China; Collaborative Research Center, Shanghai University of Medicine and Health Sciences, Shanghai 201318, China. Electronic…

Zitieren

APA

Miao, M., Zhang, R., Gao, H., Zhang, H., Que, L., Gu, X., Chang, D., & Pan, H. (2024). A simple electrochemical immunosensor based on MWCNTs-COOH/Fc-COOH@CoAl-LDH, nanocomposite for sensitive detection of the tumor marker CA724. International Immunopharmacology. https://doi.org/10.1016/j.intimp.2024.113406

Vancouver

Miao M, Zhang R, Gao H, Zhang H, Que L, Gu X, et al. A simple electrochemical immunosensor based on MWCNTs-COOH/Fc-COOH@CoAl-LDH, nanocomposite for sensitive detection of the tumor marker CA724. International Immunopharmacology. 2024. doi:10.1016/j.intimp.2024.113406

BibTeX
@article{Miao2024,
  author = {Miao, Meng and Zhang, Ruyi and Gao, Hongmin and Zhang, Hehua and Que, Longbin and Gu, Xin and Chang, Dong and Pan, Hongzhi},
  title = {A simple electrochemical immunosensor based on MWCNTs-COOH/Fc-COOH@CoAl-LDH, nanocomposite for sensitive detection of the tumor marker CA724},
  journal = {International Immunopharmacology},
  year = {2024},
  doi = {10.1016/j.intimp.2024.113406}
}
Kostenlos starten →