Validation of a Cable-Free and User-Friendly thermoAcoustic Range Verifier for a 2 Gy Proton Field Delivered to the Liver of an Anthropomorphic Phantom
International Journal of Radiation Oncology*Biology*Physics · 2026
DOI: 10.1016/j.ijrobp.2026.08.040 ↗Kurz gefasst
Validierungsstudie in der Zeitschrift International Journal of Radiation Oncology*Biology*Physics, 2026. Verfasst von Sarah K Patch und 4 weiteren Autor:innen. Beteiligte Einrichtung: Acoustic Range Estimates, Chicago, Ilinois; UW-Milwaukee, Department of Physics, Milwaukee, Wisconsin. Electronic address: Sarah.Patch@AcousticRangeE…. Hauptthemen laut MEDLINE: Phantoms, Imaging, Proton Therapy, Liver, Radiotherapy Planning, Computer-Assisted. Die Arbeit nennt 24 Literaturangaben. Gefördert von: National Institutes of Health, National Institutes of Health National Cancer Institute.
Abstract (Originalsprache)
PURPOSE: Detection and offline correction of proton range errors due to incorrect Hounsfield Unit-to-stopping power conversion with minimal (5-10 minute) extension of the patient-facing clinical workflow using a cable-free thermoAcoustic Range Verifier (tARV). METHODS AND MATERIALS: The accelerator (Mevion Hyperscan) delivered at a conventional dose rate. To generate quantitative results, the target was a multimodality anthropomorphic phantom (Sun Nuclear) composed primarily of hydrogels with known stoichiometry. The tARV includes 6 sonar receivers packed around the active imaging face of a wireless ultrasound probe (Clarius PAHD3). METHODS: Two dual-field plans were computed using Raystation 2023b's Monte Carlo (MC) engine. "Clinical MC" and "stoichiometric MC" plans were generated using clinical settings and by applying stoichiometry and range measurements (Ion Beam Applications (IBA), Zebra) of phantom hydrogels. We evaluated the posterior-to-anterior field that delivered approximately 2 Gy (maximum 27 proton pulses/spot) to 7 energy layers (135.6-162.8 MeV). The tARV was positioned anteriorly and distal to the Bragg peak, nearly in-line with the posterior-to-anterior beam trajectory, and rotated to collect a sagittal ultrasound image. Acoustic paths to receivers 5 to 6 were frequently obstructed by ribs. Measured thermoacoustic pulses generated by each spot were averaged to increase signal-to-noise ratio (SNR). Time shifts between measured and simulated pulses were used to compute range errors. Range errors were averaged with respect to energy layer. Clinical MC dose was recomputed by shifting each beamlet by the layer's average range error, and imported back into the treatment planning system. RESULTS: Overall, thermoacoustic estimates of range errors for clinical and stoichiometric dosemaps were 6.7±1.7mm and 0.5±1.6mm, respectively. Dose-volume histograms of recomputed and stoichiometric MC dosemaps were indistinguishable from each other but were clearly distinguishable from the clinical MC dose-volume histogram. CONCLUSIONS: Using thermoacoustic range estimates to recompute largely corrected the clinical dosemap. This preliminary, but quantitative study demonstrates that the tARV warrants and is poised for efficient evidence development, such as noninterventional patient safety and offline adaptive studies.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- Acoustic Range Estimates, Chicago, Ilinois; UW-Milwaukee, Department of Physics, Milwaukee, Wisconsin. Electronic address: Sarah.Patch@AcousticRangeEstimates.com.
- University of Utah, Department of Radiation Oncology, Salt Lake, Utah. Electronic address: Vikren.Sarkar@hci.utah.edu.
- Acoustic Range Estimates, Chicago, Ilinois. Electronic address: anjali.chiravuri@gmail.com.
- Phantoms and Alignment, Sun Nuclear Corporation, Norfolk, Virginia. Electronic address: rvallejos@mirion.com.
- Phantoms and Alignment, Sun Nuclear Corporation, Norfolk, Virginia. Electronic address: jolynch@mirion.com.
Förderung
- National Institutes of Health
- National Institutes of Health National Cancer Institute
Zitieren
APA
Patch, S. K., Sarkar, V., Chiravuri, A., Vallejos, R. J., & Lynch, J. E. (2026). Validation of a Cable-Free and User-Friendly thermoAcoustic Range Verifier for a 2 Gy Proton Field Delivered to the Liver of an Anthropomorphic Phantom. International Journal of Radiation Oncology*Biology*Physics. https://doi.org/10.1016/j.ijrobp.2026.08.040
Vancouver
Patch SK, Sarkar V, Chiravuri A, Vallejos RJ, Lynch JE. Validation of a Cable-Free and User-Friendly thermoAcoustic Range Verifier for a 2 Gy Proton Field Delivered to the Liver of an Anthropomorphic Phantom. International Journal of Radiation Oncology*Biology*Physics. 2026. doi:10.1016/j.ijrobp.2026.08.040
BibTeX
@article{Patch2026,
author = {Patch, Sarah K and Sarkar, Vikren and Chiravuri, Anjali and Vallejos, Roger J and Lynch, John E},
title = {Validation of a Cable-Free and User-Friendly thermoAcoustic Range Verifier for a 2 Gy Proton Field Delivered to the Liver of an Anthropomorphic Phantom},
journal = {International Journal of Radiation Oncology*Biology*Physics},
year = {2026},
doi = {10.1016/j.ijrobp.2026.08.040}
}
EUVIMED