Nanosecond and microsecond pulse field ablation for paroxysmal atrial fibrillation: Muscle response, procedural safety, and 6-month outcomes
Heart Rhythm · 2026-09 · 23 (9) · e1886-e1896
DOI: 10.1016/j.hrthm.2026.05.003 ↗Kurz gefasst
Fachartikel in der Zeitschrift Heart Rhythm, 2026-09. Verfasst von Lei Ding und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, State Key Laboratory of Cardiovascular Disease, Cardiovascular Institute, Fuwai Hospital, National Center for Cardiovascula…. Hauptthemen laut MEDLINE: Atrial Fibrillation, Catheter Ablation, Pulmonary Veins, Muscle Contraction. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 33 Literaturangaben. Gefördert von: National Natural Science Foundation of China.
Abstract (Originalsprache)
BACKGROUND: Nanosecond pulse field ablation (nanoPFA) is an emerging non-thermal modality that may reduce muscle contractions and enable atrial fibrillation (AF) ablation under conscious sedation. OBJECTIVE: This study aimed to compare the muscle response, safety, and 6-month outcomes between nanoPFA and microsecond PFA (microPFA). METHODS: Patients with symptomatic paroxysmal AF underwent nanoPFA under conscious sedation or microPFA under general anesthesia. In the nanoPFA group, muscle contraction, pain (Visual Analogue Scale [VAS]), and patient experience (Likert scale questionnaire) were assessed. Outcomes were evaluated at 3 and 6 months. Propensity score matching (PSM) was performed as sensitivity analysis. RESULTS: Among 151 patients (60.2 ± 11.4 years, 32.5% female), 57 received nanoPFA and 94 underwent microPFA. Pulmonary vein and superior vena cava isolation was completed in all, with no device-related complications. During nanoPFA, muscle contraction was absent in 43.9%, slight in 47.4%, and severe in 8.8%, most frequently at the right superior pulmonary vein. Most patients reported mild/no discomfort (VAS 0-4: 71.9% [41/57]); none reported severe pain, and 87.7% preferred conscious sedation. Six-month Kaplan-Meier freedom from any atrial tachyarrhythmia was 91.8% with nanoPFA and 83.9% with microPFA (log-rank P = .57); after PSM, 6-month freedom remained similar between groups (90.0% vs 88.9%, log-rank P = .56). CONCLUSION: NanoPFA performed under conscious sedation provides comparable safety and similar 6-month arrhythmia outcomes to microPFA under general anesthesia, with minimal pain and muscle responses, supporting the feasibility of a conscious-sedation workflow in our center.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, State Key Laboratory of Cardiovascular Disease, Cardiovascular Institute, Fuwai Hospital, National Center for Cardiovascular Diseases, Chinese Academy of Medical Sciences, and Peking Union Medi…
- Cardiac Center, The Affiliated Taian City Central Hospital of Qingdao University, Taian city, Shandong Province, China.
- Department of Cardiology, Jinling Hospital, Affiliated Hospital of Medical School, Nanjing University, Xuanwu District, Nanjing, China.
Förderung
- National Natural Science Foundation of China
Zitieren
APA
Ding, L., Xiang, B., Yu, F., Qi, Y., Jiang, Z., Zhang, K., Liu, Y., Li, J., Zhang, H., & Tang, M. (2026). Nanosecond and microsecond pulse field ablation for paroxysmal atrial fibrillation: Muscle response, procedural safety, and 6-month outcomes. Heart Rhythm, 23(9), e1886-e1896. https://doi.org/10.1016/j.hrthm.2026.05.003
Vancouver
Ding L, Xiang B, Yu F, Qi Y, Jiang Z, Zhang K, et al. Nanosecond and microsecond pulse field ablation for paroxysmal atrial fibrillation: Muscle response, procedural safety, and 6-month outcomes. Heart Rhythm. 2026;23(9):e1886-e1896. doi:10.1016/j.hrthm.2026.05.003
BibTeX
@article{Ding2026,
author = {Ding, Lei and Xiang, Bing and Yu, Fengyuan and Qi, Yingjie and Jiang, Zihan and Zhang, Kuo and Liu, Yuandong and Li, Jianhua and Zhang, Hongda and Tang, Min},
title = {Nanosecond and microsecond pulse field ablation for paroxysmal atrial fibrillation: Muscle response, procedural safety, and 6-month outcomes},