Improved implantable cardioverter-defibrillator shock efficacy using programmable pulse width

Michael S Katcher, Kevin Davis, Nima Badie, Luke C McSpadden, Ulrika Birgersdotter-Green, Jacqueline Joza, Robert Schaller, Gabriel Mouchawar

Heart Rhythm · 2026

DOI: 10.1016/j.hrthm.2026.04.035 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Heart Rhythm, 2026. Verfasst von Michael S Katcher und 7 weiteren Autor:innen. Beteiligte Einrichtung: Mass General Brigham-Salem Hospital, Salem, Massachusetts. Electronic address: mkatcher@mgb.org. Hauptthemen laut MEDLINE: Defibrillators, Implantable, Electric Countershock, Tachycardia, Ventricular, Ventricular Fibrillation. Die Arbeit nennt 6 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Abbott Laboratories, Abbott Pharmaceuticals.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Each phase of a biphasic implantable cardioverter-defibrillator (ICD) shock waveform can be characterized by either the percentage of peak voltage that decays (tilt) or the time taken for that decay (pulse width). Although shock programming has typically relied on fixing the tilt, patient-specific fixed pulse widths may improve shock success. OBJECTIVE: The purpose of this study was to compare shock energies and success rates to treat spontaneous ventricular tachycardia/ventricular fibrillation using ICD/CRT-D (cardiac resynchronization therapy - defibrillator) devices with fixed pulse width vs fixed tilt waveforms. METHODS: US patients with single-coil ICD or CRT-D devices (Ellipse, Assura, Unify, and Fortify, Abbott) were identified who had received defibrillation or cardioversion shock therapy (≥1 episode) when programmed to both 65% fixed tilt and fixed pulse width modes for distinct periods. Patient demographic characteristics, programming, and shock episodes were downloaded from the Merlin.net Patient Care Network (Abbott). Shock energies and success rates were compared across shock waveform types within each patient. RESULTS: In total, 277 patients experiencing at least 1 shock-delivered episode with each shock type were evaluated (mean age 70 ± 14 years; 157 [57%] with ICDs; 120 [43%] with CRT-Ds). The first-shock success rate was significantly greater with fixed pulse width vs fixed tilt (median [25th, 75th percentile] 100.0% [75.0%, 100.0%] of episodes vs 87.5% [42.1%, 100.0%] of episodes; P < .001), and more patients experienced 100% first-shock success with fixed pulse width vs fixed tilt (69.3% of episodes vs 46.2% of patients; P < .001). With fixed pulse width, 79.4% of patients (220 of 277) experienced greater-or-equal first-shock success, and they did so by delivering comparable energies (33.0 [29.6, 35.1] J vs 32.9 [25.7, 36.9] J; P = .53). CONCLUSION: Reprogramming shock waveforms from 65% fixed tilt to fixed pulse width significantly improved intrapatient shock success.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Defibrillators, ImplantableElectric CountershockTachycardia, VentricularVentricular FibrillationHumansFemaleMaleTreatment OutcomeMiddle AgedRetrospective StudiesAgedEquipment Design

Stichwörter der Autor:innen

Fixed pulse widthFixed tiltImplantable cardioverter-defibrillatorShock programmingShocks

Autor:innen und Institutionen

  • Mass General Brigham-Salem Hospital, Salem, Massachusetts. Electronic address: mkatcher@mgb.org.
  • Abbott, Sylmar, California.
  • University of California San Diego Health, San Diego, California.
  • McGill University Health Centre, Montreal, Quebec, Canada.
  • Penn Medicine, University of Pennsylvania, Philadelphia, Pennsylvania.

Förderung

  • Abbott Laboratories
  • Abbott Pharmaceuticals

Zitieren

APA

Katcher, M. S., Davis, K., Badie, N., McSpadden, L. C., Birgersdotter-Green, U., Joza, J., Schaller, R., & Mouchawar, G. (2026). Improved implantable cardioverter-defibrillator shock efficacy using programmable pulse width. Heart Rhythm. https://doi.org/10.1016/j.hrthm.2026.04.035

Vancouver

Katcher MS, Davis K, Badie N, McSpadden LC, Birgersdotter-Green U, Joza J, et al. Improved implantable cardioverter-defibrillator shock efficacy using programmable pulse width. Heart Rhythm. 2026. doi:10.1016/j.hrthm.2026.04.035

BibTeX
@article{Katcher2026,
  author = {Katcher, Michael S and Davis, Kevin and Badie, Nima and McSpadden, Luke C and Birgersdotter-Green, Ulrika and Joza, Jacqueline and Schaller, Robert and Mouchawar, Gabriel},
  title = {Improved implantable cardioverter-defibrillator shock efficacy using programmable pulse width},
  journal = {Heart Rhythm},
  year = {2026},
  doi = {10.1016/j.hrthm.2026.04.035}
}
Kostenlos starten →