Sine wave electroporation in patients with atrial fibrillation: Initial results of the BURST-AF study
Heart Rhythm · 2025
DOI: 10.1016/j.hrthm.2025.04.042 ↗Kurz gefasst
Multizentrische Studie in der Zeitschrift Heart Rhythm, 2025. Verfasst von Ante Anic und 10 weiteren Autor:innen. Beteiligte Einrichtung: Cardiovascular Department, University Hospital Center Split, Split, Croatia. Electronic address: anteanic@gmail.com. Hauptthemen laut MEDLINE: Atrial Fibrillation, Catheter Ablation, Electroporation, Heart Conduction System. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
BACKGROUND: Challenges or opportunities remain for improvement of pulsed field ablation (PFA) for atrial fibrillation (AF). OBJECTIVE: BURST-AF evaluated safety and lesion durability of a novel PFA system, Coherent Sine Burst Electroporation (CSE). METHODS: Patients with AF who failed ≥ 1 antiarrhythmic drug and undergoing de novo ablation were included. The CSE system includes a PFA generator and stylet-configurable catheter (circular, linear, focal). Pulmonary veins (PVs) were isolated with overlapping circular PFA applications. Ablation of additional lesions (posterior wall [PW]/cavotricuspid isthmus [CTI]) was performed at investigator discretion. Acute end point was isolation of PVs after a 20-minute waiting period. Chronic end point was isolation of PVs, bidirectional block (CTI), and/or absence of PW electrical activity at remap. Freedom from > 30 seconds of atrial arrhythmias was evaluated till 6 months. RESULTS: A total of 90 patients were enrolled (54% men, 62 years old, 60% paroxysmal AF). Targeted PVs (344) were isolated with 6.5 ± 1.7 PFA lesions/PV. PFA lesions for CTI and PW were 8.3 ± 4.7 and 14.4 ± 3.9, respectively. Procedure, catheter dwell, and fluoroscopy times were 93 ± 28 minutes, 54 ± 28 minutes, and 13.3 ± 5.5 minutes, respectively. Per PV isolation was 94.1% (317/337), and per patient isolation was 78.4% (69/88). Additional lesion set durability was 94.4% (102/108). At 6-month follow-up, 92.6% (63/68 patients) were free from atrial arrhythmias. One primary safety event of renal insufficiency was reported. CONCLUSION: CSE PFA system durably isolated PVs, PW, and CTI with a low rate of complications in this first-in-human study. (CSE System Pilot Study in Patients with Atrial Fibrillation - BURST-AF, NCT05572047).
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Cardiovascular Department, University Hospital Center Split, Split, Croatia. Electronic address: anteanic@gmail.com.
- Cardiovascular Department, University Hospital Center Zagreb, Zagreb, Croatia.
- Cardiovascular Department, Israeli-Georgian Medical Research Clinic, Tbilisi, Georgia.
- Cardiovascular Department, University Hospital Center Split, Split, Croatia.
- Argá Medtech, Inc, San Diego, California.
Zitieren
APA
Anic, A., Velagic, V., Papiashvili, G., Sikiric, I., Lorenzo, M., Prepolec, I., Aranza, I., Kapanadze, N., Dagelic, M., Breskovic, T., & Jurisic, Z. (2025). Sine wave electroporation in patients with atrial fibrillation: Initial results of the BURST-AF study. Heart Rhythm. https://doi.org/10.1016/j.hrthm.2025.04.042
Vancouver
Anic A, Velagic V, Papiashvili G, Sikiric I, Lorenzo M, Prepolec I, et al. Sine wave electroporation in patients with atrial fibrillation: Initial results of the BURST-AF study. Heart Rhythm. 2025. doi:10.1016/j.hrthm.2025.04.042
BibTeX
@article{Anic2025,
author = {Anic, Ante and Velagic, Vedran and Papiashvili, Giorgi and Sikiric, Ivan and Lorenzo, Melvin and Prepolec, Ivan and Aranza, Ivan and Kapanadze, Nino and Dagelic, Marija and Breskovic, Toni and Jurisic, Zrinka},