Significant hemolysis is present during irreversible electroporation of cardiomyocytes in vitro

Ivana Fiserova, Ondrej Fiser, Marek Novak, Jan Trnka, Antonia Gibalova, David Kvapil, Barbora Bacova, Marek Hozman, Dalibor Herman, Klara Benesova, Pavel Osmancik

Heart Rhythm · 2025

DOI: 10.1016/j.hrthm.2024.08.019 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Heart Rhythm, 2025. Verfasst von Ivana Fiserova und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biochemistry, Cell and Molecular Biology, Charles University, Prague, Czech Republic. Hauptthemen laut MEDLINE: Myocytes, Cardiac, Hemolysis, Electroporation, Atrial Fibrillation. Die Arbeit wurde bisher 36-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: European Union.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Pulsed field ablation (PFA) of atrial fibrillation is a new method in clinical practice. Despite a favorable safety profile of PFA in atrial fibrillation ablation, rare cases of renal failure, probably due to hemolysis, have recently been reported. OBJECTIVE: The aim of this study was to determine the rate of hemolysis and cardiac cell death during in vitro PFA with different electric field intensities. METHODS: Blood samples from healthy volunteers and mouse HL-1 cardiomyocyte cell lines were subjected to in vitro irreversible electroporation using 216 bipolar pulses, each lasting 2 μs with intervals of 5 μs, repeated 20 times at a frequency of 1 Hz. These pulses varied from 500 V to 1500 V. Cell-free hemoglobin levels were assessed spectrophotometrically, and red blood cell microparticles were evaluated by flow cytometry. Cardiomyocyte death was quantified with propidium iodide. RESULTS: Pulsed field energy (1000 V/cm, 1250 V/cm, and 1500 V/cm) was associated with a significant increase in cell-free hemoglobin (0.32 ± 0.16 g/L, 2.2 ± 0.96 g/L, and 5.7 ± 0.39 g/L; P < .01) and similar increase in the concentration of red blood cell microparticles. Significant rates of cardiomyocyte death were observed at electric field strengths of 750 V/cm, 1000 V/cm, 1250 V/cm, and 1500 V/cm (26.5% ± 5.9%, 44.3% ± 6.2%, 55.5% ± 6.9%, and 74.5% ± 17.8% of cardiomyocytes; P < .01). CONCLUSION: The most effective induction of cell death in vitro was observed at 1500 V/cm. This intensity was also associated with a significant degree of hemolysis.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Myocytes, CardiacHemolysisElectroporationAtrial FibrillationCatheter AblationHumansAnimalsMiceMale

Stichwörter der Autor:innen

Atrial fibrillationCardiomyocytesHemolysisIrreversible electroporationPulsed field ablation

Autor:innen und Institutionen

  • Department of Biochemistry, Cell and Molecular Biology, Charles University, Prague, Czech Republic.
  • Department of Biomedical Technology, Faculty of Biomedical Engineering, Czech Technical University in Prague, Prague, Czech Republic.
  • Department of Laboratory Hematology, Central Laboratories, University Hospital Kralovske Vinohrady, Prague, Czech Republic.
  • Department of Cardiology, University Hospital Kralovske Vinohrady, Charles University, Prague, Czech Republic.
  • Institute for Biostatistical Analyses, Masaryk University, Brno, Czech Republic.
  • Department of Cardiology, University Hospital Kralovske Vinohrady, Charles University, Prague, Czech Republic. Electronic address: pavel.osmancik@gmail.com.

Förderung

  • European Union

Zitieren

APA

Fiserova, I., Fiser, O., Novak, M., Trnka, J., Gibalova, A., Kvapil, D., Bacova, B., Hozman, M., Herman, D., Benesova, K., & Osmancik, P. (2025). Significant hemolysis is present during irreversible electroporation of cardiomyocytes in vitro. Heart Rhythm. https://doi.org/10.1016/j.hrthm.2024.08.019

Vancouver

Fiserova I, Fiser O, Novak M, Trnka J, Gibalova A, Kvapil D, et al. Significant hemolysis is present during irreversible electroporation of cardiomyocytes in vitro. Heart Rhythm. 2025. doi:10.1016/j.hrthm.2024.08.019

BibTeX
@article{Fiserova2025,
  author = {Fiserova, Ivana and Fiser, Ondrej and Novak, Marek and Trnka, Jan and Gibalova, Antonia and Kvapil, David and Bacova, Barbora and Hozman, Marek and Herman, Dalibor and Benesova, Klara and Osmancik, Pavel},
  title = {Significant hemolysis is present during irreversible electroporation of cardiomyocytes in vitro},
  journal = {Heart Rhythm},
  year = {2025},
  doi = {10.1016/j.hrthm.2024.08.019}
}
Kostenlos starten →