EUVIMEDEUVIMED entdecken
Randomisierte kontrollierte StudieMultizentrische Studie

Lesion delivery and scar formation in catheter ablation for atrial fibrillation: The DECAAF II trial

Nazem Akoum, Mario Mekhael, Felipe Bisbal, Oussama Wazni, Christopher McGann, Hyejung Lee, Tyler Bardsley, Tom Greene, J Michael Dean, Lilas Dagher, Eugene Kholmovski, Moussa Mansour und 15 weitere

Heart Rhythm · 2025

DOI: 10.1016/j.hrthm.2024.08.062 ↗

Kurz gefasst

Randomisierte kontrollierte Studie in der Zeitschrift Heart Rhythm, 2025. Verfasst von Nazem Akoum und 26 weiteren Autor:innen. Beteiligte Einrichtung: Cardiology Department, University of Washington Medical Center, Seattle, Washington. Hauptthemen laut MEDLINE: Atrial Fibrillation, Catheter Ablation, Pulmonary Veins, Cicatrix. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 24 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: The Efficacy of Delayed Enhancement MRI-Guided Fibrosis Ablation vs Conventional Catheter Ablation of Atrial Fibrillation randomized trial showed no difference in atrial fibrillation (AF) recurrence with additional delayed enhancement magnetic resonance imaging (DE-MRI) fibrosis-targeted ablation to pulmonary vein isolation (PVI) in persistent AF. OBJECTIVE: We evaluated the effect of lesion delivery on ablation-induced scarring and AF recurrence. METHODS: Lesions delivered, targeting fibrotic and nonfibrotic areas identified from preablation DE-MRI, were studied in relation to ablation-induced scarring on 3-month DE-MRI, including their association with arrhythmia recurrence. RESULTS: A total of 593 patients treated with radiofrequency were analyzed: 293 (49.4%) underwent PVI and 300 (50.6%) underwent additional fibrosis-guided ablation. Lesion analysis showed that 80.9% in the MRI fibrosis-guided group vs 16.5% in the PVI group (P < .001) had ≥40% of baseline fibrosis targeted. MRI assessment of ablation-induced scar showed that 44.8% of fibrosis-guided ablation and 15.5% of PVI had ≥40% of their fibrosis covered by scar (P < .001), demonstrating significant attenuation from lesions delivered to scar formed. In the overall population, fibrosis coverage with scar was not associated with recurrence (hazard ratio [HR] 0.90; 95% confidence interval [CI] 0.80-1.01; P = .08 per 20% increase). In patients with baseline fibrosis < 20%, fibrosis coverage with scar was associated with lower recurrence than PVI (HR 0.85; 95% CI 0.73-0.97; P = .03), whereas the association was not significant when baseline fibrosis ≥ 20% (HR 0.97; 95% CI 0.80-1.17; P = .77). Significant center variation was observed in fibrosis targeting and coverage with scarring. CONCLUSION: Radiofrequency ablation lesions do not uniformly result in scar formation. A post hoc analysis suggests reduced arrhythmia recurrence when ablation-induced scarring covers fibrotic regions in patients with low baseline fibrosis.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Atrial FibrillationCatheter AblationPulmonary VeinsCicatrixHumansMaleFemaleMiddle AgedRecurrenceTreatment OutcomeMagnetic Resonance ImagingFibrosisFollow-Up StudiesAged

Stichwörter der Autor:innen

Atrial fibrillationAtrial fibrosisCardiac magnetic resonance imagingCatheter ablationLesion delivery

Autor:innen und Institutionen

  • Cardiology Department, University of Washington Medical Center, Seattle, Washington.
  • Cardiology Department, Tulane University School of Medicine, New Orleans, Louisiana.
  • Heart Institute, University Hospital Germans Trias i Pujol, Barcelona, Spain.
  • Cleveland Clinic, Cleveland, Ohio.
  • Swedish Heart Center, Seattle, Washington.
  • University of Utah, Salt Lake City, Utah.

Zitieren

APA

Akoum, N., Mekhael, M., Bisbal, F., Wazni, O., McGann, C., Lee, H., Bardsley, T., Greene, T., Dean, J. M., Dagher, L., Kholmovski, E., Mansour, M., Marchlinski, F., Wilber, D., Hindricks, G., Mahnkopf, C., Wells, D., Jaïs, P., Sanders, P., …, & DECAAF II investigators (2025). Lesion delivery and scar formation in catheter ablation for atrial fibrillation: The DECAAF II trial. Heart Rhythm. https://doi.org/10.1016/j.hrthm.2024.08.062

Vancouver

Akoum N, Mekhael M, Bisbal F, Wazni O, McGann C, Lee H, et al. Lesion delivery and scar formation in catheter ablation for atrial fibrillation: The DECAAF II trial. Heart Rhythm. 2025. doi:10.1016/j.hrthm.2024.08.062

BibTeX
@article{Akoum2025,
  author = {Akoum, Nazem and Mekhael, Mario and Bisbal, Felipe and Wazni, Oussama and McGann, Christopher and Lee, Hyejung and Bardsley, Tyler and Greene, Tom and Dean, J Michael and Dagher, Lilas and Kholmovski, Eugene and Mansour, Moussa and Marchlinski, Francis and Wilber, David and Hindricks, Gerhard and Mahnkopf, Christian and Wells, Darryl and Jaïs, Pierre and Sanders, Prashanthan and Brachmann, Johannes and Bax, Jeroen J and Morrison-de Boer, Leonie and Deneke, Thomas and Calkins, Hugh and Sohns, Christian and Marrouche, Nassir and DECAAF II investigators, DECAAF II investigators},
  title = {Lesion delivery and scar formation in catheter ablation for atrial fibrillation: The DECAAF II trial},
  journal = {Heart Rhythm},
  year = {2025},
  doi = {10.1016/j.hrthm.2024.08.062}
}
Kostenlos starten →