Acute human defibrillation performance of a subcutaneous implantable cardioverter-defibrillator with an additional coil electrode

Sing-Chien Yap, Erik F J Oosterwerff, Lucas V A Boersma, Willeke van der Stuijt, Anneke Lenssen, Stephen J Hahn, Reinoud E Knops

Heart Rhythm · 2023

DOI: 10.1016/j.hrthm.2023.08.009 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Heart Rhythm, 2023. Verfasst von Sing-Chien Yap und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Thorax Center, Cardiovascular Institute, Erasmus Medisch Centrum, Rotterdam, The Netherlands. Electronic address: s.c.yap@e…. Hauptthemen laut MEDLINE: Cardiomyopathies, Defibrillators, Implantable. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Boston Scientific Corp.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: The subcutaneous implantable cardioverter-defibrillator (S-ICD) delivers 80 J shocks from an 8 cm left-parasternal coil to a 59 cm3 left lateral pulse generator (PG). A system that defibrillates with lower energy could significantly reduce PG size. Computer modeling and animal studies suggested that a second shock coil either parallel to the left-parasternal coil or transverse from the xiphoid to the PG pocket would significantly reduce the defibrillation threshold. OBJECTIVE: The purpose of this study was to acutely assess the defibrillation efficacy of parallel and transverse configurations in patients receiving an S-ICD. METHODS: Testing was performed in patients receiving a conventional S-ICD system. Success at 65 J was required before investigational testing. A second electrode was temporarily inserted from the xiphoid incision connected to the PG with an investigational Y-adapter. Phase 1 (n = 11) tested the parallel configuration. Phase 2 (n = 21) tested both parallel and transverse configurations in random order. RESULTS: This study enrolled 35 patients (28 males (80%); mean age 51 ± 17 years; left ventricular ejection fraction 40% ± 15%; body mass index 26 ± 4 kg/m2; prior myocardial infarction 46%; congestive heart failure 49%; cardiomyopathy 63%). Compared to the conventional S-ICD system, mean shock impedance decreased for both parallel (69 ± 15 Ω vs 86 ± 20 Ω; n = 33; P < .001) and transverse (56 ± 14 Ω vs 81 ± 21 Ω; n = 20; P < .001) configurations. Shock success rates at 20, 30, and 40 J were 55%, 79%, 97%, and 25%, 70%, 90% for parallel and transverse configurations, respectively. Defibrillation threshold testing was well tolerated with no serious adverse events. CONCLUSION: Adding a second shock coil, particularly in the parallel configuration, significantly reduced the impedance and had a high likelihood of defibrillation success at energies ≤40 J. This may enable the development of a smaller S-ICD.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

CardiomyopathiesDefibrillators, ImplantableAdultAgedHumansMaleMiddle AgedBody Mass IndexElectric CountershockStroke VolumeVentricular FibrillationVentricular Function, Left

Stichwörter der Autor:innen

DefibrillationElectrodesHumanImplantable cardioverter-defibrillatorSubcutaneous

Autor:innen und Institutionen

  • Department of Cardiology, Thorax Center, Cardiovascular Institute, Erasmus Medisch Centrum, Rotterdam, The Netherlands. Electronic address: s.c.yap@erasmusmc.nl.
  • Department of Cardiology, Isala Klinieken, Zwolle, The Netherlands.
  • Department of Cardiology, St Antonious Ziekenhuis, Nieuwegein, The Netherlands; Department of Cardiology, Amsterdam University Medical Center, Amsterdam, The Netherlands.
  • Department of Cardiology, Amsterdam University Medical Center, Amsterdam, The Netherlands.
  • Boston Scientific, Amsterdam, The Netherlands.
  • Boston Scientific CRM, St. Paul, Minnesota.

Förderung

  • Boston Scientific Corp

Zitieren

APA

Yap, S. C., Oosterwerff, E. F. J., Boersma, L. V. A., van der Stuijt, W., Lenssen, A., Hahn, S. J., & Knops, R. E. (2023). Acute human defibrillation performance of a subcutaneous implantable cardioverter-defibrillator with an additional coil electrode. Heart Rhythm. https://doi.org/10.1016/j.hrthm.2023.08.009

Vancouver

Yap SC, Oosterwerff EFJ, Boersma LVA, van der Stuijt W, Lenssen A, Hahn SJ, et al. Acute human defibrillation performance of a subcutaneous implantable cardioverter-defibrillator with an additional coil electrode. Heart Rhythm. 2023. doi:10.1016/j.hrthm.2023.08.009

BibTeX
@article{Yap2023,
  author = {Yap, Sing-Chien and Oosterwerff, Erik F J and Boersma, Lucas V A and van der Stuijt, Willeke and Lenssen, Anneke and Hahn, Stephen J and Knops, Reinoud E},
  title = {Acute human defibrillation performance of a subcutaneous implantable cardioverter-defibrillator with an additional coil electrode},
  journal = {Heart Rhythm},
  year = {2023},
  doi = {10.1016/j.hrthm.2023.08.009}
}
Kostenlos starten →