Pulmonary vein isolation using a novel balloon-in-basket pulsed field ablation system: The Lübeck how-to protocol
Heart Rhythm O2 · 2026-06 · 7 (6) · 1042-1052
DOI: 10.1016/j.hroo.2026.02.022 ↗Kurz gefasst
Studienprotokoll in der Zeitschrift Heart Rhythm O2, 2026-06. Verfasst von Roland Richard Tilz und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Rhythmology, University Heart Center Lübeck, University Hospital Schleswig-Holstein, Lübeck, Germany. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 13 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Single-shot pulsed field ablation (PFA) is increasingly used for pulmonary vein isolation (PVI) in atrial fibrillation (AF). Although balloon-in-basket (BiB)-PFA system integrates ablation and mapping into a single catheter, standardized procedural workflows remain limited.
Objective
This study aimed to describe a standardized, step-by-step workflow for AF ablation using the BiB-PFA system and evaluate procedural performance, acute safety, and the learning curve in clinical practice.
Methods
This prospective, single-center study included 40 consecutive patients undergoing index PVI using the BiB-PFA system at a tertiary high-volume electrophysiology center. Procedures followed standardized protocol including balloon preparation, ultrasound-guided access, fluoroscopy-guided transseptal puncture, selective pulmonary vein (PV) angiography, and a structured 2-step ablation approach for each vein. Posterior wall ablation (PWA) was performed in selected patients. Procedural metrics and acute complications were analyzed descriptively.
Results
Patients were 64.5 years old (56.5, 73.3); 41% were female, and 57.5% had nonparoxysmal AF. The median procedural duration was 41.5 minutes (35.0, 59.0), left atrial dwell time 27.5 minutes (25.0, 36.0), and BiB dwell time 23.0 minutes (20.0, 25.8). A median of 4 applications per PV was used. Reduced-energy delivery was required in 30% of right inferior and 25% of right superior PV. PWA was performed in 35% of patients using a median of 5 applications. No major intraprocedural complication occurred. The mean procedural duration decreased from 50.0 minutes (38.0, 60.5) in the first 10 cases to 34.5 minutes (31.5, 52.0) in the last 10 cases (P = .059).
Conclusion
A standardized BiB-PFA workflow enables efficient, safe, and reproducible PVI with optional PWA and can be rapidly implemented during early operator experience.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Rhythmology, University Heart Center Lübeck, University Hospital Schleswig-Holstein, Lübeck, Germany.
- German Center for Cardiovascular Research (DZHK), Partner Site Hamburg/Kiel/Lübeck, Lübeck, Germany.
Zitieren
APA
Tilz, R. R., Hatahet, S., Eitel, C., Popescu, S., Subin, B., Kuck, K. H., Mamaev, R., & Wenzel, J. P. (2026). Pulmonary vein isolation using a novel balloon-in-basket pulsed field ablation system: The Lübeck how-to protocol. Heart Rhythm O2, 7(6), 1042-1052. https://doi.org/10.1016/j.hroo.2026.02.022
Vancouver
Tilz RR, Hatahet S, Eitel C, Popescu S, Subin B, Kuck KH, et al. Pulmonary vein isolation using a novel balloon-in-basket pulsed field ablation system: The Lübeck how-to protocol. Heart Rhythm O2. 2026;7(6):1042-1052. doi:10.1016/j.hroo.2026.02.022
BibTeX
@article{Tilz2026,
author = {Tilz, Roland Richard and Hatahet, Sascha and Eitel, Charlotte and Popescu, Sorin and Subin, Behnam and Kuck, Karl-Heinz and Mamaev, Roman and Wenzel, Jan-Per},
title = {Pulmonary vein isolation using a novel balloon-in-basket pulsed field ablation system: The Lübeck how-to protocol},
journal = {Heart Rhythm O2},
year = {2026},
volume = {7},
number = {6},
pages = {1042--1052},
doi = {10.1016/j.hroo.2026.02.022}
}
EUVIMED