Impact of heart rate on coronary artery stenosis grading accuracy using deep learning-based fast kV-switching CT: A phantom study

Ryoji Mikayama, Tsukasa Kojima, Takashi Shirasaka, Shiho Yamane, Ryohei Funatsu, Toyoyuki Kato, Hidetake Yabuuchi

Physica Medica · 2025

DOI: 10.1016/j.ejmp.2025.105046 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physica Medica, 2025. Verfasst von Ryoji Mikayama und 6 weiteren Autor:innen. Beteiligte Einrichtung: Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Fukuoka, Japan; Department of Health Sciences, Graduate School o…. Hauptthemen laut MEDLINE: Phantoms, Imaging, Deep Learning, Coronary Stenosis, Heart Rate. Die Arbeit nennt 30 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

INTRODUCTION: Deep learning-based fast kV-switching CT (DL-FKSCT) generates complete sinograms for fast kV-switching dual-energy CT (DECT) scans by using a trained neural network to restore missing views. Such restoration significantly enhances the image quality of coronary CT angiography (CCTA), and the allowable heart rate (HR) may vary between DECT and single-energy CT (SECT). This study aimed to examine HR's effect onCCTA using DL-FKSCT. METHODS: We scanned stenotic coronary artery phantoms attached to a pulsating cardiac phantom with DECT and SECT modes on a DL-FKSCT scanner. The phantom unit was operated with simulated HRs ranging from 0 (static) to 50-70 beats per minute (bpm). The sharpness and stenosis ratio of the coronary model were quantitatively compared between DECT and SECT, stratified by simulated HR settings using the paired t-test (significance was set at p < 0.01 with a Bonferroni adjustment for multiple comparisons). RESULTS: Regarding image sharpness, DECT showed significant superiority over SECT. In terms of the stenosis ratio compared to a static image reference, 70 keV virtual monochromatic image in DECT exhibited errors exceeding 10 % at HRs surpassing 65 bpm (p < 0.01), whereas 120 kVp SECT registered errors below 10 % across all HR settings, with no significant differences observed. CONCLUSIONS: In DL-FKSCT, DECT exhibited a lower upper limit of HR than SECT. Therefore, HR control is important for DECT scans in DL-FKSCT.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Phantoms, ImagingDeep LearningCoronary StenosisHeart RateTomography, X-Ray ComputedComputed Tomography AngiographyHumansImage Processing, Computer-AssistedCoronary AngiographyTime Factors

Stichwörter der Autor:innen

Cardiac computed tomography angiographyDual-energy computed tomographyHeart ratePhantom study

Autor:innen und Institutionen

  • Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Fukuoka, Japan; Department of Health Sciences, Graduate School of Medical Sciences, Kyushu University, Fukuoka, Japan. Electronic addr…
  • Department of Health Sciences, Faculty of Medical Sciences, Kyushu University, Fukuoka, Japan.
  • Division of Radiology, Department of Medical Technology, Kyushu University Hospital, Fukuoka, Japan.

Zitieren

APA

Mikayama, R., Kojima, T., Shirasaka, T., Yamane, S., Funatsu, R., Kato, T., & Yabuuchi, H. (2025). Impact of heart rate on coronary artery stenosis grading accuracy using deep learning-based fast kV-switching CT: A phantom study. Physica Medica. https://doi.org/10.1016/j.ejmp.2025.105046

Vancouver

Mikayama R, Kojima T, Shirasaka T, Yamane S, Funatsu R, Kato T, et al. Impact of heart rate on coronary artery stenosis grading accuracy using deep learning-based fast kV-switching CT: A phantom study. Physica Medica. 2025. doi:10.1016/j.ejmp.2025.105046

BibTeX
@article{Mikayama2025,
  author = {Mikayama, Ryoji and Kojima, Tsukasa and Shirasaka, Takashi and Yamane, Shiho and Funatsu, Ryohei and Kato, Toyoyuki and Yabuuchi, Hidetake},
  title = {Impact of heart rate on coronary artery stenosis grading accuracy using deep learning-based fast kV-switching CT: A phantom study},
  journal = {Physica Medica},
  year = {2025},
  doi = {10.1016/j.ejmp.2025.105046}
}
Kostenlos starten →