Dosimetric characterisation of commercial optically stimulated luminescence and tin-filtered radiophotoluminescence glass dosimeters for accurate radiation dose measurement in general radiography

Sumalee Yabsantia, Kosuke Matsubara, Chirasak Khamfongkhruea, Sawanee Suntiwong, Nuntawat Udee, Titipong Kaewlek, Darrin Madtohed, Nobpamadsiri Boonkrachang, Pratchayakan Hompeng, Thunyarat Chusin

Physica Medica · 2025

DOI: 10.1016/j.ejmp.2025.105030 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physica Medica, 2025. Verfasst von Sumalee Yabsantia und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiological Technology, Faculty of Allied Health Sciences, Naresuan University, Phitsanulok 65000, Thailand; Interdisciplinary Health …. Hauptthemen laut MEDLINE: Glass, Radiation Dosimeters, Radiation Dosage, Optically Stimulated Luminescence Dosimetry. Die Arbeit nennt 30 Literaturangaben. Gefördert von: Thailand Science Research and Innovation, Naresuan University.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Accurate radiation dose measurements rely on well-characterised dosimeter properties. This study investigated the characteristics of optically stimulated luminescence dosimeters (OSLD) and tin-filtered radiophotoluminescence glass dosimeters (RPLGD) in measuring incident air kerma (Ki), entrance surface air kerma (Ke), and absorbed dose (D) during general X-ray imaging. METHODS: We evaluated the radiation sensitivity and signal stability of OSLD and RPLGD. Their linear responses across doses from 0.5-7.0 mGy, energy dependence between 60 and 120 kVp, and variability in accumulated dose measurements were characterised using an ionisation chamber under Ki, Ke, and D measurement conditions. RESULTS: The sensitivity correction factors for OSLD and RPLGD ranged from 0.96 to 1.04 (CoV: 1.99 %) and 0.90-1.06 (CoV: 3.75 %), respectively. Signal depletion per reading was -0.40 ± 0.11 % for OSLD and -0.03 ± 0.04 % for RPLGD. Over 60 days, signal fading was -6.83 ± 0.40 % for OSLD and -2.70 ± 0.20 % for RPLGD. Both dosimeters showed excellent dose-response linearity (R2 > 0.995), with correction factors of 0.98-1.01 (CoV: 1.20 %) for OSLD and 1.01-1.05 (CoV: 1.95 %) for RPLGD. Beam quality correction factors ranged from 0.93 to 1.14 (CoV: 5.76 %) for OSLD and 0.87-1.07 (CoV: 5.36 %) for RPLGD, relative to 80 kVp calibration in Ki condition. Reproducibility of accumulated dose readings across nine fractions was within 2.32 % and 7.96 % for OSLD and RPLGD. CONCLUSIONS: OSLD demonstrated superior uniformity and greater reproducibility in accumulated measurements of Ki, Ke, and D, whereas RPLGD exhibited enhanced signal stability. Both dosimeters displayed distinct characteristics, underscoring the importance of applying correction factors for accurate radiation dose measurement.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

GlassRadiation DosimetersRadiation DosageOptically Stimulated Luminescence DosimetryTinRadiographyLuminescenceTitaniumRadiometry

Stichwörter der Autor:innen

General radiographyOptically stimulated luminescence dosimeterRadiation dose measurementRadiophotoluminescence glass dosimeter

Substanzen

TinTitanium

Autor:innen und Institutionen

  • Department of Radiological Technology, Faculty of Allied Health Sciences, Naresuan University, Phitsanulok 65000, Thailand; Interdisciplinary Health and Data Sciences Research Unit, Faculty of Allied Health Sciences, Na…
  • Department of Quantum Medical Technology, Faculty of Health Sciences, Institute of Medical, Pharmaceutical and Health Sciences, Kanazawa University, Ishikawa 920-0942, Japan.
  • Medical Physics Program, Princess Srisavangavadhana College of Medicine, Chulabhorn Royal Academy, Bangkok 10210, Thailand; Radiation Oncology Department, Chulabhorn Hospital, Chulabhorn Royal Academy, Bangkok 10210, Th…
  • Radiation Oncology Department, Chulabhorn Hospital, Chulabhorn Royal Academy, Bangkok 10210, Thailand.
  • Department of Radiological Technology, Faculty of Allied Health Sciences, Naresuan University, Phitsanulok 65000, Thailand.

Förderung

  • Thailand Science Research and Innovation
  • Naresuan University

Zitieren

APA

Yabsantia, S., Matsubara, K., Khamfongkhruea, C., Suntiwong, S., Udee, N., Kaewlek, T., Madtohed, D., Boonkrachang, N., Hompeng, P., & Chusin, T. (2025). Dosimetric characterisation of commercial optically stimulated luminescence and tin-filtered radiophotoluminescence glass dosimeters for accurate radiation dose measurement in general radiography. Physica Medica. https://doi.org/10.1016/j.ejmp.2025.105030

Vancouver

Yabsantia S, Matsubara K, Khamfongkhruea C, Suntiwong S, Udee N, Kaewlek T, et al. Dosimetric characterisation of commercial optically stimulated luminescence and tin-filtered radiophotoluminescence glass dosimeters for accurate radiation dose measurement in general radiography. Physica Medica. 2025. doi:10.1016/j.ejmp.2025.105030

BibTeX
@article{Yabsantia2025,
  author = {Yabsantia, Sumalee and Matsubara, Kosuke and Khamfongkhruea, Chirasak and Suntiwong, Sawanee and Udee, Nuntawat and Kaewlek, Titipong and Madtohed, Darrin and Boonkrachang, Nobpamadsiri and Hompeng, Pratchayakan and Chusin, Thunyarat},
  title = {Dosimetric characterisation of commercial optically stimulated luminescence and tin-filtered radiophotoluminescence glass dosimeters for accurate radiation dose measurement in general radiography},
  journal = {Physica Medica},
  year = {2025},
  doi = {10.1016/j.ejmp.2025.105030}
}
Kostenlos starten →