Patient-specific organ dose and in-vivo image quality assessment in clinical CT

Wanyi Fu, Shobhit Sharma, Justin Solomon, Francesco Ria, Hananiel Setiawan, Aiping Ding, William P Segars, Ehsan Samei

Physica Medica · 2025-08 · 136 · 105017

DOI: 10.1016/j.ejmp.2025.105017 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physica Medica, 2025-08. Verfasst von Wanyi Fu und 7 weiteren Autor:innen. Beteiligte Einrichtung: Center for Virtual Imaging Trials, Carl E. Ravin Advanced Imaging Labs, and Department of Electrical and Computer Engineering, Duke University, Durha…. Hauptthemen laut MEDLINE: Radiation Dosage, Tomography, X-Ray Computed. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 31 Literaturangaben. Gefördert von: National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: To develop and characterize individualized dose and quality measures at organ level compared to their generic counterparts across a clinical CT dataset. MATERIALS AND METHODS: The study included 9801 chest-abdomen-pelvis and abdomen-pelvis CT exams (7,763 patients, mean age, 56 ± 17 years; 4113 women) representing 20 unique protocols. For each exam, patient-specific organ dose of all radiosensitive organs was estimated using a validated method by generating personalized computational phantoms and Monte Carlo simulations. Effective dose (EOD) was calculated by weighted sum of the organ doses. Liver dose, ODliver, noise in the liver, Nliver, and observer model detectability, d', were assessed within the liver as examples of individualized, organ-based image assessment measurements. The organ-based measurements (ODliver, EOD, and Nliver) were compared to their generic counterparts: ssize-specific ddose estimates (SSDE), effective dose based on dose length product (EDLP), and whole-body noise (Nglobal), respectively. RESULTS: Generic dose values were substantially higher than individualized estimates for SSDE vs. ODliver (median of all exams: 51.2 %, p < 0.001) and EDLP vs. EDOD (median: 41.0 %, p < 0.001). Nglobal was generally lower than Nliver (median: -7.2 %, p < 0.001). The correlation relationships of EOD and d' were substantially varied (R2 range: 0-0.5) for different patient sizes and scan parameters. CONCLUSIONS: Demonstrated across a population of exams, individualized organ-based measurements of dose and quality are feasible. Generic measures cannot fully represent individualized organ-based values. The correlation relationships between individualized dose and image quality values varies for different vendors and protocols, implying imaging optimization is best when done semi-independently for each factor using individualized measurements.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiation DosageTomography, X-Ray ComputedHumansFemaleMiddle AgedMaleMonte Carlo MethodPhantoms, ImagingAdultAgedImage Processing, Computer-AssistedQuality Control

Stichwörter der Autor:innen

Computed tomographyMonte CarloQuality controlcomputational phantomorgan dose

Autor:innen und Institutionen

  • Center for Virtual Imaging Trials, Carl E. Ravin Advanced Imaging Labs, and Department of Electrical and Computer Engineering, Duke University, Durham, NC 27708, USA. Electronic address: wanyifu@tsinghua.edu.cn.
  • Center for Virtual Imaging Trials, Carl E. Ravin Advanced Imaging Labs, and Medical Physics Graduate Program Duke University, Durham, NC 27708, USA. Electronic address: shsharma@mru.medical.canon.
  • Center for Virtual Imaging Trials, Carl E. Ravin Advanced Imaging Labs, Clinical Imaging Physics Group, Duke University Health System, Durham, NC 27710, USA. Electronic address: justin.solomon@duke.edu.
  • Center for Virtual Imaging Trials, Carl E. Ravin Advanced Imaging Labs, Clinical Imaging Physics Group, Duke University Health System, Durham, NC 27710, USA. Electronic address: francesco.ria@duke.edu.
  • Center for Virtual Imaging Trials, Carl E. Ravin Advanced Imaging, Medical Physics Graduate Program, Duke University, Durham, NC 27708, USA. Electronic address: hananiel.setiawan@duke.edu.
  • Clinical Imaging Physics Group, Duke University Health System, Durham, NC 27710, USA. Electronic address: aiping.ding@duke.edu.

Förderung

  • National Institutes of Health

Zitieren

APA

Fu, W., Sharma, S., Solomon, J., Ria, F., Setiawan, H., Ding, A., Segars, W. P., & Samei, E. (2025). Patient-specific organ dose and in-vivo image quality assessment in clinical CT. Physica Medica, 136, 105017. https://doi.org/10.1016/j.ejmp.2025.105017

Vancouver

Fu W, Sharma S, Solomon J, Ria F, Setiawan H, Ding A, et al. Patient-specific organ dose and in-vivo image quality assessment in clinical CT. Physica Medica. 2025;136:105017. doi:10.1016/j.ejmp.2025.105017

BibTeX
@article{Fu2025,
  author = {Fu, Wanyi and Sharma, Shobhit and Solomon, Justin and Ria, Francesco and Setiawan, Hananiel and Ding, Aiping and Segars, William P and Samei, Ehsan},
  title = {Patient-specific organ dose and in-vivo image quality assessment in clinical CT},
  journal = {Physica Medica},
  year = {2025},
  volume = {136},
  pages = {105017},
  doi = {10.1016/j.ejmp.2025.105017}
}
Kostenlos starten →