EUVIMEDEUVIMED entdecken
VergleichsstudieMultizentrische StudieRandomisierte kontrollierte Studie

Impact of Flow Status on Effective Orifice Area in Transcatheter Heart Valves: A COMPARE-TAVI 1 Substudy

Lytfi Krasniqi, Christian Juhl Terkelsen, Henrik Nissen, Philip Freeman, Christian Alcaraz Frederiksen, Henrik Vase, Kristian Hejlesen, Christian Byskov Fur, Troels Thim, Ashkan Eftekhari, Frederik Uttenthal, Julia Ellert und 4 weitere

Journal of the American Society of Echocardiography · 2026

DOI: 10.1016/j.echo.2025.09.017 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Journal of the American Society of Echocardiography, 2026. Verfasst von Lytfi Krasniqi und 15 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiothoracic Surgery, Odense University Hospital, Odense, Denmark; Department of Clinical Research, University of Southern Denmark, O…. Hauptthemen laut MEDLINE: Transcatheter Aortic Valve Replacement, Aortic Valve Stenosis, Heart Valve Prosthesis, Echocardiography, Doppler. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 23 Literaturangaben. Gefördert von: Hjerteforeningen, Region Midtjylland, Meril Life Sciences.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE: The aim of this study was to study the impact of flow status on effective orifice area (EOA) in patients treated with the balloon-expandable Myval and Sapien transcatheter heart valves (THVs). METHODS: We collected the core laboratory-measured EOA, mean and peak gradients, Doppler velocity index (DVI), and stroke volume index (SVi) from the 30-day echocardiograms in patients treated with the balloon-expandable Myval and Sapien THVs in the COMPARE-TAVI 1 trial. Patients were stratified according to flow into low flow (SVi 50 mL/m2). RESULTS: A total of 1,031 patients were included in COMPARE-TAVI 1. Myval THVs (20.0-32.0 mm) exhibited overall higher mean EOA and lower mean gradients (1.99 ± 0.55 cm2, 9.46 ± 3.84 mm Hg) compared to Sapien THVs (20-29 mm; 1.81 ± 0.56 cm2, 11.47 ± 4.47 mm Hg). Compared to normal-flow patients, low-flow patients exhibited smaller EOA (P < .001) and lower DVI (P < .001) in all valve sizes irrespective of THV platform, while high-flow patients had larger EOA (P < .001) and higher DVI (P < .001). Female sex was associated with lower stroke volume, resulting in lower mean and peak gradients independent of valve size and platform (P < .001). CONCLUSION: The study indicates that EOA may be underestimated after TAVI in patients with low-flow status, which can impact the assessment of prosthesis-patient mismatch.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Transcatheter Aortic Valve ReplacementAortic Valve StenosisHeart Valve ProsthesisEchocardiography, DopplerAortic ValveHumansFemaleMaleProsthesis DesignAged, 80 and overBlood Flow VelocityAgedTreatment OutcomeStroke Volume

Stichwörter der Autor:innen

COMPARE-TAVI 1MyvalRCTSapienTAVI

Autor:innen und Institutionen

  • Department of Cardiothoracic Surgery, Odense University Hospital, Odense, Denmark; Department of Clinical Research, University of Southern Denmark, Odense, Denmark.
  • Department of Cardiology, Aarhus University Hospital, Aarhus, Denmark; Department of Clinical Medicine, Aarhus University Hospital, Aarhus, Denmark.
  • Department of Cardiology, Odense University Hospital, Odense, Denmark.
  • Department of Cardiology, Aalborg University Hospital, Aalborg, Denmark.
  • Department of Cardiology, Aarhus University Hospital, Aarhus, Denmark.
  • Department of Cardiology, Odense University Hospital, Odense, Denmark; Department of Cardiovascular Medicine, Mayo Clinic, Rochester, Minnesota; Department of Clinical Research, University of Southern Denmark, Odense, D…

Förderung

  • Hjerteforeningen
  • Region Midtjylland
  • Meril Life Sciences
  • Odense University Hospital

Zitieren

APA

Krasniqi, L., Terkelsen, C. J., Nissen, H., Freeman, P., Frederiksen, C. A., Vase, H., Hejlesen, K., Fur, C. B., Thim, T., Eftekhari, A., Uttenthal, F., Ellert, J., Borg Mogensen, N. S., Haujir, A., Christiansen, E. H., & Dahl, J. S. (2026). Impact of Flow Status on Effective Orifice Area in Transcatheter Heart Valves: A COMPARE-TAVI 1 Substudy. Journal of the American Society of Echocardiography. https://doi.org/10.1016/j.echo.2025.09.017

Vancouver

Krasniqi L, Terkelsen CJ, Nissen H, Freeman P, Frederiksen CA, Vase H, et al. Impact of Flow Status on Effective Orifice Area in Transcatheter Heart Valves: A COMPARE-TAVI 1 Substudy. Journal of the American Society of Echocardiography. 2026. doi:10.1016/j.echo.2025.09.017

BibTeX
@article{Krasniqi2026,
  author = {Krasniqi, Lytfi and Terkelsen, Christian Juhl and Nissen, Henrik and Freeman, Philip and Frederiksen, Christian Alcaraz and Vase, Henrik and Hejlesen, Kristian and Fur, Christian Byskov and Thim, Troels and Eftekhari, Ashkan and Uttenthal, Frederik and Ellert, Julia and Borg Mogensen, Nils Sofus and Haujir, Amal and Christiansen, Evald Høj and Dahl, Jordi Sanchez},
  title = {Impact of Flow Status on Effective Orifice Area in Transcatheter Heart Valves: A COMPARE-TAVI 1 Substudy},
  journal = {Journal of the American Society of Echocardiography},
  year = {2026},
  doi = {10.1016/j.echo.2025.09.017}
}
Kostenlos starten →