Comparison of spectral performance of three dual-energy CT scanners equipped with a deep-learning image reconstruction algorithm and one photon counting CT scanner: A phantom study

Joël Greffier, Cécile Salvat, Maxime Pastor, Valérie Bousson, Salim Si-Mohamed, Jean-Paul Beregi, Lama Hadid-Beurrier, Djamel Dabli

Diagnostic and Interventional Imaging · 2026

DOI: 10.1016/j.diii.2025.11.007 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Diagnostic and Interventional Imaging, 2026. Verfasst von Joël Greffier und 7 weiteren Autor:innen. Beteiligte Einrichtung: IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30900 Nîmes, France. Hauptthemen laut MEDLINE: Algorithms, Tomography Scanners, X-Ray Computed, Deep Learning, Tomography, X-Ray Computed. Die Arbeit wurde bisher 7-mal zitiert (Crossref) und nennt 31 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: The purpose of this study was to compare the spectral performance of three dual-energy CT (DECT) scanners and one photon-counting CT (PCCT) scanner on virtual monoenergetic images (VMIs) at low-energy levels and on iodine maps. MATERIALS AND METHODS: A spectral phantom was scanned using one PCCT scanner and three different DECT scanners that included a rapid kV-switching CT (R-KVSCT), an ultrafast kV-switching (U-KVSCT) and a dual-layer CT (DLCT) scanner. Acquisitions were obtained with each CT system using classical abdominal and pelvic examination parameters, as well as a volume CT dose index at 11 mGy. VMI at 40/50/60/70 keV and iodine maps were reconstructed for each scanner. Noise power spectrum (NPS) and task-based transfer function (TTF) were evaluated. Detectability indexes (d') were computed to model the detection task of two contrast-enhanced lesions. RESULTS: Noise magnitude decreased from 40 to 70 keV for all DECT scanners and this decrease was greater for R-KVSCT (-80.0 ± 0.1 [standard deviation (SD)] %) and less pronounced for DLCT (-14.4 ± 0.8 [SD] %) scanners. The average NPS spatial frequency (fav) values decreased from 40 to 70 keV (0.26 to 0.17 mm-1) for R-KVSCT, increased for DLCT (0.18 to 0.25 mm-1) but were similar for U-KVSCT (0.19 ± 0.002 [SD] mm-1) and PCCT (0.21 ± [SD] 0.008 mm-1) scanner. For R-KVSCT and PCCT scanners, TTF at 50 % (f50) values increased from 40 to 70 keV for both inserts. For U-KVSCT and DLCT scanners, similar f50 values were found according to energy level for both inserts. For both contrast-enhanced lesions, d' values decreased from 40 to 70 keV for PCCT, DLCT and U-KVSCT scanners. For R-KVSCT scanner, d' values peaked at 60 keV. At 40 and 50 keV, the greatest d' values were found with DLCT and PCCT scanners. CONCLUSION: At 40 or 50 keV, the best combined results (objective and subjective assessments) are obtained with DLCT and PCCT scanners.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AlgorithmsTomography Scanners, X-Ray ComputedDeep LearningTomography, X-Ray ComputedImage Processing, Computer-AssistedRadiography, Dual-Energy Scanned ProjectionPhantoms, ImagingPhotonsHumans

Stichwörter der Autor:innen

Computed tomographyDual-energy CTImage qualityPhoton-counting CTTask-based image quality assessment

Autor:innen und Institutionen

  • IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30900 Nîmes, France.
  • Radiation Protection and Medical Physics Department, Hôpital Lariboisière, AP-HP, Université Paris Cité, 75010 Paris, France.
  • Department of Radiology, Hôpital Lariboisière-Fernand Widal, AP-HP, Université Paris Cité, 75010 Paris, France.
  • Université de Lyon, INSA-Lyon, Université Claude Bernard Lyon 1, UJM-Saint Etienne, CNRS, INSERM, CREATIS UMR 5220, U1206, 69100 Villeurbanne, France; Department of Radiology, Hôpital Louis Pradel, Hospices Civils de Ly…

Zitieren

APA

Greffier, J., Salvat, C., Pastor, M., Bousson, V., Si-Mohamed, S., Beregi, J. P., Hadid-Beurrier, L., & Dabli, D. (2026). Comparison of spectral performance of three dual-energy CT scanners equipped with a deep-learning image reconstruction algorithm and one photon counting CT scanner: A phantom study. Diagnostic and Interventional Imaging. https://doi.org/10.1016/j.diii.2025.11.007

Vancouver

Greffier J, Salvat C, Pastor M, Bousson V, Si-Mohamed S, Beregi JP, et al. Comparison of spectral performance of three dual-energy CT scanners equipped with a deep-learning image reconstruction algorithm and one photon counting CT scanner: A phantom study. Diagnostic and Interventional Imaging. 2026. doi:10.1016/j.diii.2025.11.007

BibTeX
@article{Greffier2026,
  author = {Greffier, Joël and Salvat, Cécile and Pastor, Maxime and Bousson, Valérie and Si-Mohamed, Salim and Beregi, Jean-Paul and Hadid-Beurrier, Lama and Dabli, Djamel},
  title = {Comparison of spectral performance of three dual-energy CT scanners equipped with a deep-learning image reconstruction algorithm and one photon counting CT scanner: A phantom study},
  journal = {Diagnostic and Interventional Imaging},
  year = {2026},
  doi = {10.1016/j.diii.2025.11.007}
}
Kostenlos starten →