User evaluation of detector performance in clinical photon‐counting and energy‐integrating CT scanners using DICOM images
Medical Physics · 2025
DOI: 10.1002/mp.70045 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Medical Physics, 2025. Verfasst von Ke Li und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Imaging Physics, UT MD Anderson Cancer Center, Houston, Texas, USA. Hauptthemen laut MEDLINE: Photons, Image Processing, Computer-Assisted, Tomography Scanners, X-Ray Computed, Tomography, X-Ray Computed. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Background
Clinical users have a critical need to routinely assess the performance of photon‐counting detectors (PCDs) in PCD‐CT scanners. Such assessments provide insights into detector characteristics, support protocol optimization, and inform decisions on future scanner acquisitions. Historically, this has been challenging due to limited access to raw detector data, which restricts direct evaluation of PCD performance.
Purpose
To evaluate the zero‐frequency detective quantum efficiency () and detector deadtime of PCDs from an end‐user perspective using reconstructed DICOM images.
Methods
Detector performance was evaluated on two Siemens NAEOTOM Alpha PCD‐CT scanners and one Siemens SOMATOM Force energy‐integrating detector CT (EID‐CT) scanner. Air‐only scans were performed in service mode across a range of tube potentials (70–140 kV) and tube currents (4–1200 mA). DICOM images were reconstructed on the scanner using a linear algorithm with a soft‐tissue kernel (Br44). The noise power spectrum (NPS) of the images was used to estimate the mean detector output counts. Mean input photon numbers were estimated based on beam quality and exposure measurements. For the PCD‐CT systems, tube current‐sweep experiments were used to generate image variance—mA curves, from which detector deadtime was estimated using a previously validated parametric model.
Results
The EID and PCD demonstrated comparable values (EID: 72%–74%; PCD: 72%–77%). showed no significant dependence on tube potential. The estimated PCD deadtime ranged from 5.3 to 7.0 ns. Detector performance was consistent across both PCD‐CT systems. Additionally, the ultra‐high‐resolution (UHR) and standard acquisition modes exhibited equivalent deadtime.
Conclusions
The of the PCD‐CT and EID‐CT detectors are comparable, with performance primarily limited by geometric efficiency rather than sensor absorption efficiency. Under clinically relevant conditions, pileup‐induced count losses in the evaluated PCD‐CT scanners are minimal and appear to be effectively corrected by the manufacturer.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Imaging Physics, UT MD Anderson Cancer Center, Houston, Texas, USA.
- Department of Interventional Radiology, UT MD Anderson Cancer Center, Houston, Texas, USA.
- Department of Abdominal Imaging, UT MD Anderson Cancer Center, Houston, Texas, USA.
Zitieren
APA
Li, K., Liu, X., Jacobsen, M. C., Rong, J., Jensen, C. T., Tamm, E. P., & Dong, F. (2025). User evaluation of detector performance in clinical photon‐counting and energy‐integrating CT scanners using DICOM images. Medical Physics. https://doi.org/10.1002/mp.70045
Vancouver
Li K, Liu X, Jacobsen MC, Rong J, Jensen CT, Tamm EP, et al. User evaluation of detector performance in clinical photon‐counting and energy‐integrating CT scanners using DICOM images. Medical Physics. 2025. doi:10.1002/mp.70045
BibTeX
@article{Li2025,
author = {Li, Ke and Liu, Xinming and Jacobsen, Megan C and Rong, John and Jensen, Corey T and Tamm, Eric P and Dong, Frank},
title = {User evaluation of detector performance in clinical photon‐counting and energy‐integrating CT scanners using DICOM images},
journal = {Medical Physics},
year = {2025},
doi = {10.1002/mp.70045}
}
EUVIMED