Impact of tin filter on the image quality of ultra-low dose chest CT: A phantom study on three CT systems

Joël Greffier, Isabelle Fitton, Claire Van Ngoc Ty, Julien Frandon, Jean-Paul Beregi, Djamel Dabli

Diagnostic and Interventional Imaging · 2023

DOI: 10.1016/j.diii.2023.05.005 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Diagnostic and Interventional Imaging, 2023. Verfasst von Joël Greffier und 5 weiteren Autor:innen. Beteiligte Einrichtung: IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30029 Nîmes, France. Electronic address: joel.gr…. Hauptthemen laut MEDLINE: Tin, Tomography, X-Ray Computed. Die Arbeit wurde bisher 19-mal zitiert (Crossref) und nennt 31 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: The purpose of this study was to assess the impact of a tin filter on the image quality of ultra-low dose (ULD) chest computed tomography (CT) on three different CT systems. MATERIALS AND METHODS: An image quality phantom was scanned on three CT systems including two split-filter dual-energy CT (SFCT-1 and SFCT-2) scanners and one dual-source CT scanner (DSCT). Acquisitions were performed with a volume CT dose index (CTDIvol) of 0.4 mGy, first at 100 kVp without tin filter (Sn), and second, at Sn100/Sn140 kVp, Sn100/Sn110/Sn120/Sn130/Sn140/Sn150 kVp and Sn100/Sn150 kVp for SFCT-1, SFCT-2 and DSCT respectively. Noise-power-spectrum and task-based transfer function were computed. The detectability index (d') was computed to model the detection of two chest lesions. RESULTS: For DSCT and SFCT-1, noise magnitude values were higher with 100kVp than with Sn100 kVp and with Sn140 kVp or Sn150 kVp than with Sn100 kVp. For SFCT-2, noise magnitude increased from Sn110 kVp to Sn150 kVp and was higher at Sn100 kVp than at Sn110 kVp. For most kVp with the tin filter, the noise amplitude values were lower than those obtained at 100 kVp. For each CT system, noise texture and spatial resolution values were similar with 100 kVp and with all kVp used with a tin filter. For all simulated chest lesions, the highest d' values were obtained at Sn100 kVp for SFCT-1 and DSCT and at Sn110 kVp for SFCT-2. CONCLUSION: For ULD chest CT protocols, the lowest noise magnitude and highest detectability values for simulated chest lesions are obtained with Sn100 kVp for the SFCT-1 and DSCT CT systems and at Sn110 kVp for SFCT-2.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

TinTomography, X-Ray ComputedHumansRadiation DosageThoraxTomography Scanners, X-Ray ComputedPhantoms, Imaging

Stichwörter der Autor:innen

ChestMultidetector computed tomographyTask-based image quality assessmentTin filterUltra-low dose

Substanzen

Tin

Autor:innen und Institutionen

  • IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30029 Nîmes, France. Electronic address: joel.greffier@chu-nimes.fr.
  • Université Paris Cité, 75006 Paris, France, Department of Radiology, Assistance Publique-Hôpitaux de Paris, Hôpital Européen Georges Pompidou, 75015 Paris, France.
  • IMAGINE UR UM 103, Montpellier University, Department of Medical Imaging, Nîmes University Hospital, 30029 Nîmes, France.

Zitieren

APA

Greffier, J., Fitton, I., Ngoc Ty, C. V., Frandon, J., Beregi, J. P., & Dabli, D. (2023). Impact of tin filter on the image quality of ultra-low dose chest CT: A phantom study on three CT systems. Diagnostic and Interventional Imaging. https://doi.org/10.1016/j.diii.2023.05.005

Vancouver

Greffier J, Fitton I, Ngoc Ty CV, Frandon J, Beregi JP, Dabli D. Impact of tin filter on the image quality of ultra-low dose chest CT: A phantom study on three CT systems. Diagnostic and Interventional Imaging. 2023. doi:10.1016/j.diii.2023.05.005

BibTeX
@article{Greffier2023,
  author = {Greffier, Joël and Fitton, Isabelle and Ngoc Ty, Claire Van and Frandon, Julien and Beregi, Jean-Paul and Dabli, Djamel},
  title = {Impact of tin filter on the image quality of ultra-low dose chest CT: A phantom study on three CT systems},
  journal = {Diagnostic and Interventional Imaging},
  year = {2023},
  doi = {10.1016/j.diii.2023.05.005}
}
Kostenlos starten →