Fachartikel in der Zeitschrift PLOS ONE, 2023. Verfasst von Simone Schüle und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic and Interventional Radiology and Neuroradiology, German Armed Forces Hospital Ulm, Ulm, Baden-Wurttemberg, Germany. Hauptthemen laut MEDLINE: Tin, Sinusitis. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objectives
Paranasal sinus imaging due to chronic inflammatory disease is one of the most common examinations in head and neck radiology with CT imaging considered the current gold standard. In this phantom study we analyzed different low dose CT protocols in terms of image quality, radiation exposure and subjective evaluation in order to establish an optimized scanning protocol.
Methods
In a phantom study, an Alderson phantom was scanned using 12 protocols between 70–120 kV and 25–200 mAs with and without tin filtration. For all datasets, iterative reconstruction was used. Data were objectively evaluated (image noise, (dose-weighted) contrast-to-noise ratio) and for subjective evaluation an online survey using a Likert scale was performed to reach a large group of clinically experienced reader (n = 62). The protocol was considered diagnostically insufficient if the median score was 4 and above and if more than 10% of raters scored 4 and above on the Likert scale. For an interreader agreement an ICC was calculated. To compare clinical value in relation to the applied dose and the objective image parameters, we calculated a figure of merit (FOM) and ranked the protocols accordingly.
Results
There was an overall moderate agreement between the 62 readers for the 12 examined CT protocols. In this phantom study, protocols with 100 kV with spectral shaping and 50–100 mAs obtained the best results for its combination of dose, image quality and clinical information value for diagnosing sinusitis (FOM 1 st – 2 nd place) with the 70 kV and 50 mAs as a good alternative as well (Sinusitis: FOM shared 2 nd ). For preoperative planning, where a higher dose is necessary, 100 kV with spectral shaping and 100 mAs achieved the overall best results (FOM 1 st place) with 70 kV and 50 mAs ranking 4 th .
Conclusion
100-kV protocols with spectral shaping or low kV protocols (70 kV) with a similarly low dose showed the best figure of merit for imaging sinonasal disease and preoperative planning. With modern scanner technology available, spectral shaping or low KV protocols should be used for sinusitis imaging.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Substanzen
Autor:innen und Institutionen
- Department of Diagnostic and Interventional Radiology and Neuroradiology, German Armed Forces Hospital Ulm, Ulm, Baden-Wurttemberg, Germany.
- Department of Otorhinolaryngology and Head and Neck Surgery, German Armed Forces Central Hospital Koblenz, Koblenz, Rhineland-Palatinate, Germany.
- Department of Radiology, University Hospital of Ulm, Ulm, Baden-Wurttemberg, Germany.
ORCID-Profile: Simone Schüle, Kai Johannes Lorenz, Carsten Hackenbroch
Zitieren
APA
Schüle, S., Strobel, J. R. B., Lorenz, K. J., Beer, M., & Hackenbroch, C. (2023). Tin filter compared to low kV protocols - optimizing sinonasal imaging in computed tomography. PLOS ONE. https://doi.org/10.1371/journal.pone.0279907
Vancouver
Schüle S, Strobel JRB, Lorenz KJ, Beer M, Hackenbroch C. Tin filter compared to low kV protocols - optimizing sinonasal imaging in computed tomography. PLOS ONE. 2023. doi:10.1371/journal.pone.0279907
BibTeX
@article{Schle2023,
author = {Schüle, Simone and Strobel, Joachim Rudolf Balthasar and Lorenz, Kai Johannes and Beer, Meinrad and Hackenbroch, Carsten},
title = {Tin filter compared to low kV protocols - optimizing sinonasal imaging in computed tomography},
journal = {PLOS ONE},
year = {2023},
doi = {10.1371/journal.pone.0279907}
}
EUVIMED