Ultra-high throughput dual channel liquid chromatography with tandem mass spectrometry for quantification of four immunosuppressants in whole blood for therapeutic drug monitoring

Tanja R Zijp, Kai van Hateren, Hiltjo Kuiper, Erwin M Jongedijk, Daan J Touw

Journal of Chromatography A · 2023

DOI: 10.1016/j.chroma.2023.464086 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Journal of Chromatography A, 2023. Verfasst von Tanja R Zijp und 4 weiteren Autor:innen. Beteiligte Einrichtung: University of Groningen, University Medical Center Groningen, Department of Clinical Pharmacy and Pharmacology, Hanzeplein 1, 9700 RB Groningen, the …. Hauptthemen laut MEDLINE: Immunosuppressive Agents, Tandem Mass Spectrometry. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Liquid chromatography with tandem mass spectrometry (LC-MS/MS) is the golden standard for immunosuppressants analyses, where optimising throughput by parallel chromatography can reduce costs and turnaround time. We aimed to double our system throughput using a dual LC-MS/MS setup. Therefore, two independent UPLC systems were hyphenated to one triple quadrupole MS, with staggered injections from one autosampler on alternating columns. The method simultaneously measured the analytes tacrolimus, sirolimus, everolimus, and cyclosporin A in whole blood using isotope dilution, with a run time of 1.5 min. Using the dual LC-MS/MS system, net run-to-run time improved from 2.3 to 0.98 min per injection, where throughput increased from 26 to 61 injections per hour. For Performance Qualification, 1101 clinical samples were measured on the dual LC-MS/MS system in addition to the standard system, during a period of one month, and the results were compared using Passing Bablok regression and Bland Altman analysis. There was excellent agreement for all four analytes, with regression slopes of 0.98-1.02x and intercepts of -0.11-0.88 µg/L. Minor bias was demonstrated between the systems with mean differences from -0.93 to 1.43%. In conclusion, the throughput was doubled and idle MS time was reduced with good agreement to the standard system. Currently, the method is applied for clinical routine with frequent peak intensities of >180 injections per day.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Immunosuppressive AgentsTandem Mass SpectrometryChromatography, LiquidDrug MonitoringCyclosporine

Stichwörter der Autor:innen

Dual LC MS MSDual channel chromatographyOptimal MS MS utilisationProductivity optimisationUltrahigh-speed

Substanzen

Immunosuppressive AgentsCyclosporine

Autor:innen und Institutionen

  • University of Groningen, University Medical Center Groningen, Department of Clinical Pharmacy and Pharmacology, Hanzeplein 1, 9700 RB Groningen, the Netherlands.
  • University of Groningen, University Medical Center Groningen, Department of Clinical Pharmacy and Pharmacology, Hanzeplein 1, 9700 RB Groningen, the Netherlands; University of Groningen, Department of Pharmaceutical Ana…

Zitieren

APA

Zijp, T. R., van Hateren, K., Kuiper, H., Jongedijk, E. M., & Touw, D. J. (2023). Ultra-high throughput dual channel liquid chromatography with tandem mass spectrometry for quantification of four immunosuppressants in whole blood for therapeutic drug monitoring. Journal of Chromatography A. https://doi.org/10.1016/j.chroma.2023.464086

Vancouver

Zijp TR, van Hateren K, Kuiper H, Jongedijk EM, Touw DJ. Ultra-high throughput dual channel liquid chromatography with tandem mass spectrometry for quantification of four immunosuppressants in whole blood for therapeutic drug monitoring. Journal of Chromatography A. 2023. doi:10.1016/j.chroma.2023.464086

BibTeX
@article{Zijp2023,
  author = {Zijp, Tanja R and van Hateren, Kai and Kuiper, Hiltjo and Jongedijk, Erwin M and Touw, Daan J},
  title = {Ultra-high throughput dual channel liquid chromatography with tandem mass spectrometry for quantification of four immunosuppressants in whole blood for therapeutic drug monitoring},
  journal = {Journal of Chromatography A},
  year = {2023},
  doi = {10.1016/j.chroma.2023.464086}
}
Kostenlos starten →