Preliminary analytical evaluation of the fully automated Cobas i601 LC-MS/MS system for steroid hormones and vitamin D in a clinical laboratory
Clinica Chimica Acta · 2027-02 · 593 · 121312
DOI: 10.1016/j.cca.2026.121312 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Clinica Chimica Acta, 2027-02. Verfasst von Mariana Serres-Gómez und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Laboratory Medicine, La Paz University Hospital, Madrid 28046, Spain; IdiPaz-Hospital La Paz Institute for Health Research, Madrid 2804…. Hauptthemen laut MEDLINE: Tandem Mass Spectrometry, Vitamin D, Liquid Chromatography-Mass Spectrometry, Steroids. Die Arbeit nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Roche Diagnostics Corporation.
Abstract (Originalsprache)
BACKGROUND: Fully automated liquid chromatography-tandem mass spectrometry (LC-MS/MS) platforms aim to combine the analytical specificity and sensitivity of mass spectrometry with the workflow efficiency of automated routine analyzers. This study performed a preliminary analytical evaluation of the Cobas i601 LC-MS/MS system for estradiol, testosterone, aldosterone, androstenedione, 25-hydroxyvitamin D (25(OH)D) and 24,25-dihydroxyvitamin D (24,25(OH)2D), including method comparison with routine immunoassays in a real-world clinical laboratory setting. METHODS: Imprecision, bias, and linearity were assessed according to Clinical and Laboratory Standards Institute (CLSI) EP10-A3 using three levels of commercial quality control material. Results were compared to analytical performance specifications (APS) based on biological variation. Method comparison and bias estimation with routine immunoassays were performed following CLSI EP09-A3 using serum samples. Predicted biases at clinical decision levels (CDL) were evaluated against allowable limits. RESULTS: All assays demonstrated acceptable analytical performance (maximum coefficient of variation 3.4%; maximum bias 7.8%) and fulfilled predefined APS. Linearity was confirmed across the tested ranges (p > 0.05). Testosterone showed acceptable biases at CDL (1.7% to 9.3%) according to the predefined APS, although greater variability was observed at lower concentrations. In contrast, androstenedione, estradiol, aldosterone and 25(OH)D exhibited substantial systematic bias compared with immunoassays (20%, -36.5%, -28.5% and - 16% respectively at intermediate CDL). CONCLUSIONS: The Cobas i601 demonstrated acceptable analytical performance in this first preliminary evaluation. However, systematic differences relative to immunoassays for several analytes highlight the need for close clinician-laboratory communication when implementing this new automated LC-MS/MS platform, and underscore the need for further analytical and clinical validation.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Laboratory Medicine, La Paz University Hospital, Madrid 28046, Spain; IdiPaz-Hospital La Paz Institute for Health Research, Madrid 28046, Spain. Electronic address: mserres.gomez@gmail.com.
- Department of Laboratory Medicine, La Paz University Hospital, Madrid 28046, Spain; IdiPaz-Hospital La Paz Institute for Health Research, Madrid 28046, Spain.
- Department of Laboratory Medicine, La Paz University Hospital, Madrid 28046, Spain.
Förderung
- Roche Diagnostics Corporation
Zitieren
APA
Serres-Gómez, M., González-Casaus, M. L., Serrano Olmedo, M. G., Fernández-Calle, P., Villar Torrubiano, D., Iturzaeta Sánchez, J. M., González Almodóvar, B., & Buño Soto, A. (2027). Preliminary analytical evaluation of the fully automated Cobas i601 LC-MS/MS system for steroid hormones and vitamin D in a clinical laboratory. Clinica Chimica Acta, 593, 121312. https://doi.org/10.1016/j.cca.2026.121312
Vancouver
Serres-Gómez M, González-Casaus ML, Serrano Olmedo MG, Fernández-Calle P, Villar Torrubiano D, Iturzaeta Sánchez JM, et al. Preliminary analytical evaluation of the fully automated Cobas i601 LC-MS/MS system for steroid hormones and vitamin D in a clinical laboratory. Clinica Chimica Acta. 2027;593:121312. doi:10.1016/j.cca.2026.121312
BibTeX
@article{SerresGmez2027,
author = {Serres-Gómez, Mariana and González-Casaus, María Luisa and Serrano Olmedo, María Gemma and Fernández-Calle, Pilar and Villar Torrubiano, Daniel and Iturzaeta Sánchez, José Manuel and González Almodóvar, Bernardo and Buño Soto, Antonio},
title = {Preliminary analytical evaluation of the fully automated Cobas i601 LC-MS/MS system for steroid hormones and vitamin D in a clinical laboratory},
journal = {Clinica Chimica Acta},
year = {2027},
volume = {593},
pages = {121312},
doi = {10.1016/j.cca.2026.121312}
}
EUVIMED