Evaluation of an enzymatic one-step assay for Beta-Hydroxybutyrate blood testing on clinical chemistry platforms using HILIC-MS/MS as a non-enzymatic reference method

Max K Steinbach, Holger Müller, Yanik Reichard, Konstantin Eulberg, Tobias Frömel, Alexandra Lein, Christoph Peter, Kai Prager, Matthias Grimmler

Clinica Chimica Acta · 2026

DOI: 10.1016/j.cca.2025.120703 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Clinica Chimica Acta, 2026. Verfasst von Max K Steinbach und 8 weiteren Autor:innen. Beteiligte Einrichtung: Hochschule Fresenius, University of Applied Sciences, Idstein, Germany. Hauptthemen laut MEDLINE: Tandem Mass Spectrometry, 3-Hydroxybutyric Acid, Blood Chemical Analysis, Hydroxybutyrate Dehydrogenase. Die Arbeit nennt 26 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: β-Hydroxybutyrate (β-HB), the predominant circulating ketone body, serves as alternative energy source during periods of glucose deprivation. Accurate quantification of β-HB is essential for the clinical assessment and monitoring of ketosis and diabetic ketoacidosis (DKA). METHODS: We evaluated the analytical performance of the DiaSys β-Hydroxybutyrate 21 FS assay (BHB 21 FS), a one-step enzymatic assay based on β-hydroxybutyrate dehydrogenase-catalyzed oxidation of β-HB to generate NADH, which is photometrically measured at 340 nm. Performance characteristics, including linearity, repeatability, and total precision, were assessed in serum samples following CLSI guidelines. The BHB 21 FS assay was benchmarked against two commercial two-step assays and a non-enzymatic reference method (HILIC-ESI-MS/MS). RESULTS: The BHB 21 FS assay showed a linear measuring range from 0.05 up to 6.0-8.0 mmol/L and CLSI-compliant total precision (CV ≤3.3 %), notably at the clinically relevant 3 mmol/L cut-off (mean CV: 1.93 %). Method comparison demonstrated excellent correlation with EKF BHB LiquiColor® across three different analyzers (r = 0.999; bias ≤1 %), and high concordance with DiaSys BHB FS (r = 0.998; bias 5.0 %) and the HILIC-MS/MS reference method (r = 0.998; bias 4.0 %). Interference testing indicated robust performance even in highly hemolytic, lipemic, or icteric samples (≤10 % deviation). On platforms with active cooling compartments, calibration and onboard stability were maintained for 10-16 weeks. CONCLUSION: The DiaSys BHB 21 FS assay combines robust analytical performance and high concordance with both enzymatic and mass spectrometric methods. These features support its use as a reliable one-step assay for routine clinical assessment of ketosis and ketoacidosis.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Tandem Mass Spectrometry3-Hydroxybutyric AcidBlood Chemical AnalysisHydroxybutyrate DehydrogenaseEnzyme AssaysHumansReference StandardsChromatography, Liquid

Stichwörter der Autor:innen

Diabetic ketoacidosisEnzyme assayKetonesLiquid chromatographyMass spectrometryMethod validationPerformance evaluation

Substanzen

3-Hydroxybutyric AcidHydroxybutyrate Dehydrogenase

Autor:innen und Institutionen

  • Hochschule Fresenius, University of Applied Sciences, Idstein, Germany.
  • DiaSys Diagnostic Systems GmbH, Holzheim, Germany.
  • Institute of Molecular Medicine I, Medical Faculty, Heinrich Heine University Düsseldorf, Düsseldorf, Germany.
  • Hochschule Fresenius, University of Applied Sciences, Idstein, Germany; DiaServe Laboratories GmbH, Iffeldorf, Germany. Electronic address: matthias.grimmler@ext.hs-fresenius.de.

Zitieren

APA

Steinbach, M. K., Müller, H., Reichard, Y., Eulberg, K., Frömel, T., Lein, A., Peter, C., Prager, K., & Grimmler, M. (2026). Evaluation of an enzymatic one-step assay for Beta-Hydroxybutyrate blood testing on clinical chemistry platforms using HILIC-MS/MS as a non-enzymatic reference method. Clinica Chimica Acta. https://doi.org/10.1016/j.cca.2025.120703

Vancouver

Steinbach MK, Müller H, Reichard Y, Eulberg K, Frömel T, Lein A, et al. Evaluation of an enzymatic one-step assay for Beta-Hydroxybutyrate blood testing on clinical chemistry platforms using HILIC-MS/MS as a non-enzymatic reference method. Clinica Chimica Acta. 2026. doi:10.1016/j.cca.2025.120703

BibTeX
@article{Steinbach2026,
  author = {Steinbach, Max K and Müller, Holger and Reichard, Yanik and Eulberg, Konstantin and Frömel, Tobias and Lein, Alexandra and Peter, Christoph and Prager, Kai and Grimmler, Matthias},
  title = {Evaluation of an enzymatic one-step assay for Beta-Hydroxybutyrate blood testing on clinical chemistry platforms using HILIC-MS/MS as a non-enzymatic reference method},
  journal = {Clinica Chimica Acta},
  year = {2026},
  doi = {10.1016/j.cca.2025.120703}
}
Kostenlos starten →