Development of an SI traceable value assigned amino acid matrix-matched material to underpin European external quality assessment

Emily Whyte, Rachel Carling, Simon Cowen, Patrick Sears, Chris Hopley

Analytical and Bioanalytical Chemistry · 2025

DOI: 10.1007/s00216-025-05793-4 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Analytical and Bioanalytical Chemistry, 2025. Verfasst von Emily Whyte und 4 weiteren Autor:innen. Beteiligte Einrichtung: LGC, Teddington, Middlesex, TW11 0LY, UK. emilywhyte2017@gmail.com. Hauptthemen laut MEDLINE: Amino Acids, Quality Assurance, Health Care. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Department for Science, Innovation and Technology, Synnovis Innovation Accelerator Fund.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Plasma amino acid analysis plays a vital role in monitoring inherited metabolic diseases (IMDs), such as phenylketonuria (PKU), where dietary therapy is the cornerstone of management and patients are regularly monitored using blood or plasma phenylalanine throughout life. Recent findings from external quality assessment (EQA) schemes demonstrate large variability between results from different clinical laboratories. EQA schemes are hampered by factors including suboptimal sample type and frequency of distributions. Aiming to harmonise and standardise EQA schemes, this paper describes value assignment of an SI traceable matrix-matched amino acid material, and its use in an interlaboratory comparison (ERNDIM QAA-2106 study) that included 89 clinical laboratories from throughout Europe. The outcomes of the interlaboratory comparison are valuable in enabling clinical laboratories to refine their methodologies and, in a longer term, will lead to enhanced accuracy of results in clinical measurements and, ultimately, improved patient care in IMD management. A frozen, pooled human plasma sample was distributed to participating laboratories ( n = 88). The sample was value assigned for phenylalanine by the National Measurement Laboratory (NML) traceable to the international system of units (SI), at approximately the target value for dietary monitoring of phenylketonuria (360 µmol/L). Participants used their routine method of analysis to measure 24 amino acids in triplicate and results were compared against desirable analytical performance. Mean intra- and inter-laboratory variability were acceptable for 20/20 and 19/20 amino acids respectively. Mean bias relative to the consensus values was acceptable for 17/20 amino acids. Overall performance of phenylalanine was acceptable; however, 15/87 laboratories failed to meet acceptable imprecision and 8/87 had unacceptable bias; additionally, an overall negative bias for phenylalanine measurement was identified. This study demonstrates the importance of ensuring a matrix material with an assigned value of the analyte(s) of interest (traceable to the SI) is available for regular EQA. The significant bias in the routine measurement of phenylalanine identified has the potential to impact disease management, which aims to control blood phenylalanine concentrations for preventing adverse neurological outcomes.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Amino AcidsQuality Assurance, Health CareHumansEuropePhenylketonuriasPhenylalanineQuality Control

Stichwörter der Autor:innen

Amino acidsCertified reference materialExternal quality assessmentPhenylketonuriaPlasmaStandardisation

Substanzen

Amino AcidsPhenylalanine

Autor:innen und Institutionen

  • LGC, Teddington, Middlesex, TW11 0LY, UK. emilywhyte2017@gmail.com.
  • Faculty of Engineering and Physical Sciences, Univeristy of Surrey, Surrey, GU2 7XH, UK. emilywhyte2017@gmail.com.
  • Biochemical Sciences, Synnovis, Guys & St Thomas' NHSFT, London, SE1 7EH, UK.
  • European Research Network for Evaluation and Improvement of Screening, Diagnosis and Treatment of Inherited Disorders of Metabolism, Sheepbridge Centre, Chesterfield, Derbyshire, S41 9RX, UK.
  • GKT School Medical Education, Kings College London, Strand, London, WC2R 2LS, UK.
  • LGC, Teddington, Middlesex, TW11 0LY, UK.

Förderung

  • Department for Science, Innovation and Technology
  • Synnovis Innovation Accelerator Fund

Zitieren

APA

Whyte, E., Carling, R., Cowen, S., Sears, P., & Hopley, C. (2025). Development of an SI traceable value assigned amino acid matrix-matched material to underpin European external quality assessment. Analytical and Bioanalytical Chemistry. https://doi.org/10.1007/s00216-025-05793-4

Vancouver

Whyte E, Carling R, Cowen S, Sears P, Hopley C. Development of an SI traceable value assigned amino acid matrix-matched material to underpin European external quality assessment. Analytical and Bioanalytical Chemistry. 2025. doi:10.1007/s00216-025-05793-4

BibTeX
@article{Whyte2025,
  author = {Whyte, Emily and Carling, Rachel and Cowen, Simon and Sears, Patrick and Hopley, Chris},
  title = {Development of an SI traceable value assigned amino acid matrix-matched material to underpin European external quality assessment},
  journal = {Analytical and Bioanalytical Chemistry},
  year = {2025},
  doi = {10.1007/s00216-025-05793-4}
}
Kostenlos starten →