A novel whole-head RF coil design tailored for concurrent multichannel brain stimulation and imaging at 3T
Brain Stimulation · 2023-07 · 16 (4) · 1021-1031
DOI: 10.1016/j.brs.2023.05.025 ↗Kurz gefasst
Fachartikel in der Zeitschrift Brain Stimulation, 2023-07. Verfasst von Lucia I Navarro de Lara und 9 weiteren Autor:innen. Beteiligte Einrichtung: Athinoula A. Martinos Center for Biomedical Imaging, Dept. of Radiology, Massachusetts General Hospital, Charlestown, MA, 02129, USA; Harvard Medical…. Hauptthemen laut MEDLINE: Brain, Transcranial Magnetic Stimulation. Die Arbeit wurde bisher 19-mal zitiert (Crossref) und nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: NIDCD NIH HHS, NIBIB NIH HHS, NIH HHS.
Abstract (Originalsprache)
PURPOSE: Multichannel Transcranial Magnetic Stimulation (mTMS) [1] is a novel non-invasive brain stimulation technique allowing multiple sites to be stimulated simultaneously or sequentially under electronic control without movement of the coils. To enable simultaneous mTMS and MR imaging, we have designed and constructed a whole-head 28-channel receive-only RF coil at 3T. METHODS: A helmet-shaped structure was designed considering a specific layout for a mTMS system with holes for positioning the TMS units next to the scalp. Diameter of the TMS units defined the diameter of RF loops. The placement of the preamplifiers was designed to minimize possible interactions and to allow straightforward positioning of the mTMS units around the RF coil. Interactions between TMS-MRI were analyzed for the whole-head system extending the results presented in previous publications [2]. Both SNR- and g-factors maps were obtained to compare the imaging performance of the coil with commercial head coils. RESULTS: Sensitivity losses for the RF elements containing TMS units show a well-defined spatial pattern. Simulations indicate that the losses are predominantly caused by eddy currents on the coil wire windings. The average SNR performance of the TMSMR 28-channel coil is about 66% and 86% of the SNR of the 32/20-channel head coil respectively. The g-factor values of the TMSMR 28-channel coil are similar to the 32-channel coil and significantly better than the 20-channel coil. CONCLUSION: We present the TMSMR 28-channel coil, a head RF coil array to be integrated with a multichannel 3-axisTMS coil system, a novel tool that will enable causal mapping of human brain function.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Athinoula A. Martinos Center for Biomedical Imaging, Dept. of Radiology, Massachusetts General Hospital, Charlestown, MA, 02129, USA; Harvard Medical School, Boston, MA, 02115, USA. Electronic address: lnavarrodelara@mg…
- Athinoula A. Martinos Center for Biomedical Imaging, Dept. of Radiology, Massachusetts General Hospital, Charlestown, MA, 02129, USA; Harvard Medical School, Boston, MA, 02115, USA.
- Institute of Medical Physics and Radiation Protection, Department of Life Science Engineering, TH Mittelhessen University of Applied Sciences, Giessen, Germany; Department of Diagnostic and Interventional Radiology, Uni…
- Athinoula A. Martinos Center for Biomedical Imaging, Dept. of Radiology, Massachusetts General Hospital, Charlestown, MA, 02129, USA.
- Athinoula A. Martinos Center for Biomedical Imaging, Dept. of Radiology, Massachusetts General Hospital, Charlestown, MA, 02129, USA; Harvard Medical School, Boston, MA, 02115, USA; Department of Electrical and Computer…
Förderung
- NIDCD NIH HHS
- NIBIB NIH HHS
- NIH HHS
- NIMH NIH HHS
- National Institutes of Health
- Siemens USA
- Massachusetts General Hospital
- Rappaport Foundation
Zitieren
APA
Navarro de Lara, L. I., Stockmann, J. P., Meng, Q., Keil, B., Mareyam, A., Uluç, I., Daneshzand, M., Makarov, S., Wald, L. L., & Nummenmaa, A. (2023). A novel whole-head RF coil design tailored for concurrent multichannel brain stimulation and imaging at 3T. Brain Stimulation, 16(4), 1021-1031. https://doi.org/10.1016/j.brs.2023.05.025
Vancouver
Navarro de Lara LI, Stockmann JP, Meng Q, Keil B, Mareyam A, Uluç I, et al. A novel whole-head RF coil design tailored for concurrent multichannel brain stimulation and imaging at 3T. Brain Stimulation. 2023;16(4):1021-1031. doi:10.1016/j.brs.2023.05.025
BibTeX
@article{NavarrodeLara2023,
author = {Navarro de Lara, Lucia I and Stockmann, Jason P and Meng, Qinglei and Keil, Boris and Mareyam, Azma and Uluç, Işıl and Daneshzand, Mohammad and Makarov, Sergey and Wald, Lawrence L and Nummenmaa, Aapo},
title = {A novel whole-head RF coil design tailored for concurrent multichannel brain stimulation and imaging at 3T},
journal = {Brain Stimulation},
year = {2023},
volume = {16},
number = {4},
pages = {1021--1031},
doi = {10.1016/j.brs.2023.05.025}
}
EUVIMED