EUVIMEDEUVIMED entdecken
Crossref

Automated needle reconstruction for HDR prostate brachytherapy using a commercially compatible 3D-printed transrectal ultrasound sleeve with integrated EM tracking sensors

Catherine Holly Frank, Andrew Moyo, Dalton Griner, Kelli Anderson, Jonathan M Morris, Bradley J Stish, Jessica M Wilson, Mark R Waddle, Christopher L Deufel

Brachytherapy · 2026-07 · 25 (4) · 673-683

DOI: 10.1016/j.brachy.2026.04.002 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Brachytherapy, 2026-07. Verfasst von Catherine Holly Frank und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Mayo Clinic, Rochester, MN. Hauptthemen laut MEDLINE: Brachytherapy, Prostatic Neoplasms, Printing, Three-Dimensional. Die Arbeit nennt 21 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: Electromagnetic (EM) tracking can enhance the quality and efficiency of ultrasound-based HDR prostate brachytherapy by enabling automated digital needle reconstruction during treatment planning. This work integrates EM tracking sensors within a biocompatible, 3D-printed transrectal ultrasound (TRUS) sleeve engineered for compatibility with a commercially available probe and stepper to support clinical translation. METHODS AND MATERIALS: A clinical TRUS probe was digitally scanned, and a conformal sleeve was designed in CAD and 3D-printed using sterilizable materials. The sleeve incorporated channels for EM sensors and snap-on mounting features to ensure reproducible positioning relative to the probe. Sensors were placed at the probe apex within the TRUS axial imaging plane to minimize uncertainties from mounting or EM field distortions. Multiple sleeves were fabricated to evaluate consistency and performance. EM tracking-based reconstruction accuracy was assessed in a water-tank phantom containing seven needles placed to approximate the peripheral needle positions of a typical HDR prostate implant. RESULTS: EM-tracked needle reconstruction showed close agreement with manual, image-based reconstruction, yielding median (interquartile range) errors of 0.60 (0.28-0.78) mm for needle depths and 0.64 (0.40-0.93) mm for needle shafts. Sleeve mounting reproducibility had 90% confidence intervals whose absolute bounds were below 0.17 mm (3D offset), 0.09° (pitch), 0.06° (yaw), and 0.51° (roll). A patient-specific calibration procedure was developed to correct potential variability in 3D printing and sensor placement. The potential net reduction in procedure time using EM-based reconstruction versus manual reconstruction was approximately 28 minutes. CONCLUSIONS: Integrating EM sensors into a 3D-printed TRUS sleeve provided automated needle reconstruction with sub-millimeter accuracy for HDR prostate brachytherapy. Future clinical implementation is supported by the use of sterilizable, patient-safe materials and compatibility with standard TRUS systems.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

BrachytherapyProstatic NeoplasmsPrinting, Three-DimensionalMaleHumansNeedlesEquipment DesignPhantoms, ImagingUltrasonography

Stichwörter der Autor:innen

AutomationBrachytherapyDigitizationElectromagnetic tracking

Autor:innen und Institutionen

  • Department of Radiation Oncology, Mayo Clinic, Rochester, MN.
  • Anatomic Modeling Unit, Mayo Clinic, Rochester MN.
  • Department of Radiology, Mayo Clinic, Rochester, MN.
  • Department of Radiation Oncology, Mayo Clinic, Rochester, MN. Electronic address: deufel.christopher@mayo.edu.

Zitieren

APA

Frank, C. H., Moyo, A., Griner, D., Anderson, K., Morris, J. M., Stish, B. J., Wilson, J. M., Waddle, M. R., & Deufel, C. L. (2026). Automated needle reconstruction for HDR prostate brachytherapy using a commercially compatible 3D-printed transrectal ultrasound sleeve with integrated EM tracking sensors. Brachytherapy, 25(4), 673-683. https://doi.org/10.1016/j.brachy.2026.04.002

Vancouver

Frank CH, Moyo A, Griner D, Anderson K, Morris JM, Stish BJ, et al. Automated needle reconstruction for HDR prostate brachytherapy using a commercially compatible 3D-printed transrectal ultrasound sleeve with integrated EM tracking sensors. Brachytherapy. 2026;25(4):673-683. doi:10.1016/j.brachy.2026.04.002

BibTeX
@article{Frank2026,
  author = {Frank, Catherine Holly and Moyo, Andrew and Griner, Dalton and Anderson, Kelli and Morris, Jonathan M and Stish, Bradley J and Wilson, Jessica M and Waddle, Mark R and Deufel, Christopher L},
  title = {Automated needle reconstruction for HDR prostate brachytherapy using a commercially compatible 3D-printed transrectal ultrasound sleeve with integrated EM tracking sensors},
  journal = {Brachytherapy},
  year = {2026},
  volume = {25},
  number = {4},
  pages = {673--683},
  doi = {10.1016/j.brachy.2026.04.002}
}
Kostenlos starten →