Fachartikel in der Zeitschrift Brachytherapy, 2024. Verfasst von Cullen Boyle und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Thomas Jefferson University Hospital, Philadelphia, PA. Hauptthemen laut MEDLINE: Cone-Beam Computed Tomography, Brachytherapy, Phantoms, Imaging. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 22 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: A new mobile cone beam computed tomography (CBCT) imaging ring (IRm, Elekta, v2.10.6, Veenendaal, Netherlands) has recently been proposed for brachytherapy to improve procedure efficiency. We describe the commissioning process and end-to-end tests for GYN HDR brachytherapy employing IRm CBCT imaging. MATERIALS AND METHODS: Commissioning included imaging isocenter test, 3D image quality, 2D imaging quality, image dose, and tube characteristics. CIRS HDR GYN phantom and Venezia CT/MR gynecological applicator were used to perform end-to-end (E2E) tests and optimize workflow. Venezia applicator and four interstitial needles were inserted into the phantom and IRm CBCT images were acquired. Phantom and applicator were scanned with CT scanner (Siemens SOMATOM go.Open Pro) using department's pelvis imaging protocol. MR imaging was performed using 0.35T MR Linac TRUFI pulse sequence. CBCT images were registered to CT and MR using rigid registration to assess image quality and applicator geometry fidelity. RESULTS: All physics tests passed within acceptance tolerances. Registration of CBCT images to MR and CT scans was acceptable for applicator placement. Applicator registration of CBCT images to CT demonstrated excellent agreement of most distal source dwell position (

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Cone-Beam Computed TomographyBrachytherapyPhantoms, ImagingHumansFemaleGenital Neoplasms, FemaleRadiotherapy DosageRadiotherapy, Image-GuidedEquipment Design

Stichwörter der Autor:innen

BrachytherapyCBCTCTHDRMRIimaging

Autor:innen und Institutionen

  • Department of Radiation Oncology, Thomas Jefferson University Hospital, Philadelphia, PA.
  • Department of Radiation Oncology, Thomas Jefferson University Hospital, Philadelphia, PA. Electronic address: reza.taleei@jefferson.edu.

Zitieren

APA

Boyle, C., Mourtada, F., Anne, R., Wan, S., Chen, Y., Vinogradskiy, Y., & Taleei, R. (2024). Comprehensive commissioning of a cone beam CT imaging ring for treatment of HDR GYN patients. Brachytherapy. https://doi.org/10.1016/j.brachy.2024.07.008

Vancouver

Boyle C, Mourtada F, Anne R, Wan S, Chen Y, Vinogradskiy Y, et al. Comprehensive commissioning of a cone beam CT imaging ring for treatment of HDR GYN patients. Brachytherapy. 2024. doi:10.1016/j.brachy.2024.07.008

BibTeX
@article{Boyle2024,
  author = {Boyle, Cullen and Mourtada, Firas and Anne, Rani and Wan, Shuying and Chen, Yingxuan and Vinogradskiy, Yevgeniy and Taleei, Reza},
  title = {Comprehensive commissioning of a cone beam CT imaging ring for treatment of HDR GYN patients},
  journal = {Brachytherapy},
  year = {2024},
  doi = {10.1016/j.brachy.2024.07.008}
}
Kostenlos starten →