Mechanical needle guidance for ultrasound-guided parasagittal oblique in-plane paravertebral blocks: a cadaveric study
Brazilian Journal of Anesthesiology (English Edition) · 2026
DOI: 10.1016/j.bjane.2025.844716 ↗Kurz gefasst
Randomisierte kontrollierte Studie in der Zeitschrift Brazilian Journal of Anesthesiology (English Edition), 2026. Verfasst von Hesham Elsharkawy und 12 weiteren Autor:innen. Beteiligte Einrichtung: Case Western University, Pain and Healing Center, MetroHealth, The MetroHealth System, Case Western Reserve University School of Medicine, Cleveland,…. Hauptthemen laut MEDLINE: Ultrasonography, Interventional, Nerve Block. Die Arbeit nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Paravertebral blocks provide analgesia for a range of thoracoabdominal surgeries. However, visualizing the needle tip during the procedure can be challenging, especially for clinicians with limited experience, because the target is deep. We therefore tested the primary hypothesis that needle guidance by the Infiniti Plus system improves ultrasound visualization of the needle tip during thoracic paravertebral blocks performed by novice residents.
Methods
Nineteen clinical anesthesia residents each performed 20 bilateral ultrasound-guided thoracic paravertebral blocks (T2–T11) on 17 unembalmed cadavers, with and without the use of a fixed-angle mechanical needle guide in a randomized crossover design. The primary outcome, percent perfect needle visibility, was compared between guided and unguided methods using a paired t-test. Secondary outcomes, including time to needle visualization, number of needle insertion attempts, and subjective ease-of-use ratings, were analyzed using paired t-tests and Wilcoxon signed-rank tests, respectively. Inter-rater reliability for overall perception ratings was assessed using the Intraclass Correlation Coefficient (ICC).
Results
There were no significant differences in needle-target visualization (62% ± 17% with guidance vs. 64% ± 18% without, p = 0.15), time to target (HR = 1.00 [95% CI 0.86–1.16], p = 0.99), procedural difficulty scores, or number of insertion attempts between guided and unguided blocks.
Conclusion
The Infiniti Plus mechanical needle guide did not demonstrate improved ultrasound needle tip visualization during thoracic paravertebral blocks performed by novice clinicians in cadavers.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Case Western University, Pain and Healing Center, MetroHealth, The MetroHealth System, Case Western Reserve University School of Medicine, Cleveland, OH, USA; Department of Outcomes Research, Houston, Texas, USA; Madrid…
- Department of Outcomes Research, Houston, Texas, USA; Madrid Morphological Research Center, Madrid, Spain; Ataturk University, Anesthesiology Clinical Research Office, Erzurum, Turkey.
- Department of Outcomes Research, Houston, Texas, USA; National Hospital Organization, Murayama Medical Center, Department of Anesthesiology, Musashimurayama, Tokyo, Japan; National Defense Medical College, Department of…
- Department of Outcomes Research, Houston, Texas, USA; Madrid Morphological Research Center, Madrid, Spain; Ataturk University, Anesthesiology Clinical Research Office, Erzurum, Turkey; Penn State University Milton S Her…
- Department of Outcomes Research, Houston, Texas, USA; Cleveland Clinic, Quantitative Health Sciences, Cleveland, OH, USA.
- The MetroHealth System, Case Western Reserve University School of Medicine, Department of Anesthesiology and Perioperative Services, Cleveland, OH, USA. Electronic address: nicolas.masdalessandro@gmail.com.
Zitieren
APA
Elsharkawy, H., Altinpulluk, E. Y., Shimada, T., Ince, I., Mao, G., Mas D Alessandro, N. M., Soliman, L. M., Harsha, N., Kelava, M., Drake, R., Pu, X., Sessler, D. I., & Turan, A. (2026). Mechanical needle guidance for ultrasound-guided parasagittal oblique in-plane paravertebral blocks: a cadaveric study. Brazilian Journal of Anesthesiology (English Edition). https://doi.org/10.1016/j.bjane.2025.844716
Vancouver
Elsharkawy H, Altinpulluk EY, Shimada T, Ince I, Mao G, Mas D Alessandro NM, et al. Mechanical needle guidance for ultrasound-guided parasagittal oblique in-plane paravertebral blocks: a cadaveric study. Brazilian Journal of Anesthesiology (English Edition). 2026. doi:10.1016/j.bjane.2025.844716
BibTeX
@article{Elsharkawy2026,
author = {Elsharkawy, Hesham and Altinpulluk, Ece Yamak and Shimada, Tetsuya and Ince, Ilker and Mao, Guangmei and Mas D Alessandro, Nicolas Mario and Soliman, Loran Mounir and Harsha, Nair and Kelava, Marta and Drake, Richard and Pu, Xuan and Sessler, Daniel I and Turan, Alparslan},
title = {Mechanical needle guidance for ultrasound-guided parasagittal oblique in-plane paravertebral blocks: a cadaveric study},
journal = {Brazilian Journal of Anesthesiology (English Edition)},
year = {2026},
doi = {10.1016/j.bjane.2025.844716}
}
EUVIMED