Five commercial ‘depth of anaesthesia’ monitors provide discordant clinical recommendations in response to identical emergence-like EEG signals

Darren Hight, Matthias Kreuzer, Gesar Ugen, Peter Schuller, Frank Stüber, Jamie Sleigh, Heiko A Kaiser

British Journal of Anaesthesia · 2023-05 · 130 (5) · 536-545

DOI: 10.1016/j.bja.2022.12.026 ↗

Kurz gefasst

Fachartikel in der Zeitschrift British Journal of Anaesthesia, 2023-05. Verfasst von Darren Hight und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Anaesthesiology and Pain Medicine, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland. Electronic address: da…. Hauptthemen laut MEDLINE: Monitoring, Intraoperative, Anesthesiology. Die Arbeit wurde bisher 71-mal zitiert (Crossref) und nennt 36 Literaturangaben. Gefördert von: University of Bern, Gottfried und Julia Bangerter-Rhyner-Stiftung.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: 'Depth of anaesthesia' monitors claim to measure hypnotic depth during general anaesthesia from the EEG, and clinicians could reasonably expect agreement between monitors if presented with the same EEG signal. We took 52 EEG signals showing intraoperative patterns of diminished anaesthesia, similar to those that occur during emergence (after surgery) and subjected them to analysis by five commercially available monitors. METHODS: We compared five monitors (BIS, Entropy-SE, Narcotrend, qCON, and Sedline) to see if index values remained within, or moved out of, each monitors' recommended index range for general anaesthesia for at least 2 min during a period of supposed lighter anaesthesia, as observed by changes in the EEG spectrogram obtained in a previous study. RESULTS: Of the 52 cases, 27 (52%) had at least one monitor warning of potentially inadequate hypnosis (index above range) and 16 of the 52 cases (31%) had at least one monitor signifying excessive hypnotic depth (index below clinical range). Of the 52 cases, only 16 (31%) showed concordance between all five monitors. Nineteen cases (36%) had one monitor discordant compared with the remaining four, and 17 cases (33%) had two monitors in disagreement with the remaining three. CONCLUSIONS: Many clinical providers still rely on index values and manufacturer's recommended ranges for titration decision making. That two-thirds of cases showed discordant recommendations given identical EEG data, and that one-third signified excessive hypnotic depth where the EEG would suggest a lighter hypnotic state, emphasizes the importance of personalised EEG interpretation as an essential clinical skill.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Monitoring, IntraoperativeAnesthesiologyHumansAnesthesia, GeneralHypnotics and SedativesElectroencephalography

Stichwörter der Autor:innen

BISNarcotrendSedlineSpectral Entropydepth of anaesthesia monitorelectroencephalographygeneral anaesthesiaqCON

Substanzen

Hypnotics and Sedatives

Autor:innen und Institutionen

  • Department of Anaesthesiology and Pain Medicine, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland. Electronic address: darren.hight@insel.ch.
  • Department of Anesthesiology and Intensive Care Medicine, Technical University of Munich, School of Medicine, Munich, Germany.
  • Department of Anaesthesiology and Pain Medicine, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland.
  • Department of Anaesthesia, Cairns Hospital, Cairns, QLD, Australia.
  • Department of Anaesthesia, Waikato Clinical School, University of Auckland, Hamilton, New Zealand.
  • Department of Anaesthesiology and Pain Medicine, Inselspital, Bern University Hospital, University of Bern, Bern, Switzerland; Centre for Anaesthesiology and Intensive Care Medicine, Hirslanden Klinik Aarau, Hirslanden…

Förderung

  • University of Bern
  • Gottfried und Julia Bangerter-Rhyner-Stiftung

Zitieren

APA

Hight, D., Kreuzer, M., Ugen, G., Schuller, P., Stüber, F., Sleigh, J., & Kaiser, H. A. (2023). Five commercial ‘depth of anaesthesia’ monitors provide discordant clinical recommendations in response to identical emergence-like EEG signals. British Journal of Anaesthesia, 130(5), 536-545. https://doi.org/10.1016/j.bja.2022.12.026

Vancouver

Hight D, Kreuzer M, Ugen G, Schuller P, Stüber F, Sleigh J, et al. Five commercial ‘depth of anaesthesia’ monitors provide discordant clinical recommendations in response to identical emergence-like EEG signals. British Journal of Anaesthesia. 2023;130(5):536-545. doi:10.1016/j.bja.2022.12.026

BibTeX
@article{Hight2023,
  author = {Hight, Darren and Kreuzer, Matthias and Ugen, Gesar and Schuller, Peter and Stüber, Frank and Sleigh, Jamie and Kaiser, Heiko A},
  title = {Five commercial ‘depth of anaesthesia’ monitors provide discordant clinical recommendations in response to identical emergence-like EEG signals},
  journal = {British Journal of Anaesthesia},
  year = {2023},
  volume = {130},
  number = {5},
  pages = {536--545},
  doi = {10.1016/j.bja.2022.12.026}
}
Kostenlos starten →