High-performance molecularly imprinted electrochemical sensor based on 3D Ti3C2Tx/Ni(OH)2 for trace detection of lomefloxacin
Biosensors and Bioelectronics · 2026-09 · 307 · 118774
DOI: 10.1016/j.bios.2026.118774 ↗Kurz gefasst
Evaluationsstudie in der Zeitschrift Biosensors and Bioelectronics, 2026-09. Verfasst von Jing Tan und 9 weiteren Autor:innen. Beteiligte Einrichtung: State Key Laboratory of Chemistry and Utilization of Carbon Based Energy Resources, College of Chemistry, Xinjiang University, Urumqi, Xinjiang, 8300…. Hauptthemen laut MEDLINE: Titanium, Molecular Imprinting, Biosensing Techniques, Electrochemical Techniques. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 39 Literaturangaben. Gefördert von: National Natural Science Foundation of China.
Abstract (Originalsprache)
To address the challenge of recognizing lomefloxacin (LMF) residues in the environment and food, this work describes the development of a molecularly imprinted electrochemical sensor (MIECS) that combines a 3D Ti3C2Tx/Ni(OH)2 composite with molecularly imprinted polymers (MIPs). The 3D Ti3C2Tx/Ni(OH)2 framework has a larger surface area and more active sites, allowing for efficient electron and ion transport. Meanwhile, pyrrole, which serves as the functional monomer for MIP synthesis, adds imide functional groups and strong proton conductivity, considerably improving the sensor's particular identification capacity for LMF. The MIECS's sensitivity and selectivity are further enhanced by a potent synergistic interaction between these components. The sensor's exceptional recognition capability toward LMF. A wide linear response was recorded spanning the concentration range of 0.5 to 105 nM, having the limit of detection as low as 0.0157 nM, alongside remarkable stability and anti-interference performance. The sensor also delivered consistent recovery rates in real sample analyses-such as milk and lake water-validated by comparative studies with liquid chromatography. The use of 3D Ti3C2Tx-based MIECS in ultrasensitive antibiotic monitoring is made possible by this work.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- State Key Laboratory of Chemistry and Utilization of Carbon Based Energy Resources, College of Chemistry, Xinjiang University, Urumqi, Xinjiang, 830017, PR China.
- State Key Laboratory of Chemistry and Utilization of Carbon Based Energy Resources, College of Chemistry, Xinjiang University, Urumqi, Xinjiang, 830017, PR China. Electronic address: tursunabdir@xju.edu.cn.
- State Key Laboratory of Chemistry and Utilization of Carbon Based Energy Resources, College of Chemical Engineering, Xinjiang University, Urumqi, Xinjiang, 830017, PR China. Electronic address: jruxangul@xju.edu.cn.
- Xinjiang Key Laboratory of New Energy Materials and Green Chemical Engineering, Xinjiang Institute of Engineering, Urumqi, Xinjiang Uygur Autonomous Region, 830023, PR China.
Förderung
- National Natural Science Foundation of China
Zitieren
APA
Tan, J., Zhang, G., Abdiryim, T., Jamal, R., Ali, A., Xu, F., Zhou, Y., Song, K., Li, J., & Wang, L. (2026). High-performance molecularly imprinted electrochemical sensor based on 3D Ti3C2Tx/Ni(OH)2 for trace detection of lomefloxacin. Biosensors and Bioelectronics, 307, 118774. https://doi.org/10.1016/j.bios.2026.118774
Vancouver
Tan J, Zhang G, Abdiryim T, Jamal R, Ali A, Xu F, et al. High-performance molecularly imprinted electrochemical sensor based on 3D Ti3C2Tx/Ni(OH)2 for trace detection of lomefloxacin. Biosensors and Bioelectronics. 2026;307:118774. doi:10.1016/j.bios.2026.118774
BibTeX
@article{Tan2026,
author = {Tan, Jing and Zhang, Guoliang and Abdiryim, Tursun and Jamal, Ruxangul and Ali, Ahmat and Xu, Feng and Zhou, Yanqiang and Song, Kai and Li, Jiabei and Wang, Luyao},
title = {High-performance molecularly imprinted electrochemical sensor based on 3D Ti3C2Tx/Ni(OH)2 for trace detection of lomefloxacin},
journal = {Biosensors and Bioelectronics},
year = {2026},
volume = {307},
pages = {118774},
doi = {10.1016/j.bios.2026.118774}
}
EUVIMED