CapSense-MIP: Self-operating molecularly imprinted polymer (MIP) biosensor for point-of-care diagnostics
Biosensors and Bioelectronics · 2025-10 · 286 · 117599
DOI: 10.1016/j.bios.2025.117599 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biosensors and Bioelectronics, 2025-10. Verfasst von Azam Zare und 6 weiteren Autor:innen. Beteiligte Einrichtung: BioMEMS and Bioinspired Microfluidic Laboratory, Department of Biomedical Engineering, University of Calgary, Calgary, Alberta, T2N 1N4, Canada; Depa…. Hauptthemen laut MEDLINE: Biosensing Techniques, Molecularly Imprinted Polymers, Point-of-Care Systems. Die Arbeit wurde bisher 16-mal zitiert (Crossref) und nennt 64 Literaturangaben. Gefördert von: Canadian Institutes of Health Research, Canada Research Chairs Program, Natural Sciences and Engineering Research Council of Canada.
Abstract (Originalsprache)
Molecularly imprinted polymer (MIP) biosensors hold immense promise for point-of-care (POC) diagnostics due to their exceptional robustness, long shelf-life stability, selectivity, and ability to detect diverse biomarkers across their (patho)physiological ranges. However, their full potential within practical POC devices remains constrained by technical challenges, including the need for precise incubation control, effective washing of non-specific bindings, and consistent fluid handling in miniaturized systems. Without addressing these limitations, their ability to reliably operate in complex bodily fluids and within critical physiological detection ranges is significantly compromised. To address these barriers, we developed the CapSense-MIP, a capillary microfluidic biosensing platform that automates key MIP biosensing steps-sample aliquoting, incubation, washing, and sensing-using a fully integrated and self-powered system driven by sequential liquid delivery to the biosensing chamber. It incorporates an inlet regulator for precise aliquoting and incubation control, a suction-enhanced fiber for optimized washing, and the regulated channels and valve network to ensure high sensitivity and reproducibility, and a wide dynamic detection range that meets (patho)physiological needs. As a proof of concept, the CapSense-MIP was validated for agmatine detection, achieving a wide linear range of 1.0 nM-10 μM and an impressive limit of detection of 0.1 nM, with optimized protocols for both phosphate-buffered saline and human plasma. By addressing these critical challenges and leveraging the long shelf-life stability of MIP biosensors, the CapSense-MIP not only eliminates traditionally manual workflows but also delivers scalable, rapid, and cost-effective solutions for POC diagnostics, unlocking transformative potential for applications in healthcare and environmental monitoring.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- BioMEMS and Bioinspired Microfluidic Laboratory, Department of Biomedical Engineering, University of Calgary, Calgary, Alberta, T2N 1N4, Canada; Department of Biomedical Engineering, University of Calgary, Calgary, Albe…
- BioMEMS and Bioinspired Microfluidic Laboratory, Department of Biomedical Engineering, University of Calgary, Calgary, Alberta, T2N 1N4, Canada; Department of Mechanical and Manufacturing Engineering, University of Calg…
- BioMEMS and Bioinspired Microfluidic Laboratory, Department of Biomedical Engineering, University of Calgary, Calgary, Alberta, T2N 1N4, Canada; Department of Electrical and Software Engineering, University of Calgary,…
- Department of Biological Sciences, University of Calgary, 2500 University Drive, NW, Calgary, Alberta, T2N 1N4, Canada.
- Department of Physiology, Applied Physiology Research Center, Cardiovascular Research Institute, Isfahan University of Medical Sciences, Isfahan, 81746-73461, Iran.
Förderung
- Canadian Institutes of Health Research
- Canada Research Chairs Program
- Natural Sciences and Engineering Research Council of Canada
- Institute of Infection and Immunity
- Mitacs Inc
Zitieren
APA
Zare, A., Babamiri, B., Hassani, M., Khalghollah, M., Mohammadi, M., Haghjooy Javanmard, S., & Sanati Nezhad, A. (2025). CapSense-MIP: Self-operating molecularly imprinted polymer (MIP) biosensor for point-of-care diagnostics. Biosensors and Bioelectronics, 286, 117599. https://doi.org/10.1016/j.bios.2025.117599
Vancouver
Zare A, Babamiri B, Hassani M, Khalghollah M, Mohammadi M, Haghjooy Javanmard S, et al. CapSense-MIP: Self-operating molecularly imprinted polymer (MIP) biosensor for point-of-care diagnostics. Biosensors and Bioelectronics. 2025;286:117599. doi:10.1016/j.bios.2025.117599
BibTeX
@article{Zare2025,
author = {Zare, Azam and Babamiri, Bahareh and Hassani, Mohsen and Khalghollah, Mahmood and Mohammadi, Mehdi and Haghjooy Javanmard, Shaghayegh and Sanati Nezhad, Amir},
title = {CapSense-MIP: Self-operating molecularly imprinted polymer (MIP) biosensor for point-of-care diagnostics},
journal = {Biosensors and Bioelectronics},
year = {2025},
volume = {286},
pages = {117599},
doi = {10.1016/j.bios.2025.117599}
}
EUVIMED