EUVIMEDEUVIMED entdecken
EvaluationsstudieOpen Access · CC BY-NC-ND 4.0

A functionalized microwave biosensor for rapid, reagent-free detection of E. coli in water samples

Weijia Cui, Jin Wang, Clarissa Ding, Phillippe Van Cappellen, Emmanuel A Ho, Carolyn L Ren

Biosensors and Bioelectronics · 2025-06 · 278 · 117334

DOI: 10.1016/j.bios.2025.117334 ↗

Kurz gefasst

Evaluationsstudie in der Zeitschrift Biosensors and Bioelectronics, 2025-06. Verfasst von Weijia Cui und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Mechanical and Mechatronics Engineering, University of Waterloo, Canada. Hauptthemen laut MEDLINE: Biosensing Techniques, Water Microbiology, Escherichia coli O157, Escherichia coli. Die Arbeit wurde bisher 20-mal zitiert (Crossref) und nennt 33 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Canada Excellence Research Chairs, University of Waterloo, Natural Sciences and Engineering Research Council of Canada.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Escherichia coli (E. coli) O157:H7 (O157), one of the most common Shiga toxin-producing E. coli, can contaminate water systems causing severe illnesses often accompanied with diarrhea and sometimes life threatening. Frequent monitoring of E. coli in water systems is critical to protect public health. Most traditional methods for E. coli detection are slow in responding to E. coli outbreaks due to the need for sample transportation from the site to the lab, expensive equipment, and highly trained personnel for the detection. This work presents a novel reagent-free detection method that employs a microwave biosensor functionalized with an antibody specific to E. coli to offer rapid and sensitive E. coli detection. By monitoring the resonance frequency shift caused by the binding between the E. coli in the water sample and the antibody coated on the sensor using a vector network analyzer (VNA), this microwave-based biosensor achieved a limit of detection (LOD) of 647 CFU/ml. This LOD can be further reduced to 6.47 CFU/ml with a simple preconcentration step prior to the sensing procedure. The sensor has also been tested to detect E. coli in natural water systems with a low-cost, palm-sized portable VNA, suggesting its excellent feasibility for real-time on-site E.coli detection.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Biosensing TechniquesWater MicrobiologyEscherichia coli O157Escherichia coliMicrowavesLimit of DetectionEquipment DesignAntibodies, ImmobilizedHumans

Stichwörter der Autor:innen

BiosensorsE. coliMicrowave sensingReagent freeWater quality control

Substanzen

Antibodies, Immobilized

Autor:innen und Institutionen

  • Department of Mechanical and Mechatronics Engineering, University of Waterloo, Canada.
  • School of Pharmacy, University of Waterloo, Canada.
  • Department of Earth and Environmental Sciences, University of Waterloo, Canada; Water Institute, University of Waterloo, Canada.
  • School of Pharmacy, University of Waterloo, Canada; Water Institute, University of Waterloo, Canada.
  • Department of Mechanical and Mechatronics Engineering, University of Waterloo, Canada; Water Institute, University of Waterloo, Canada; Waterloo Institute for Nanotechnology, University of Waterloo, Canada. Electronic a…

Förderung

  • Canada Excellence Research Chairs
  • University of Waterloo
  • Natural Sciences and Engineering Research Council of Canada
  • Water Institute of the Gulf
  • Canada Research Chairs

Zitieren

APA

Cui, W., Wang, J., Ding, C., Van Cappellen, P., Ho, E. A., & Ren, C. L. (2025). A functionalized microwave biosensor for rapid, reagent-free detection of E. coli in water samples. Biosensors and Bioelectronics, 278, 117334. https://doi.org/10.1016/j.bios.2025.117334

Vancouver

Cui W, Wang J, Ding C, Van Cappellen P, Ho EA, Ren CL. A functionalized microwave biosensor for rapid, reagent-free detection of E. coli in water samples. Biosensors and Bioelectronics. 2025;278:117334. doi:10.1016/j.bios.2025.117334

BibTeX
@article{Cui2025,
  author = {Cui, Weijia and Wang, Jin and Ding, Clarissa and Van Cappellen, Phillippe and Ho, Emmanuel A and Ren, Carolyn L},
  title = {A functionalized microwave biosensor for rapid, reagent-free detection of E. coli in water samples},
  journal = {Biosensors and Bioelectronics},
  year = {2025},
  volume = {278},
  pages = {117334},
  doi = {10.1016/j.bios.2025.117334}
}