Hollow-microsphere-integrated optofluidic immunochip for myocardial infarction biomarker microanalysis
Biosensors and Bioelectronics · 2024-03 · 248 · 115970
DOI: 10.1016/j.bios.2023.115970 ↗Kurz gefasst
Fachartikel in der Zeitschrift Biosensors and Bioelectronics, 2024-03. Verfasst von Panpan Niu und 9 weiteren Autor:innen. Beteiligte Einrichtung: School of Precision Instrument and Opto-electronics Engineering, Tianjin University, Tianjin, 300072, China; Key Laboratory of Opto-electronics Infor…. Hauptthemen laut MEDLINE: Biosensing Techniques, Myocardial Infarction. Die Arbeit wurde bisher 50-mal zitiert (Crossref) und nennt 45 Literaturangaben.
Abstract (Originalsprache)
This work developed an optofluidic immunochip that uses whispering gallery mode with fiber laser enhancement, for the rapid detection of a key biomarker cardiac troponin I for acute myocardial infarction (AMI). The immunochip adopted an innovative design, using perforated hollow glass microspheres (HGMS) as carriers, with antibodies immobilized on the inner surface of the HGMS, thereby achieving ultra-low sample consumption. The performance of the immunochip was improved by fiber laser, including spectral width compression to 0.019 nm, optical signal-to-noise ratio amplification to 63.17 dB, and an enhancement in the limit of detection to 5 pg/mL. Moreover, this immunochip can provide results within 15 min, making it highly suitable for early AMI risk management. Compared to the standard electrochemiluminescence detection method, although some differences exist in the results of the immunochip due to the principle of detection and differences in antibody affinity, its positive reference value can be calculated as 0.0754 ng/mL, with a successful recognition rate of 88% for positive patients. The immunosensor is integrated on a polydimethylsiloxane substrate, with a compact size suitable for use in point-of-care devices and AMI self-screening, as well as rapid disease screening and microanalysis of various biomarkers, offering new possibilities for applications in these fields.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- School of Precision Instrument and Opto-electronics Engineering, Tianjin University, Tianjin, 300072, China; Key Laboratory of Opto-electronics Information Technology (Tianjin University), Key Laboratory of Micro Opto-e…
- Department of Cardiology, Tianjin Medical University General Hospital, Tianjin, 300052, China.
Zitieren
APA
Niu, P., Jiang, J., Liu, K., Zhou, X., Wang, S., Xu, T., Wang, T., Li, Y., Yang, Q., & Liu, T. (2024). Hollow-microsphere-integrated optofluidic immunochip for myocardial infarction biomarker microanalysis. Biosensors and Bioelectronics, 248, 115970. https://doi.org/10.1016/j.bios.2023.115970
Vancouver
Niu P, Jiang J, Liu K, Zhou X, Wang S, Xu T, et al. Hollow-microsphere-integrated optofluidic immunochip for myocardial infarction biomarker microanalysis. Biosensors and Bioelectronics. 2024;248:115970. doi:10.1016/j.bios.2023.115970
BibTeX
@article{Niu2024,
author = {Niu, Panpan and Jiang, Junfeng and Liu, Kun and Zhou, Xin and Wang, Shuang and Xu, Tianhua and Wang, Tong and Li, Yongle and Yang, Qing and Liu, Tiegen},
title = {Hollow-microsphere-integrated optofluidic immunochip for myocardial infarction biomarker microanalysis},
journal = {Biosensors and Bioelectronics},
year = {2024},
volume = {248},
pages = {115970},
doi = {10.1016/j.bios.2023.115970}
}