Long-term results of a vitamin E–blended polyethylene liner in total hip arthroplasty: a minimum ten year computer aided design based evaluation of wear and component positioning

Eren Demir, Mahfuz Rasuli, Michael Nienhaus, Philipp Schippers, Martin Naisan, Yazan Noufal, Philipp Drees, Yama Afghanyar

International Orthopaedics · 2026

DOI: 10.1007/s00264-026-06997-8 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Orthopaedics, 2026. Verfasst von Eren Demir und 7 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedics and Traumatology, University Medical Center of the Johannes Gutenberg University, Mainz, Germany. eren.demir@unimedizin-ma…. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip, Vitamin E, Hip Prosthesis, Polyethylene. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Mathys Ltd., Bettlach, Switzerland, Universitätsmedizin der Johannes Gutenberg-Universität Mainz.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Aseptic loosening remains a leading cause of revision in total hip arthroplasty (THA), with polyethylene wear representing a key contributing factor. Vitamin E–blended highly cross-linked polyethylene (VEPE) has been introduced to improve oxidative stability and long-term wear performance; however, long-term clinical data remain limited.

Methods

This retrospective study evaluates long-term polyethylene wear and component positioning in a cohort of patients who underwent primary THA using a vitamin E–blended polyethylene monoblock acetabular component (RM Pressfit vitamys®). Patients were followed for a minimum of ten years. Radiographic assessment was performed using a fully automated, observer-independent CAD-based software (RayMatch®), enabling measurement of cup inclination, anteversion, and polyethylene wear through 2D–3D registration of implant CAD models.

Results

A total of 57 hips with complete radiographic follow-up were analyzed. Mean polyethylene wear rates were 26.9 µm/year at five years and 18.2 µm/year at ten years, remaining well below commonly reported thresholds associated with osteolysis. Cup inclination remained stable over time, whereas a gradual increase in measured anteversion was observed during follow-up.

Conclusion

At a minimum ten year follow-up, the investigated vitamin E–blended polyethylene monoblock cup demonstrated favorable long-term wear characteristics and stable component positioning. Further comparative studies are warranted to compare its long-term performance with other acetabular cup and liner designs.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, HipVitamin EHip ProsthesisPolyethyleneProsthesis DesignComputer-Aided DesignHumansRetrospective StudiesProsthesis FailureFemaleAgedMiddle AgedMaleFollow-Up StudiesAcetabulumHip JointRadiographyAdultAged, 80 and over

Stichwörter der Autor:innen

Acetabular component positioningPolyethylene wearTotal hip arthroplastyVitamin E–blended polyethylene

Substanzen

Vitamin EPolyethylene

Autor:innen und Institutionen

  • Department of Orthopaedics and Traumatology, University Medical Center of the Johannes Gutenberg University, Mainz, Germany. eren.demir@unimedizin-mainz.de.
  • Department of Orthopaedics and Traumatology, Johannes Gutenberg University Mainz, Mainz, Germany.
  • Department of Orthopaedic Surgery, St. Josefs Hospital, Wiesbaden, Germany.

Förderung

  • Mathys Ltd., Bettlach, Switzerland
  • Universitätsmedizin der Johannes Gutenberg-Universität Mainz

Zitieren

APA

Demir, E., Rasuli, M., Nienhaus, M., Schippers, P., Naisan, M., Noufal, Y., Drees, P., & Afghanyar, Y. (2026). Long-term results of a vitamin E–blended polyethylene liner in total hip arthroplasty: a minimum ten year computer aided design based evaluation of wear and component positioning. International Orthopaedics. https://doi.org/10.1007/s00264-026-06997-8

Vancouver

Demir E, Rasuli M, Nienhaus M, Schippers P, Naisan M, Noufal Y, et al. Long-term results of a vitamin E–blended polyethylene liner in total hip arthroplasty: a minimum ten year computer aided design based evaluation of wear and component positioning. International Orthopaedics. 2026. doi:10.1007/s00264-026-06997-8

BibTeX
@article{Demir2026,
  author = {Demir, Eren and Rasuli, Mahfuz and Nienhaus, Michael and Schippers, Philipp and Naisan, Martin and Noufal, Yazan and Drees, Philipp and Afghanyar, Yama},
  title = {Long-term results of a vitamin E–blended polyethylene liner in total hip arthroplasty: a minimum ten year computer aided design based evaluation of wear and component positioning},
  journal = {International Orthopaedics},
  year = {2026},
  doi = {10.1007/s00264-026-06997-8}
}
Kostenlos starten →