EUVIMEDEUVIMED entdecken
Crossref

Wear of Third-Generation Cross-Linked Polyethylene in Primary Total Hip Arthroplasty: A 10-Year Analysis

Omid Jalali, Joseph Fiske, Michael G DeJardin, Travis S Scudday, Steven L Barnett, Robert S Gorab

The Journal of Arthroplasty · 2024

DOI: 10.1016/j.arth.2023.11.016 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of Arthroplasty, 2024. Verfasst von Omid Jalali und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopaedic Surgery, University of California, San Diego, San Diego, California. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Hip, Hip Prosthesis, Osteolysis. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 31 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Sequential modifications to the manufacturing process of highly cross-linked polyethylene (HXLPE) have improved the wear resistance and implant survival of these liners in total hip arthroplasty (THA). However, no study has examined the long-term (mean 10 year) wear rates and clinical outcomes of third-generation HXLPE in THA. The aim of our study was to report the longest-to-date analysis of wear rates and clinical outcomes of a third-generation HXLPE liner. METHODS: A series of 133 THAs using a specific HXLPE acetabular liner were retrospectively evaluated. Linear and volumetric wear rates were determined using a validated radiographic technique and clinical outcomes were analyzed. Multivariate analyses were performed to determine risk factors for accelerated wear. RESULTS: At a mean follow-up of 10.4 years (range, 8 to 13.4), the mean linear wear rate was 0.0172 mm/year and the mean volumetric wear rate was 16.99 mm3/year. There were no instances of osteolysis or mechanical failures at any time point and there was a 100% acetabular component survival rate. Younger age and use of offset liners were independent risk factors for increased wear (P < .01 for both). CONCLUSIONS: Our series of a third-generation HXLPE demonstrated very low wear rates and excellent implant survival at a mean of 10.4 years following primary THA. Future comparative studies at the 15- and 20-year follow-up timepoints are necessary to determine if such findings translate to true improvements in the tribological properties and longevity of these liners when compared to previous generations of HXLPE liners.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Arthroplasty, Replacement, HipHip ProsthesisOsteolysisHumansPolyethyleneRetrospective StudiesProsthesis FailureProsthesis DesignFollow-Up Studies

Stichwörter der Autor:innen

cross-linked polyethyleneimplant designosteolysistotal hip arthroplastywear

Substanzen

Polyethylene

Autor:innen und Institutionen

  • Department of Orthopaedic Surgery, University of California, San Diego, San Diego, California.
  • College of Osteopathic Medicine, Touro University California, Vallejo, California.
  • Department of Orthopedic Research, Hoag Orthopedics, Irvine, California.
  • Department of Orthopaedic Surgery, Hoag Orthopedic Institute, Irvine, California.

Zitieren

APA

Jalali, O., Fiske, J., DeJardin, M. G., Scudday, T. S., Barnett, S. L., & Gorab, R. S. (2024). Wear of Third-Generation Cross-Linked Polyethylene in Primary Total Hip Arthroplasty: A 10-Year Analysis. The Journal of Arthroplasty. https://doi.org/10.1016/j.arth.2023.11.016

Vancouver

Jalali O, Fiske J, DeJardin MG, Scudday TS, Barnett SL, Gorab RS. Wear of Third-Generation Cross-Linked Polyethylene in Primary Total Hip Arthroplasty: A 10-Year Analysis. The Journal of Arthroplasty. 2024. doi:10.1016/j.arth.2023.11.016

BibTeX
@article{Jalali2024,
  author = {Jalali, Omid and Fiske, Joseph and DeJardin, Michael G and Scudday, Travis S and Barnett, Steven L and Gorab, Robert S},
  title = {Wear of Third-Generation Cross-Linked Polyethylene in Primary Total Hip Arthroplasty: A 10-Year Analysis},
  journal = {The Journal of Arthroplasty},
  year = {2024},
  doi = {10.1016/j.arth.2023.11.016}
}
Kostenlos starten →