EUVIMEDEUVIMED entdecken
Open Access

Has Wrought Cobalt-Chromium-Molybdenum Alloy Changed for the Worse Over Time?

Stephanie M McCarthy, Deborah J Hall, Steven P Mell, Brett R Levine, Joshua J Jacobs, Robin Pourzal

The Journal of Arthroplasty · 2023

DOI: 10.1016/j.arth.2023.03.091 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of Arthroplasty, 2023. Verfasst von Stephanie M McCarthy und 5 weiteren Autor:innen. Beteiligte Einrichtung: Rush University Medical Center, Chicago, Illinois. Hauptthemen laut MEDLINE: Hip Prosthesis, Arthroplasty, Replacement, Hip. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 26 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar. Gefördert von: NIAMS NIH HHS, American Association of Hip and Knee Surgeons, National Institutes of Health.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Total hip arthroplasty (THA) failure due to tribocorrosion of modular junctions and resulting adverse local tissue reactions to corrosion debris have seemingly increased over the past few decades. Recent studies have found that chemically-induced column damage seen on the inner head taper is enabled by banding in the alloy microstructure of wrought cobalt-chromium-molybdenum alloy femoral heads, and is associated with more material loss than other tribocorrosion processes. It is unclear if alloy banding represents a recent phenomenon. The purpose of this study was to examine THAs implanted in the 1990s, 2000s, and 2010s to determine if alloy microstructure and implant susceptibility to severe damage has increased over time. METHODS: Five hundred and forty-five modular heads were assessed for damage severity and grouped based on decade of implantation to serve as a proxy measure for manufacturing date. A subset of heads (n = 120) was then processed for metallographic analysis to visualize alloy banding. RESULTS: We found that damage score distribution was consistent over the time periods, but the incidence of column damage significantly increased between the 1990s and 2000s. Banding also increased from the 1990s to 2000s, but both column damage and banding levels appear to recover slightly in the 2010s. CONCLUSION: Banding, which provides preferential corrosion sites enabling column damage, has increased over the last 3 decades. No difference between manufacturers was seen, which may be explained by shared suppliers of bar stock material. These findings are important as banding can be avoidable, reducing the risk of severe column damage to THA modular junctions and failure due to adverse local tissue reactions.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Hip ProsthesisArthroplasty, Replacement, HipHumansVitalliumChromium AlloysFemur HeadCorrosionProsthesis FailureProsthesis DesignCobalt

Stichwörter der Autor:innen

CoCrMoimplant alloysimplant materialsimplant metallurgytotal hip arthroplasty

Substanzen

VitalliumChromium AlloysCobalt

Autor:innen und Institutionen

  • Rush University Medical Center, Chicago, Illinois.

Förderung

  • NIAMS NIH HHS
  • American Association of Hip and Knee Surgeons
  • National Institutes of Health
  • National Institute of Arthritis and Musculoskeletal and Skin Diseases

Zitieren

APA

McCarthy, S. M., Hall, D. J., Mell, S. P., Levine, B. R., Jacobs, J. J., & Pourzal, R. (2023). Has Wrought Cobalt-Chromium-Molybdenum Alloy Changed for the Worse Over Time?. The Journal of Arthroplasty. https://doi.org/10.1016/j.arth.2023.03.091

Vancouver

McCarthy SM, Hall DJ, Mell SP, Levine BR, Jacobs JJ, Pourzal R. Has Wrought Cobalt-Chromium-Molybdenum Alloy Changed for the Worse Over Time?. The Journal of Arthroplasty. 2023. doi:10.1016/j.arth.2023.03.091

BibTeX
@article{McCarthy2023,
  author = {McCarthy, Stephanie M and Hall, Deborah J and Mell, Steven P and Levine, Brett R and Jacobs, Joshua J and Pourzal, Robin},
  title = {Has Wrought Cobalt-Chromium-Molybdenum Alloy Changed for the Worse Over Time?},
  journal = {The Journal of Arthroplasty},
  year = {2023},
  doi = {10.1016/j.arth.2023.03.091}
}
Kostenlos starten →