Switching From a 6° to a 20° Valgus Prosthetic Trochlear Groove Improved the Forgotten Joint and Oxford Knee Scores After Kinematically Aligned Total Knee Arthroplasty
Arthroplasty Today · 2026-02 · 37 · 101930
DOI: 10.1016/j.artd.2025.101930 ↗Kurz gefasst
Fachartikel in der Zeitschrift Arthroplasty Today, 2026-02. Verfasst von Stephen M Howell und 3 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, University of California, Davis, CA, USA. Die Arbeit wurde bisher 6-mal zitiert (Crossref) und nennt 41 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
In kinematically aligned (KA) total knee arthroplasty (TKA) using a femoral component with the traditional 6° valgus prosthetic trochlear groove (PTG), patients reported a lower Forgotten Joint Score (FJS) when the quadriceps line of pull was laterally misaligned to the groove, with an incidence of 89%. It remains unclear whether switching to a KA-optimized femoral component with a 20° valgus PTG, which properly aligns the quadriceps line of pull, can improve the FJS and the Oxford Knee Score (OKS).
Methods
The analysis of single-surgeon series of KA TKAs included 145 cases with a KA-optimized 20° valgus PTG and 292 cases with a 6° valgus PTG. Each participant reported their FJS and OKS at 2 years and underwent a postoperative coronal long-leg scan.
Results
The 20° group had a 6-point higher FJS (79) and a 16% lower incidence of poor FJS (
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Biomedical Engineering, University of California, Davis, CA, USA.
- University of California at Davis School of Medicine, Sacramento, CA, USA.
- Orthopädische Klinik König-Ludwig-Haus, Lehrstuhl für Orthopädie der Universität Würzburg, Würzburg, Germany.
- Department of Orthopedic Surgery, University of California, Davis, CA, USA.
- Department of Mechanical Engineering, University of California, Davis, CA, USA.
Zitieren
APA
Howell, S. M., Zabiba, A., Nedopil, A. J., & Hull, M. L. (2026). Switching From a 6° to a 20° Valgus Prosthetic Trochlear Groove Improved the Forgotten Joint and Oxford Knee Scores After Kinematically Aligned Total Knee Arthroplasty. Arthroplasty Today, 37, 101930. https://doi.org/10.1016/j.artd.2025.101930
Vancouver
Howell SM, Zabiba A, Nedopil AJ, Hull ML. Switching From a 6° to a 20° Valgus Prosthetic Trochlear Groove Improved the Forgotten Joint and Oxford Knee Scores After Kinematically Aligned Total Knee Arthroplasty. Arthroplasty Today. 2026;37:101930. doi:10.1016/j.artd.2025.101930
BibTeX
@article{Howell2026,
author = {Howell, Stephen M and Zabiba, Ahmed and Nedopil, Alexander J and Hull, Maury L},
title = {Switching From a 6° to a 20° Valgus Prosthetic Trochlear Groove Improved the Forgotten Joint and Oxford Knee Scores After Kinematically Aligned Total Knee Arthroplasty},
journal = {Arthroplasty Today},
year = {2026},
volume = {37},
pages = {101930},
doi = {10.1016/j.artd.2025.101930}
}