Quantitative evaluation of patient-emitted radiation in dual-injection Tc-99m MIBI myocardial perfusion scintigraphy: A multiregional, multi-distance, and scenario-based safety analysis

Ahmet Murat Şenışık, Handan Tanyıldızı Kökkülünk, Mahmut Yüksel

Applied Radiation and Isotopes · 2026-07 · 233 · 112554

DOI: 10.1016/j.apradiso.2026.112554 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Applied Radiation and Isotopes, 2026-07. Verfasst von Ahmet Murat Şenışık und 2 weiteren Autor:innen. Beteiligte Einrichtung: Altınbaş University, Vocational School of Health Services, Istanbul, Türkiye. Electronic address: ahmet.senisik@altinbas.edu.tr. Hauptthemen laut MEDLINE: Technetium Tc 99m Sestamibi, Myocardial Perfusion Imaging, Radiation Exposure, Radiopharmaceuticals. Die Arbeit nennt 20 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND Patients undergoing Tc-99m MIBI myocardial perfusion scintigraphy (MPS) temporarily act as external photon sources, creating short-term radiation exposure to staff and accompanying persons. Quantitative real-patient data describing multiregional and multi-distance dose-rate behavior under dual-injection protocols remain limited. PURPOSE To characterize patient-emitted radiation following dual-injection Tc-99m MIBI MPS and to estimate cumulative doses for realistic occupational and public-contact scenarios. METHODS Ambient dose equivalent rates (μSv/h) were measured at four anatomical levels (head, thorax, abdomen, pelvis) and five distances (25-200 cm) at three post-stress time points. Distance dependence was evaluated using power-law regression, and temporal clearance at 100 cm was modeled mono-exponentially to derive region-specific effective half-lives. Scenario-based cumulative doses were calculated using both median (typical) and 95th percentile (conservative) dose rates. RESULTS Dose rates decreased consistently with distance in all regions, with excellent regression fits (R2 ≥ 0.995) and empirical attenuation exponents of 0.53-0.76, consistent with extended-source behavior. Effective half-lives ranged from 4.65 to 5.69 h. Scenario-based cumulative exposures remained in the microsievert range, including ≤1.14 μSv for typical technologist contact and ≤9.69 μSv for conservative prolonged caregiver proximity. CONCLUSION Dual-injection Tc-99m MIBI MPS results in very low external radiation exposure to staff and accompanying persons. Measured doses are far below regulatory constraints, supporting the adequacy of routine ALARA-based clinical practice without additional restrictions.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Technetium Tc 99m SestamibiMyocardial Perfusion ImagingRadiation ExposureRadiopharmaceuticalsHumansOccupational ExposureRadiation DosageMale

Stichwörter der Autor:innen

Dose rateMyocardial perfusion scintigraphyOccupational exposurePatient-emitted radiationRadiation protectionTc-99m MIBI

Substanzen

Technetium Tc 99m SestamibiRadiopharmaceuticals

Autor:innen und Institutionen

  • Altınbaş University, Vocational School of Health Services, Istanbul, Türkiye. Electronic address: ahmet.senisik@altinbas.edu.tr.
  • Department of Arts & Sciences, Maine Maritime Academy, Castine, ME, 04420, USA. Electronic address: handan.kokkulunk@mma.edu.
  • Altınbaş University Hospital, Department of Nuclear Medicine, Istanbul, Türkiye. Electronic address: mahmut.yuksel@altinbas.edu.tr.

Zitieren

APA

Şenışık, A. M., Kökkülünk, H. T., & Yüksel, M. (2026). Quantitative evaluation of patient-emitted radiation in dual-injection Tc-99m MIBI myocardial perfusion scintigraphy: A multiregional, multi-distance, and scenario-based safety analysis. Applied Radiation and Isotopes, 233, 112554. https://doi.org/10.1016/j.apradiso.2026.112554

Vancouver

Şenışık AM, Kökkülünk HT, Yüksel M. Quantitative evaluation of patient-emitted radiation in dual-injection Tc-99m MIBI myocardial perfusion scintigraphy: A multiregional, multi-distance, and scenario-based safety analysis. Applied Radiation and Isotopes. 2026;233:112554. doi:10.1016/j.apradiso.2026.112554

BibTeX
@article{enk2026,
  author = {Şenışık, Ahmet Murat and Kökkülünk, Handan Tanyıldızı and Yüksel, Mahmut},
  title = {Quantitative evaluation of patient-emitted radiation in dual-injection Tc-99m MIBI myocardial perfusion scintigraphy: A multiregional, multi-distance, and scenario-based safety analysis},
  journal = {Applied Radiation and Isotopes},
  year = {2026},
  volume = {233},
  pages = {112554},
  doi = {10.1016/j.apradiso.2026.112554}
}
Kostenlos starten →