Image Quality Variation with Gantry Rotation Time and Reconstruction Algorithm in Ultra-high-resolution CT
Academic Radiology · 2025
DOI: 10.1016/j.acra.2025.09.027 ↗Kurz gefasst
Fachartikel in der Zeitschrift Academic Radiology, 2025. Verfasst von Minori Hoshika und 4 weiteren Autor:innen. Beteiligte Einrichtung: Graduate School of Health Sciences, Kumamoto University, 4-24-1 Kuhonji, Chuo-ku, Kumamoto 862-0976, Japan (M.H.). Electronic address: 192w5005@st.ku…. Hauptthemen laut MEDLINE: Tomography, X-Ray Computed, Algorithms, Radiographic Image Interpretation, Computer-Assisted, Image Processing, Computer-Assisted. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
RATIONALE AND OBJECTIVES: In ultra-high-resolution CT (U-HRCT), longer gantry rotation times are sometimes used to maintain image quality when using a small focal spot. This study aimed to evaluate the impact of gantry rotation time on image quality for deep learning reconstruction (DLR), model-based iterative reconstruction (MBIR), and filtered back projection (FBP). MATERIALS AND METHODS: A phantom was scanned on a U-HRCT scanner at four dose levels and four gantry rotation times, with images reconstructed using DLR, MBIR, and FBP algorithms. Image quality was evaluated for noise characteristics and high-contrast resolution. Noise was characterized using the noise power spectrum (NPS) to compute the noise magnitude ratio and central frequency ratio for MBIR and DLR relative to FBP, while high-contrast resolution was determined from the profile curve. RESULTS: MBIR and FBP demonstrated consistent image quality across all rotation times, with no statistically significant differences observed. In contrast, DLR showed significantly lower high-contrast resolution at a 1.0 s rotation time compared to 0.5-0.75 s (p
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Graduate School of Health Sciences, Kumamoto University, 4-24-1 Kuhonji, Chuo-ku, Kumamoto 862-0976, Japan (M.H.). Electronic address: 192w5005@st.kumamoto-u.ac.jp.
- Department of Radiological Technology, Tohoku University Hospital, Sendai, Japan (S.K.).
- Department of Medical Support, Division of Radiology, Okayama University Hospital, Okayama, Japan (N.A., T.I.).
- Department of Medical Image Analysis, Faculty of Life Sciences, Kumamoto University, Kumamoto, Japan (Y.F.).
Zitieren
APA
Hoshika, M., Kayano, S., Akagi, N., Inoue, T., & Funama, Y. (2025). Image Quality Variation with Gantry Rotation Time and Reconstruction Algorithm in Ultra-high-resolution CT. Academic Radiology. https://doi.org/10.1016/j.acra.2025.09.027
Vancouver
Hoshika M, Kayano S, Akagi N, Inoue T, Funama Y. Image Quality Variation with Gantry Rotation Time and Reconstruction Algorithm in Ultra-high-resolution CT. Academic Radiology. 2025. doi:10.1016/j.acra.2025.09.027
BibTeX
@article{Hoshika2025,
author = {Hoshika, Minori and Kayano, Shingo and Akagi, Noriaki and Inoue, Tomohiro and Funama, Yoshinori},
title = {Image Quality Variation with Gantry Rotation Time and Reconstruction Algorithm in Ultra-high-resolution CT},
journal = {Academic Radiology},
year = {2025},
doi = {10.1016/j.acra.2025.09.027}
}
EUVIMED