Advancing Neurovascular Imaging: Optimization of Reconstruction Kernel and Quantum Iterative Reconstruction for Ultra-high-resolution Photon-Counting Detector CT Angiography of the Head and Neck

Adrienn Tóth, Jennifer Yongjoo Cho, John Crow, Evan Wilson, Kimberly Kicielinski, Sami Al Kasab, Jennifer Joyce, Maria Gisele Matheus, Eric Bass, Maria Vittoria Spampinato

Academic Radiology · 2025

DOI: 10.1016/j.acra.2025.07.033 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Academic Radiology, 2025. Verfasst von Adrienn Tóth und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston 29425, SC (A.T., J.Y., J.C., E.W., J.J., M.G.M., E…. Hauptthemen laut MEDLINE: Computed Tomography Angiography, Radiographic Image Interpretation, Computer-Assisted, Intracranial Aneurysm, Head. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 28 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

RATIONALE AND OBJECTIVES: To find the optimal reconstruction parameters for ultra-high-resolution (UHR) photon-counting detector CT (PCD-CT) angiography of the head and neck, with a special focus on imaging intracranial aneurysms. MATERIAL AND METHODS: 18 patients with intracranial aneurysms were prospectively enrolled in this single-center study. CT angiograms were acquired in UHR mode on a clinical PCD-CT scanner. Images were reconstructed with six strength levels of a dedicated neurovascular kernel (Hv48-89) and with quantum iterative reconstruction (QIR) levels 1-4. Image noise, contrast-to-noise ratio, and vessel sharpness were determined for all reconstructions. Qualitative image quality was assessed by three readers using a 5-point Likert scale, for the best-performing reconstructions. Aneurysm dome and neck sizes were independently measured by two readers to assess inter-reader reliability. RESULTS: 18 patients (mean age, 64.3 years ± 15 [SD], 5 men,) were evaluated. Three kernels (Hv56, Hv64, Hv72) were identified as best performing in the quantitative analysis. The qualitative analysis demonstrated a consistent preference for QIR level 4 across all kernels in each evaluated category (P

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Computed Tomography AngiographyRadiographic Image Interpretation, Computer-AssistedIntracranial AneurysmHeadNeckCerebral AngiographyHumansMaleMiddle AgedFemaleProspective StudiesReproducibility of ResultsAgedAlgorithmsPhotons

Stichwörter der Autor:innen

Image qualityImage reconstructionIntracranial aneurysmPhoton-counting detector CTUltra-high-resolution

Autor:innen und Institutionen

  • Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston 29425, SC (A.T., J.Y., J.C., E.W., J.J., M.G.M., E.B., M.V.S.); Medical Imaging Centre, Semmelweis University, Budapest,…
  • Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston 29425, SC (A.T., J.Y., J.C., E.W., J.J., M.G.M., E.B., M.V.S.).
  • Department of Neurosurgery, Medical University of South Carolina, Charleston 29425, SC (K.P.K., S.A.K.).
  • Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston 29425, SC (A.T., J.Y., J.C., E.W., J.J., M.G.M., E.B., M.V.S.). Electronic address: spampin@musc.edu.

Zitieren

APA

Tóth, A., Cho, J. Y., Crow, J., Wilson, E., Kicielinski, K., Al Kasab, S., Joyce, J., Matheus, M. G., Bass, E., & Spampinato, M. V. (2025). Advancing Neurovascular Imaging: Optimization of Reconstruction Kernel and Quantum Iterative Reconstruction for Ultra-high-resolution Photon-Counting Detector CT Angiography of the Head and Neck. Academic Radiology. https://doi.org/10.1016/j.acra.2025.07.033

Vancouver

Tóth A, Cho JY, Crow J, Wilson E, Kicielinski K, Al Kasab S, et al. Advancing Neurovascular Imaging: Optimization of Reconstruction Kernel and Quantum Iterative Reconstruction for Ultra-high-resolution Photon-Counting Detector CT Angiography of the Head and Neck. Academic Radiology. 2025. doi:10.1016/j.acra.2025.07.033

BibTeX
@article{Tth2025,
  author = {Tóth, Adrienn and Cho, Jennifer Yongjoo and Crow, John and Wilson, Evan and Kicielinski, Kimberly and Al Kasab, Sami and Joyce, Jennifer and Matheus, Maria Gisele and Bass, Eric and Spampinato, Maria Vittoria},
  title = {Advancing Neurovascular Imaging: Optimization of Reconstruction Kernel and Quantum Iterative Reconstruction for Ultra-high-resolution Photon-Counting Detector CT Angiography of the Head and Neck},
  journal = {Academic Radiology},
  year = {2025},
  doi = {10.1016/j.acra.2025.07.033}
}
Kostenlos starten →