Automated Kidney Stone Composition Analysis with Photon-Counting Detector CT, a Performance Study—A Phantom Study
Academic Radiology · 2025
DOI: 10.1016/j.acra.2024.10.045 ↗Kurz gefasst
Fachartikel in der Zeitschrift Academic Radiology, 2025. Verfasst von Daniel Dillinger und 10 weiteren Autor:innen. Beteiligte Einrichtung: Department of Vascular and Endovascular Surgery, Federal Armed Services Hospital Koblenz, Koblenz, Germany (D.D.); Department of Radiology and Neuror…. Hauptthemen laut MEDLINE: Kidney Calculi, Tomography, X-Ray Computed, Radiographic Image Interpretation, Computer-Assisted, Pattern Recognition, Automated. Die Arbeit nennt 42 Literaturangaben.
Abstract (Originalsprache)
BACKGROUND: For treatment of urolithiasis, the stone composition is of particular interest, as uric acid (UA) stones can be treated by chemolitholysis. In this ex vivo study, we employed an advanced composition analysis approach for urolithiasis utilizing spectral data obtained from a photon-counting detector CT (PCDCT) to differentiate UA and non-UA stones. Our primary objective was to assess the accuracy of this analysis method. METHODS: A total of 148 urinary stones with a known composition that was measured by the standard reference method infrared spectroscopy (reference) were placed in an abdomen phantom and scanned in the PCDCT. Our objectives were to assess the stone detection rates of PCDCT and the accuracy of the prediction of the stone composition in UA vs non-UA compared to the reference. RESULTS: Automated detection recognized 86.5% of all stones, with best detection rate for stones larger > 5 mm in diameter (95.4%, 88.8% for stones larger than 3 mm, 94.7% for stones larger than 4 mm). Depending on the volume, we found a recognition rate of 92.8% for stones larger than 20 mm3 and 94.0% for stones with more than 30 mm3. Prediction of UA composition showed an overall sensitivity and a positive predictive value of 66.7% and a specificity and negative predictive value of 94.5%. Best diagnostic values volume wise were found by only including stones with a larger volume than 30 mm3, there we found a sensitivity of 91.7%, and a specificity of 92.4%. Sensitivity in dependance of the largest diameter was best for stones larger than 5 mm (85.7%), but specificity decreased with increasing diameter (to 91.3%). CONCLUSION: Automated urinary stone composition analysis with PCDCT showed a good automated detection rate of 86.5% up to 95.4% depending on stone diameter. The differentiation between non-UA and UA stones is performed with an NPV of 94.5% and a PPV of 66.7%. The prediction probability of non-UA stones was very good. This means the automatic detection and differentiation algorithm can identify the patients which will not profit from chemolitholysis.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Vascular and Endovascular Surgery, Federal Armed Services Hospital Koblenz, Koblenz, Germany (D.D.); Department of Radiology and Neuroradiology, Federal Armed Services Hospital Koblenz, Koblenz, Germany (D…
- Department of Radiology and Neuroradiology, Federal Armed Services Hospital Koblenz, Koblenz, Germany (D.D., S.W., D.O.); Institute of Neuroradiology, University Medical Centre Johannes Gutenberg University Mainz, Mainz…
- Department of Radiology and Neuroradiology, Federal Armed Services Hospital Koblenz, Koblenz, Germany (D.D., S.W., D.O.); Department of Radiology and Nuclear Medicine, University Medical Center Mannheim, Manheim, German…
- Department of Computed Tomography, Siemens Healthineers AG, Forchheim, Germany (S.F., M.J., B.S.).
- Department of Urology, Urinary Stone Analysis Centre Bonn, Bonn, Germany (A.H.).
- Department of Urology, Federal Armed Services Hospital Koblenz, Koblenz, Germany (J.S., H.S., R.S., T.N.).
Zitieren
APA
Dillinger, D., Waldeck, S., Overhoff, D., Faby, S., Jürgens, M., Schmidt, B., Hesse, A., Schoch, J., Schmelz, H., Stoll, R., & Nestler, T. (2025). Automated Kidney Stone Composition Analysis with Photon-Counting Detector CT, a Performance Study—A Phantom Study. Academic Radiology. https://doi.org/10.1016/j.acra.2024.10.045
Vancouver
Dillinger D, Waldeck S, Overhoff D, Faby S, Jürgens M, Schmidt B, et al. Automated Kidney Stone Composition Analysis with Photon-Counting Detector CT, a Performance Study—A Phantom Study. Academic Radiology. 2025. doi:10.1016/j.acra.2024.10.045
BibTeX
@article{Dillinger2025,
author = {Dillinger, Daniel and Waldeck, Stephan and Overhoff, Daniel and Faby, Sebastian and Jürgens, Markus and Schmidt, Bernhard and Hesse, Albrecht and Schoch, Justine and Schmelz, Hans and Stoll, Rico and Nestler, Tim},
title = {Automated Kidney Stone Composition Analysis with Photon-Counting Detector CT, a Performance Study—A Phantom Study},
journal = {Academic Radiology},
year = {2025},
doi = {10.1016/j.acra.2024.10.045}
}
EUVIMED