Repeatability, reproducibility, and observer variability of body composition assessment with single-slice lumbar MRI
Discover Imaging · 2025
DOI: 10.1007/s44352-025-00005-2 ↗Kurz gefasst
Fachartikel in der Zeitschrift Discover Imaging, 2025. Verfasst von Magdalena Nowak und 10 weiteren Autor:innen. Die Arbeit wurde bisher 5-mal zitiert (Crossref) und nennt 61 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0). Gefördert von: Perspectum Ltd..
Abstract (Originalsprache)
The single-slice MRI at the third lumbar vertebra (L3) offers an efficient way to assess body composition but evidence on the reliability of this method is scarce. This study evaluates the accuracy and precision of the single-slice L3 method for adipose and muscle tissue measurements. The technical performance of single-slice (L3) MRI body composition measurements was assessed in a prospective, single-center study of 12 participants, focusing on scan-rescan repeatability, cross-scanner reproducibility, and analyst variability. Participants were scanned using Philips 1.5 T (reference scanner) and Philips 3 T scanners. Additionally, retrospective data from multiple imaging centers from 36 participants were analyzed to evaluate inter-device and inter-observer (analyst vs. radiologist) variability across a wide range of body types, scanner manufacturers (Siemens, Philips, GE) and field strengths (1.5 T and 3 T). Blinded analyses were performed for visceral adipose tissue (VAT) and subcutaneous adipose tissue (SAT) cross-sectional areas (CSA), VAT and SAT indices, VAT/SAT ratio, skeletal muscle CSA, skeletal muscle index (SMI), psoas muscle CSA, and psoas muscle index. Single-slice MRI-based body composition measurements showed high scan-rescan repeatability (CoV: 1.5%–7.9%, ICC: 0.97–1.0), with low repeatability coefficients (RC) across key metrics, including 12 cm 2 for SAT CSA, 15 cm 2 for VAT CSA, 5 cm 2 for skeletal muscle CSA, and 1.4 cm 2 for psoas muscle CSA. Cross-scanner reproducibility was consistent (CoV: 2.3%–15%, ICC: 0.90–1.0). Inter- and intra-analyst variability was minimal (CoV: 0.5%–5.0%, ICC: 0.98–1.0). Analyst-radiologist comparisons showed near-perfect correlations (r = 0.97–1.00, p < 0.001) and excellent reliability (ICC: 0.96–1.0). The results demonstrate that the MRI-based single-slice method at the L3 vertebral level provides accurate, repeatable, and reproducible measurements of adipose and muscle tissue across diverse body types, consistent between trained analysts and experienced radiologists. This study establishes technical validity of the single-slice L3 method and supports its potential for use in clinical and research settings.
Förderung
- Perspectum Ltd.
Zitieren
APA
Nowak, M., Núñez, L., Hill, C. E., Davis, T., Cardiel, L. F., Schooler, G. R., Mermekli, A., Salvati, R., Brears, H. T., Pansini, M., & Robson, M. D. (2025). Repeatability, reproducibility, and observer variability of body composition assessment with single-slice lumbar MRI. Discover Imaging. https://doi.org/10.1007/s44352-025-00005-2
Vancouver
Nowak M, Núñez L, Hill CE, Davis T, Cardiel LF, Schooler GR, et al. Repeatability, reproducibility, and observer variability of body composition assessment with single-slice lumbar MRI. Discover Imaging. 2025. doi:10.1007/s44352-025-00005-2
BibTeX
@article{Nowak2025,
author = {Nowak, Magdalena and Núñez, Luis and Hill, Charles Edward and Davis, Tom and Cardiel, Luis Felipe and Schooler, Gary Robert and Mermekli, Asimenia and Salvati, Roberto and Brears, Helena Thomaides and Pansini, Michele and Robson, Matthew David},
title = {Repeatability, reproducibility, and observer variability of body composition assessment with single-slice lumbar MRI},
journal = {Discover Imaging},
year = {2025},
doi = {10.1007/s44352-025-00005-2}
}
EUVIMED