Comparative analysis of eMC algorithm dose calculations using GATE validation: impact of tissue heterogeneity on electron beam dosimetry

Mohammed Rezzoug, Mustapha Zerfaoui, Yassine Oulhouq, Abdeslem Rrhioua, Omar Hamzaoui, Dikra Bakari

Physical and Engineering Sciences in Medicine · 2025-12 · 48 (4) · 2021-2041

DOI: 10.1007/s13246-025-01641-y ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Physical and Engineering Sciences in Medicine, online erschienen am 18.09.2025. Verfasst von Mohammed Rezzoug und 5 weiteren Autor:innen. Beteiligte Einrichtung: Laboratory of Theoretical Physics, Particles, Modeling and Energy (LPTPME), Faculty of Sciences, University Mohamed 1st, Boulevard Mohamed VI, BP 717…. Hauptthemen laut MEDLINE: Algorithms, Monte Carlo Method, Electrons, Radiometry. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 19 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

PURPOSE: Electron beam radiotherapy is a crucial modality for treating superficial tumors. Accurate dose calculation is essential for treatment efficacy and minimizing side effects. While Monte Carlo (MC) simulations are considered the gold standard for dose calculation, their computational cost can be prohibitive. The electron Monte Carlo (eMC) algorithm offers a faster alternative, but its accuracy, especially in heterogeneous environments, remains a concern. METHODS AND MATERIALS: This study compares electron beam dose distributions calculated using the eMC algorithm in a treatment planning system (TPS) with those obtained from full MC simulations using the GATE platform. We evaluated the eMC algorithm's performance across various electron energies (6, 9, and 12 MeV) and field sizes (6 × 6 cm2 to 20 × 20 cm2), in both homogeneous water phantoms and heterogeneous phantoms incorporating lung-equivalent and bone-equivalent materials. RESULTS: Results in homogeneous phantoms demonstrated generally good agreement between eMC and GATE, with some discrepancies observed in penumbra regions and at higher energies, particularly for larger field sizes. In heterogeneous phantoms, significant deviations were observed, particularly in lateral dose profiles near density interfaces and at higher beam energies, with percentage of points with less than 3% difference dropping considerably. CONCLUSION: These findings highlight the limitations of the eMC algorithm in accurately modeling complex tissue heterogeneities. While eMC provides acceptable accuracy in relatively simple scenarios, its performance degrades significantly in clinically realistic heterogeneous environments, necessitating caution in treatment planning and highlighting the ongoing need for improved dose calculation algorithms.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AlgorithmsMonte Carlo MethodElectronsRadiometryPhantoms, ImagingRadiotherapy DosageRadiotherapy Planning, Computer-AssistedHumans

Stichwörter der Autor:innen

Dose distributionElectron Monte Carlo (eMC) algorithmElectron beam radiotherapyHeterogeneous phantomMonte Carlo simulation

Autor:innen und Institutionen

  • Laboratory of Theoretical Physics, Particles, Modeling and Energy (LPTPME), Faculty of Sciences, University Mohamed 1st, Boulevard Mohamed VI, BP 717, 60000, Oujda, Morocco. med.rezzoug@ump.ac.ma.
  • Department of Radiation Oncology, Hassan II Oncology Centre, Mohammed VI University Hospital, Oujda, Morocco. med.rezzoug@ump.ac.ma.
  • Laboratory of Theoretical Physics, Particles, Modeling and Energy (LPTPME), Faculty of Sciences, University Mohamed 1st, Boulevard Mohamed VI, BP 717, 60000, Oujda, Morocco.
  • National Institute of Particle Physics and Its Application, Oujda, Morocco.
  • EGIS Laboratory, National School of Applied Sciences, University Mohamed 1st, Oujda, Morocco.

Zitieren

APA

Rezzoug, M., Zerfaoui, M., Oulhouq, Y., Rrhioua, A., Hamzaoui, O., & Bakari, D. (2025). Comparative analysis of eMC algorithm dose calculations using GATE validation: impact of tissue heterogeneity on electron beam dosimetry. Physical and Engineering Sciences in Medicine, 48(4), 2021-2041. https://doi.org/10.1007/s13246-025-01641-y

Vancouver

Rezzoug M, Zerfaoui M, Oulhouq Y, Rrhioua A, Hamzaoui O, Bakari D. Comparative analysis of eMC algorithm dose calculations using GATE validation: impact of tissue heterogeneity on electron beam dosimetry. Physical and Engineering Sciences in Medicine. 2025;48(4):2021-2041. doi:10.1007/s13246-025-01641-y

BibTeX
@article{Rezzoug2025,
  author = {Rezzoug, Mohammed and Zerfaoui, Mustapha and Oulhouq, Yassine and Rrhioua, Abdeslem and Hamzaoui, Omar and Bakari, Dikra},
  title = {Comparative analysis of eMC algorithm dose calculations using GATE validation: impact of tissue heterogeneity on electron beam dosimetry},
  journal = {Physical and Engineering Sciences in Medicine},
  year = {2025},
  volume = {48},
  number = {4},
  pages = {2021--2041},
  doi = {10.1007/s13246-025-01641-y}
}
Kostenlos starten →