EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY 4.0

A retrospective comparison of [18F]FDG radiation dose following a transition from conventional to long axial field of view PET/CT

Wei-Ting Jacky Chen, William I D Rae, Peter L Kench, Kathy P Willowson, Dale L Bailey, Elizabeth A Bailey, Heidi Fearnside, Eleanor Kelliher, Steven R Meikle

Physical and Engineering Sciences in Medicine · 2025

DOI: 10.1007/s13246-025-01588-0 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Physical and Engineering Sciences in Medicine, 2025. Verfasst von Wei-Ting Jacky Chen und 8 weiteren Autor:innen. Beteiligte Einrichtung: Discipline of Medical Imaging Science, Faculty of Medicine and Health, University of Sydney, Sydney, NSW, Australia. wei-ting.chen@sydney.edu.au. Hauptthemen laut MEDLINE: Positron Emission Tomography Computed Tomography, Fluorodeoxyglucose F18, Radiation Dosage. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 44 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Sydney.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Long axial field of view (LAFOV) PET/CT scanners (> 1 m axial FOV) provide an order of magnitude higher system sensitivity compared with conventional scanners. This creates opportunities for significant radiation dose reductions for patients, without loss of diagnostic image quality or increased scan time. This study aimed to investigate changes in radiation dose received by patients undergoing whole-body [ 18 F]FDG PET/CT studies at a metropolitan hospital following the transition from the Siemens Biograph mCT (21.8 cm axial FOV) to the Siemens Biograph Vision Quadra LAFOV PET/CT (106 cm axial FOV). For the mCT and Quadra, 484 and 554 patient studies were reviewed, respectively. The radiation dose from the PET component was derived from the recorded FDG dose, calculated based on ICRP recommendations, and scaled to patient weight. The CT dose was derived from the dose-length product. The median effective dose from the PET component for the mCT and Quadra was 6.2 (IQR 5.5–6.9) and 2.9 (IQR 2.8–3.6) mSv, respectively, and 5.7 (IQR 5.1–6.5) and 2.8 (IQR 2.4–3.4) mSv, respectively, when scaled to patient weight. The median effective dose from the CT component for the mCT and Quadra was 7.7 (IQR 6.2–9.4) and 7.6 (IQR 5.9–9.4) mSv, respectively. The total median effective dose combining PET and CT components for the mCT and Quadra was 13.9 (IQR 12.4–15.7) and 10.5 (IQR 9.4–12.3) mSv, respectively, and 13.5 (IQR 12.4–15.0) and 10.3 (IQR 9.3–11.9) mSv, respectively, when scaled to patient weight. While the effective dose from PET was approximately halved due to reduced injected activity, the CT effective dose remained relatively unchanged and is now the dominant source of radiation dose to the patient for LAFOV PET/CT.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Positron Emission Tomography Computed TomographyFluorodeoxyglucose F18Radiation DosageHumansRetrospective StudiesMaleFemaleMiddle AgedAgedAdult

Stichwörter der Autor:innen

18F-FDG PETEffective doseLong axial field-of-viewPositron emission tomography/computed tomography (PET/CT)

Substanzen

Fluorodeoxyglucose F18

Autor:innen und Institutionen

  • Discipline of Medical Imaging Science, Faculty of Medicine and Health, University of Sydney, Sydney, NSW, Australia. wei-ting.chen@sydney.edu.au.
  • Discipline of Medical Imaging Science, Faculty of Medicine and Health, University of Sydney, Sydney, NSW, Australia.
  • Medical Imaging Department, Prince of Wales Hospital, Sydney, NSW, Australia.
  • Department of Nuclear Medicine, Royal North Shore Hospital, Sydney, NSW, Australia.
  • Brain and Mind Centre, University of Sydney, Sydney, NSW, Australia.
  • Sydney Imaging Core Research Facility, University of Sydney, Sydney, NSW, Australia.

Förderung

  • University of Sydney

Zitieren

APA

Chen, W. J., Rae, W. I. D., Kench, P. L., Willowson, K. P., Bailey, D. L., Bailey, E. A., Fearnside, H., Kelliher, E., & Meikle, S. R. (2025). A retrospective comparison of [18F]FDG radiation dose following a transition from conventional to long axial field of view PET/CT. Physical and Engineering Sciences in Medicine. https://doi.org/10.1007/s13246-025-01588-0

Vancouver

Chen WJ, Rae WID, Kench PL, Willowson KP, Bailey DL, Bailey EA, et al. A retrospective comparison of [18F]FDG radiation dose following a transition from conventional to long axial field of view PET/CT. Physical and Engineering Sciences in Medicine. 2025. doi:10.1007/s13246-025-01588-0

BibTeX
@article{Chen2025,
  author = {Chen, Wei-Ting Jacky and Rae, William I D and Kench, Peter L and Willowson, Kathy P and Bailey, Dale L and Bailey, Elizabeth A and Fearnside, Heidi and Kelliher, Eleanor and Meikle, Steven R},
  title = {A retrospective comparison of [18F]FDG radiation dose following a transition from conventional to long axial field of view PET/CT},
  journal = {Physical and Engineering Sciences in Medicine},
  year = {2025},
  doi = {10.1007/s13246-025-01588-0}
}
Kostenlos starten →