EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Advantages of carbon fibre-reinforced polyetheretherketone over titanium implants in spine SBRT: a phantom evaluation

Simon K Goodall, Katherine Tonkin, Peter Rampant, Pejman Rowshan Farzad, Martin Ebert

Physical and Engineering Sciences in Medicine · 2025

DOI: 10.1007/s13246-025-01587-1 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physical and Engineering Sciences in Medicine, 2025. Verfasst von Simon K Goodall und 4 weiteren Autor:innen. Beteiligte Einrichtung: School of Physics, Mathematics, and Computing, Faculty of Engineering and Mathematical Sciences, University of Western Australia, Crawley, WA, 6009, …. Hauptthemen laut MEDLINE: Ketones, Polyethylene Glycols, Titanium, Phantoms, Imaging. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: University of Western Australia.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study evaluates the clinical feasibility of spine stereotactic body radiotherapy (SBRT) in the presence of titanium and carbon fibre-reinforced polyetheretherketone (CFR-PEEK) spinal implants using custom 3D-printed phantoms. The investigation focuses on the dosimetric accuracy, imaging challenges, and achievable localisation precision. Customised 3D-printed phantoms incorporating titanium and CFR-PEEK implants were computed tomography (CT) scanned, with and without metal artefact reduction (MAR) algorithms. Localisation accuracy was tested using Elekta XVI CBCT and Brainlab ExacTrac Dynamic. The dosimetric accuracy of the Monaco treatment planning system (TPS) was assessed under simple geometric conditions and for clinically realistic VMAT plans. Patient-specific quality assurance and phantom-based measurements using ionization chambers and radiochromic film were performed to evaluate delivered dose accuracy. Both Image Guided Radiotherapy (IGRT) systems achieved sub-millimetre localisation accuracy, with maximum deviations of 0.3 mm observed for titanium implants. The Monaco treatment planning system (TPS) demonstrated accurate dose modelling, with deviations < 1% for CFR-PEEK and < 2% for titanium implants in simple homogeneous arrangements. In complex VMAT plan deliveries, dosimetric measurements showed stronger agreement with TPS predictions for CFR-PEEK implants, with deviations < 3%. Titanium-based plans exhibited greater deviations, with localised dose discrepancies exceeding clinical tolerances of 5%. The application of MAR algorithms reduced these discrepancies to < 5%, ensuring clinically acceptable dosimetric accuracy. CFR-PEEK implants enhance clinical workflows due to reduced imaging artefacts and smoother dose distributions, making MAR corrections unnecessary. For titanium implants, MAR is essential to achieve clinically acceptable dosimetric accuracy, highlighting the robustness of CFR-PEEK for spine SBRT.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

KetonesPolyethylene GlycolsTitaniumPhantoms, ImagingCarbon FiberProstheses and ImplantsRadiosurgerySpineCarbonBenzophenonesPolymersHumansRadiotherapy Planning, Computer-AssistedRadiotherapy DosagePrinting, Three-Dimensional

Stichwörter der Autor:innen

3D-printed phantomsCFR-PEEK implantsMetal artefact reduction (MAR)Spine SBRTTitanium implants

Substanzen

polyetheretherketoneBenzophenonesKetonesPolyethylene GlycolsPolymersTitaniumCarbon FiberCarbon

Autor:innen und Institutionen

  • School of Physics, Mathematics, and Computing, Faculty of Engineering and Mathematical Sciences, University of Western Australia, Crawley, WA, 6009, Australia. Simon.goodall@research.uwa.edu.au.
  • GenesisCare, 24 Salvado Road, Wembley, WA, 6014, Australia. Simon.goodall@research.uwa.edu.au.
  • GenesisCare, 24 Salvado Road, Wembley, WA, 6014, Australia.
  • School of Physics, Mathematics, and Computing, Faculty of Engineering and Mathematical Sciences, University of Western Australia, Crawley, WA, 6009, Australia.
  • Centre for Advanced Technologies in Cancer Research (CATCR), Perth, WA, 6000, Australia.
  • Department of Radiation Oncology, Sir Charles Gardiner Hospital, Nedlands, WA, 6009, Australia.

Förderung

  • University of Western Australia

Zitieren

APA

Goodall, S. K., Tonkin, K., Rampant, P., Rowshan Farzad, P., & Ebert, M. (2025). Advantages of carbon fibre-reinforced polyetheretherketone over titanium implants in spine SBRT: a phantom evaluation. Physical and Engineering Sciences in Medicine. https://doi.org/10.1007/s13246-025-01587-1

Vancouver

Goodall SK, Tonkin K, Rampant P, Rowshan Farzad P, Ebert M. Advantages of carbon fibre-reinforced polyetheretherketone over titanium implants in spine SBRT: a phantom evaluation. Physical and Engineering Sciences in Medicine. 2025. doi:10.1007/s13246-025-01587-1

BibTeX
@article{Goodall2025,
  author = {Goodall, Simon K and Tonkin, Katherine and Rampant, Peter and Rowshan Farzad, Pejman and Ebert, Martin},
  title = {Advantages of carbon fibre-reinforced polyetheretherketone over titanium implants in spine SBRT: a phantom evaluation},
  journal = {Physical and Engineering Sciences in Medicine},
  year = {2025},
  doi = {10.1007/s13246-025-01587-1}
}
Kostenlos starten →