Curvature correction factors for the independent verification of monitor units of electron treatment plans calculated in Eclipse

Luke A Slama, Talat Mahmood, Brendan Mckernan

Physical and Engineering Sciences in Medicine · 2024

DOI: 10.1007/s13246-024-01421-0 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physical and Engineering Sciences in Medicine, 2024. Verfasst von Luke A Slama und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, Sir Charles Gairdner Hospital, B Block, Hospital Ave, Nedlands, WA, 6009, Australia. Luke.Slama@health.wa.gov.au. Hauptthemen laut MEDLINE: Electrons, Radiotherapy Planning, Computer-Assisted, Monte Carlo Method, Phantoms, Imaging. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 20 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Electron beam dosimetry is sensitive to the surface contour of the patient. Over 10% difference between Treatment Planning System (TPS) and independent monitor-unit (IMU) calculations have been reported in the literature. Similar results were observed in our clinic between Radformation ClearCalc IMU and Eclipse TPS electron Monte Carlo (eMC) algorithm (v.16.1). This paper presents data measured under 3D printed spherical and cylindrical phantoms to validate the eMC algorithm in the presence of curved geometries. Measurements were performed with multiple detectors and compared to calculations made in Eclipse for the 6, 9 and 12 MeV electron energies. This data is used to create curvature correction factors (CCFs), defined as the ratio of the detector reading with the curved-surface phantom to a flat phantom at the same depth. The mean difference between the TPS calculated and measured CCFs using the NACP, Diode E, microSilicon, and microDiamond detectors were 1.3, 0.9, 0.7 and 0.7% respectively, with maximum differences of 4.5, 2.3, 1.9, and 1.8% respectively. Applying CCFs to previous failing patient IMU calculations improved agreement to the TPS. CCFs were implemented in our clinic for patient-specific IMU calculations with the assistance of a ESAPI script.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

ElectronsRadiotherapy Planning, Computer-AssistedMonte Carlo MethodPhantoms, ImagingAlgorithmsHumansRadiometryRadiotherapy Dosage

Stichwörter der Autor:innen

ElectronsIndependent MU calculationPatient Specific Quality AssuranceTreatment planning

Autor:innen und Institutionen

  • Department of Radiation Oncology, Sir Charles Gairdner Hospital, B Block, Hospital Ave, Nedlands, WA, 6009, Australia. Luke.Slama@health.wa.gov.au.
  • Department of Radiation Oncology, Sir Charles Gairdner Hospital, B Block, Hospital Ave, Nedlands, WA, 6009, Australia.

Zitieren

APA

Slama, L. A., Mahmood, T., & Mckernan, B. (2024). Curvature correction factors for the independent verification of monitor units of electron treatment plans calculated in Eclipse. Physical and Engineering Sciences in Medicine. https://doi.org/10.1007/s13246-024-01421-0

Vancouver

Slama LA, Mahmood T, Mckernan B. Curvature correction factors for the independent verification of monitor units of electron treatment plans calculated in Eclipse. Physical and Engineering Sciences in Medicine. 2024. doi:10.1007/s13246-024-01421-0

BibTeX
@article{Slama2024,
  author = {Slama, Luke A and Mahmood, Talat and Mckernan, Brendan},
  title = {Curvature correction factors for the independent verification of monitor units of electron treatment plans calculated in Eclipse},
  journal = {Physical and Engineering Sciences in Medicine},
  year = {2024},
  doi = {10.1007/s13246-024-01421-0}
}
Kostenlos starten →