EUVIMEDEUVIMED entdecken
Crossref

A Monte Carlo study on dose distribution of an orthovoltage radiation therapy system

Toshie Iseri, Yoshinori Tanabe, Ryouta Onizuka, Yuri Torigoe, Hiro Horikirizono, Kazuhito Itamoto, Hiroshi Sunahara, Harumichi Itoh, Kenji Tani, Munekazu Nakaichi

Physical and Engineering Sciences in Medicine · 2023

DOI: 10.1007/s13246-023-01237-4 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Physical and Engineering Sciences in Medicine, 2023. Verfasst von Toshie Iseri und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Veterinary Radiology, Joint Faculty of Veterinary Medicine, Yamaguchi University, 1677-1 Yoshida, Yamaguchi-shi, Yamaguchi, Yamagishi, …. Hauptthemen laut MEDLINE: Radiotherapy Planning, Computer-Assisted, Brain Neoplasms. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 19 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

It is important to plan radiotherapy treatment and establish optimal dose distribution to reduce the chances of side effects and injury. Because there are no commercially available tools for calculating dose distribution in orthovoltage radiotherapy in companion animals, we developed an algorithm to accomplish this and verified its characteristics using tumor disease cases. First, we used the Monte Carlo method to develop an algorithm to calculate the dose distribution of orthovoltage radiotherapy (280 kVp; MBR-320, Hitachi Medical Corporation, Tokyo, Japan) using BEAMnrc at our clinic. Using development of Monte Carlo method, dose distribution for tumor and normal organs were evaluated in brain tumors, squamous cell carcinomas of the head, and feline nasal lymphomas. In all cases of brain tumors, the mean dose delivered to the GTV ranged from 36.2 to 76.1% of the prescribed dose due to the decrease through the skull. In the nasal lymphoma in cats, the eyes with covered a 2 mm-thick lead plate, the respective average dose to the eyes was 71.8% and 89.9% less than that to the uncovered eyes. The findings may be useful for informed decision making in orthovoltage radiotherapy with more effective and targeted irradiation and data collection allowing detailed informed consent.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Radiotherapy Planning, Computer-AssistedBrain NeoplasmsAnimalsCatsRadiotherapy DosageMonte Carlo MethodAlgorithms

Stichwörter der Autor:innen

Monte Carlo methodOrgans at riskRadiation treatmentRadiotherapy dosage

Autor:innen und Institutionen

  • Department of Veterinary Radiology, Joint Faculty of Veterinary Medicine, Yamaguchi University, 1677-1 Yoshida, Yamaguchi-shi, Yamaguchi, Yamagishi, Japan. iseri@yamaguchi-u.ac.jp.
  • Graduate School of Health Sciences, Okayama University, Okayama, Japan.
  • Department of Therapeutic Radiology, Yamaguchi University Hospital, Yamaguchi, Japan.
  • Department of Veterinary Radiology, Joint Faculty of Veterinary Medicine, Yamaguchi University, 1677-1 Yoshida, Yamaguchi-shi, Yamaguchi, Yamagishi, Japan.
  • Animal medical center, Joint faculty of Veterinary Medicine, Yamaguchi University, Yamaguchi, Japan.
  • Department of Veterinary Surgery, Joint faculty of Veterinary Medicine, Yamaguchi University, Yamaguchi, Japan.

Zitieren

APA

Iseri, T., Tanabe, Y., Onizuka, R., Torigoe, Y., Horikirizono, H., Itamoto, K., Sunahara, H., Itoh, H., Tani, K., & Nakaichi, M. (2023). A Monte Carlo study on dose distribution of an orthovoltage radiation therapy system. Physical and Engineering Sciences in Medicine. https://doi.org/10.1007/s13246-023-01237-4

Vancouver

Iseri T, Tanabe Y, Onizuka R, Torigoe Y, Horikirizono H, Itamoto K, et al. A Monte Carlo study on dose distribution of an orthovoltage radiation therapy system. Physical and Engineering Sciences in Medicine. 2023. doi:10.1007/s13246-023-01237-4

BibTeX
@article{Iseri2023,
  author = {Iseri, Toshie and Tanabe, Yoshinori and Onizuka, Ryouta and Torigoe, Yuri and Horikirizono, Hiro and Itamoto, Kazuhito and Sunahara, Hiroshi and Itoh, Harumichi and Tani, Kenji and Nakaichi, Munekazu},
  title = {A Monte Carlo study on dose distribution of an orthovoltage radiation therapy system},
  journal = {Physical and Engineering Sciences in Medicine},
  year = {2023},
  doi = {10.1007/s13246-023-01237-4}
}
Kostenlos starten →