On the Material Constitutive Behavior of the Aortic Root in Patients with Transcatheter Aortic Valve Implantation
Cardiovascular Engineering and Technology · 2024
DOI: 10.1007/s13239-023-00699-7 ↗Kurz gefasst
Fachartikel in der Zeitschrift Cardiovascular Engineering and Technology, 2024. Verfasst von Chiara Catalano und 8 weiteren Autor:innen. Beteiligte Einrichtung: Department of Engineering, Università degli Studi di Palermo, Viale delle Scienze, Palermo, Italy. Hauptthemen laut MEDLINE: Transcatheter Aortic Valve Replacement, Aortic Valve Stenosis, Heart Valve Prosthesis. Die Arbeit wurde bisher 12-mal zitiert (Crossref) und nennt 29 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: H2020 Excellent Science, Università degli Studi di Palermo.
Abstract (Originalsprache)
Background
Transcatheter aortic valve implantation (TAVI) is a minimally invasive procedure used to treat patients with severe aortic valve stenosis. However, there is limited knowledge on the material properties of the aortic root in TAVI patients, and this can impact the credibility of computer simulations. This study aimed to develop a non-invasive inverse approach for estimating reliable material constituents for the aortic root and calcified valve leaflets in patients undergoing TAVI.
Methods
The identification of material parameters is based on the simultaneous minimization of two cost functions, which define the difference between model predictions and cardiac-gated CT measurements of the aortic wall and valve orifice area. Validation of the inverse analysis output was performed comparing the numerical predictions with actual CT shapes and post-TAVI measures of implanted device diameter.
Results
A good agreement of the peak systolic shape of the aortic wall was found between simulations and imaging, with similarity index in the range in the range of 83.7% to 91.5% for n.20 patients. Not any statistical difference was observed between predictions and CT measures of orifice area for the stenotic aortic valve. After TAVI simulations, the measurements of SAPIEN 3 Ultra (S3) device diameter were in agreement with those from post-TAVI angio-CT imaging. A sensitivity analysis demonstrated a modest impact on the S3 diameters when altering the elastic material property of the aortic wall in the range of inverse analysis solution.
Conclusions
Overall, this study demonstrates the feasibility and potential benefits of using non-invasive imaging techniques and computational modeling to estimate material properties in patients undergoing TAVI.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Engineering, Università degli Studi di Palermo, Viale delle Scienze, Palermo, Italy.
- 4RealSim Services BV, Groene Dijk 2B, 3401 NJ, IJsselstein, The Netherlands.
- Department for the Treatment and Study of Cardiothoracic Diseases and Cardiothoracic Transplantation, IRCCS-ISMETT, Palermo, Italy.
- Radiology Unit, Department of Diagnostic and Therapeutic Services, IRCCS-ISMETT, Palermo, Italy.
- 3D printing and Virtual Reality Laboratory, Department of Research, IRCCS-ISMETT, IRCCS Mediterranean Institute for Transplantation and Advanced Specialized Therapies, Via Tricomi, 5, Palermo, Italy.
- Department of Engineering, Università degli Studi di Palermo, Viale delle Scienze, Palermo, Italy. salvatore.pasta@unipa.it.
ORCID-Profile: Salvatore Pasta
Förderung
- H2020 Excellent Science
- Università degli Studi di Palermo
Zitieren
APA
Catalano, C., Turgut, T., Zahalka, O., Götzen, N., Cannata, S., Gentile, G., Agnese, V., Gandolfo, C., & Pasta, S. (2024). On the Material Constitutive Behavior of the Aortic Root in Patients with Transcatheter Aortic Valve Implantation. Cardiovascular Engineering and Technology. https://doi.org/10.1007/s13239-023-00699-7
Vancouver
Catalano C, Turgut T, Zahalka O, Götzen N, Cannata S, Gentile G, et al. On the Material Constitutive Behavior of the Aortic Root in Patients with Transcatheter Aortic Valve Implantation. Cardiovascular Engineering and Technology. 2024. doi:10.1007/s13239-023-00699-7
BibTeX
@article{Catalano2024,
author = {Catalano, Chiara and Turgut, Tahir and Zahalka, Omar and Götzen, Nils and Cannata, Stefano and Gentile, Giovanni and Agnese, Valentina and Gandolfo, Caterina and Pasta, Salvatore},
title = {On the Material Constitutive Behavior of the Aortic Root in Patients with Transcatheter Aortic Valve Implantation},
journal = {Cardiovascular Engineering and Technology},
year = {2024},
doi = {10.1007/s13239-023-00699-7}
}
EUVIMED